| Literature DB >> 32030363 |
Kenneth A Beckman1, Jodi I Luchs2, Mark S Milner3, Richard W Yee4.
Abstract
PURPOSE: To evaluate the extent to which rebound tonometry affects corneal surface properties and preoperative corneal measurements.Entities:
Keywords: Icare ic100; cataract surgery; corneal measurements; rebound tonometry; refractive surgery; same day intraocular pressure testing
Year: 2020 PMID: 32030363 PMCID: PMC6977213 DOI: 10.1177/2515841419892070
Source DB: PubMed Journal: Ther Adv Ophthalmol ISSN: 2515-8414
Figure 1.The step-by-step approach employed during the study. Fifteen patients were included from each of the four sites, for a total of 60 patients (120 eyes).
None of the 60 study eyes developed increased staining scores following IOP testing with the Icare ic100, and for both mean keratometry and total corneal cylinder, no statistically significant difference was found from the first measurement to the second measurement between the study eyes and control eyes.
| Study | PCKM | PC TotalCyl | IOLM KM | IOLM Total Cyl | Stain | IOP |
|---|---|---|---|---|---|---|
| Avg Pre | 43.647 | 0.840 | 43.697 | 0.901 | 0.233 | 14.900 |
|
| 1.667 | 0.554 | 1.700 | 0.585 | 0.563 | 3.776 |
| Avg Post | 43.622 | 0.828 | 43.721 | 0.896 | 0.217 | |
|
| 1.692 | 0.529 | 1.752 | 0.574 | 0.555 | |
| Avg Diff | 0.025 | 0.012 | 0.024 | 0.005 | 0.017 | |
| Control | PCKM | PC TotalCyl | IOLM KM | IOLM Total Cyl | Stain | IOP |
| Avg Pre | 43.600 | 0.882 | 43.728 | 0.917 | 0.250 | |
|
| 1.784 | 0.662 | 1.780 | 0.742 | 0.541 | |
| Avg Post | 43.647 | 0.968 | 43.725 | 0.951 | 0.305 | 14.610 |
|
| 1.777 | 0.655 | 1.776 | 0.724 | 0.561 | 4.131 |
| Avg Diff | 0.047 | 0.086 | 0.003 | 0.035 | 0.055 | |
| Diff-in-diff | 0.022 | 0.074 | 0.021 | 0.030 | 0.038 | |
| 0.962 | 0.634 | 0.963 | 0.861 | 0.790 | ||
Avg Diff, average difference in absolute value between first and second measurement; Diff-in-Diff, difference in absolute value between the Avg Diff for the study eyes and the control eyes; IOLM KM, IOL Master mean keratometry; IOLM Total Cyl, IOL Master total corneal cylinder; IOP, intraocular pressure; PCKM, Pentacam mean keratometry; PC TotalCyl, Pentacam total corneal cylinder; SD, standard deviation.