Shihoko Iwata1, Michiyoshi Sanuki2, Makoto Ozaki3.
Abstract
BACKGROUND: The patient state index (PSI) is a parameter of a four-channel electroencephalography (EEG)-derived variable used to assess the depth of anesthesia. A PSI value of 25-50 indicates adequate state of hypnosis, and a value of 100 indicates a fully awake state. Due to reduced interference from electronic devices like electrocautery, falsely high intraoperative PSI values are rarely reported. However, this case report cautions about falsely high PSI during cardiopulmonary bypass (CPB) with intra-aortic balloon pumping (IABP). CASEEntities:
Keywords: Artifact; Cardiopulmonary bypass; Diagnostic error; Electroencephalography; Intra-aortic balloon pumping; Patient state index
Year: 2019 PMID: 32026977 PMCID: PMC6967226 DOI: 10.1186/s40981-019-0299-x
Source DB: PubMed Journal: JA Clin Rep ISSN: 2363-9024
Fig. 1Patient state index from ascending aorta cannulation to after cardiopulmonary bypass. Patient state index was remarkably decreased after stopping intra-aortic balloon pumping (IABP) before partial clamping of the ascending aorta for anastomosis with the saphenous vein (a), rapidly increased after de-clamping the ascending aorta and restarting IABP (b), decreased after stopping IABP for hemostasis around the anastomosis (c), and again rapidly increased corresponding to the restart of IABP (d). Time 0 corresponds to ascending aortic cannulation
Fig. 2The display window of SedLine® without intra-aortic balloon pumping. Electroencephalogram (EEG) (top), patient state index (PSI) and suppression ratio (middle), and color density spectral array (DSA) (bottom) are shown. PSI was 14 and suppression ratio was 32 with almost flat EEG at the time point C in Fig. 1. The vertical white dotted lines indicate each second, and a horizontal white line shows 1 s in EEG. DSA indicates time (x-axis), frequency (y-axis), and power (color scale) with 95% spectral edge frequency (white line). Vertical black lines with a blue bar at 0 Hz timeline in DSA, conveying periods of suppression, were observed when PSI was extremely low
Fig. 3The display window of SedLine® after restarting intra-aortic balloon pumping. Patient state index (PSI) was increased from 14 to 80 after restarting intra-aortic balloon pumping (IABP) before weaning from cardiopulmonary bypass. Regular oscillation noises at about 60 Hz corresponding to the frequency of IABP (white arrow) are observed on almost flat EEG (top) at the time point D in Fig. 1. The vertical white dotted lines indicate each second, and a horizontal white line shows 1 s in EEG. Ninety-five percent spectral edge frequency was 9.4 Hz and 9.2 Hz on the left and right side, respectively. The density spectral array trend indicates the predominance of low frequency waves