Characterization of irregular electron beam for boost dose after whole breast irradiation.

Ayat M Saadeldin1, Azhar M Elwan2.   

Abstract

Irradiating a tumor bed with boost dose after whole breast irradiation helps reducing the probability of local recurrence. However, the success of electron beam treatment with a small area aiming to cover a superficial lesion is a dual challenge as it requires an adequate dosimetry beside a double check for dose coverage with an estimation of various combined uncertainty of tumor location and losing lateral electron equilibrium within small field dimensions. AIM OF WORK: this work aims to measure the electron beam fluence within different field dimensions and the deviation from measurement performed in standard square electron applicator beam flatness and symmetry, then to calculate the average range of the correction factor required to overcome the loss of lateral electron equilibrium. MATERIAL AND
METHOD: the electron beam used in this work generated from the linear accelerator model ELEKTA Precise and dosimetry system used were a pair of PTW Pin Point ion chambers for electron beam dosimetry at standard conditions and assessment of beam quality at a reference depth of measurement, with an automatic water phantom, then a Roos ion chamber was used for absolute dose measurement, and PTW 2Darray to investigate the beam fluence of four applicators 6, 10, 14 and 20 cm2 and 4 rectangular cutouts 6 × 14, 8 × 14, 6 × 17 and 8 × 17 cm2, the second part was clinical application which was performed in a precise treatment planning system and examined boost dose after whole breast irradiation.
RESULTS: revealed that lower energy (6MeV and 8MeV) showed the loss of lateral electron equilibrium and deviation from measurements of a standard applicator more than the high energy (15 MeV) which indicated that the treatment of superficial dose with 6MeV required higher monitor unit to allow for the loss of lateral electron equilibrium and higher margin as well.
© 2020 Greater Poland Cancer Centre. Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Boost dose; Electron beam; Small field; Whole breast irradiation

Year:  2020        PMID: 32021571      PMCID: PMC6994277          DOI: 10.1016/j.rpor.2020.01.007

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  22 in total

1.  Scattered radiation from applicators in clinical electron beams.

Authors:  L J van Battum; W van der Zee; H Huizenga
Journal:  Phys Med Biol       Date:  2003-08-07       Impact factor: 3.609

2.  The nth root percent depth dose method for calculating monitor units for irregularly shaped electron fields.

Authors:  Than S Kehwar; M Saiful Huq
Journal:  Med Phys       Date:  2008-04       Impact factor: 4.071

3.  Comparison of modulated electron radiotherapy to conventional electron boost irradiation and volumetric modulated photon arc therapy for treatment of tumour bed boost in breast cancer.

Authors:  Andrew Alexander; Emilie Soisson; Tarek Hijal; Arman Sarfehnia; Jan Seuntjens
Journal:  Radiother Oncol       Date:  2011-07-06       Impact factor: 6.280

4.  A sector-integration method for calculating the output factors of irregularly shaped electron fields.

Authors:  P A Jursinic; R Mueller
Journal:  Med Phys       Date:  1997-11       Impact factor: 4.071

5.  Adjuvant intraoperative radiotherapy for selected breast cancers in previously irradiated women: Evidence for excellent feasibility and favorable outcomes.

Authors:  Gladys Blandino; Marina Guenzi; Liliana Belgioia; Elisabetta Bonzano; Elena Configliacco; Elena Tornari; Francesca Cavagnetto; Davide Bosetti; Alessandra Fozza; Daniele Friedman; Renzo Corvò
Journal:  Rep Pract Oncol Radiother       Date:  2017-05-05

6.  Dosimetric impact of setup accuracy for an electron breast boost technique.

Authors:  Scott Davidson; Steven Kirsner; Bryan Mason; Kelly Kisling; Renee D Barrett; Anthony Bonetati; Matthew T Ballo
Journal:  Pract Radiat Oncol       Date:  2015-04-06

7.  Toward optimizing patient-specific IMRT QA techniques in the accurate detection of dosimetrically acceptable and unacceptable patient plans.

Authors:  Elizabeth M McKenzie; Peter A Balter; Francesco C Stingo; Jimmy Jones; David S Followill; Stephen F Kry
Journal:  Med Phys       Date:  2014-12       Impact factor: 4.071

8.  First treatment of mycosis fungoides by total skin electron beam (TSEB) therapy in Greece.

Authors:  Stefanos Diamantopoulos; Kalliopi Platoni; Vassilis Kouloulias; Panayiotis Pantelakos; Maria Dilvoi; Evangelia Papadavid; Christina Antoniou; Efstathios Efstathopoulos
Journal:  Rep Pract Oncol Radiother       Date:  2013-08-13

9.  Electron beam characteristics at extended source-to-surface distances for irregular cut-outs.

Authors:  T Arunkumar; Sanjay S Supe; M Ravikumar; S Sathiyan; M Ganesh
Journal:  J Med Phys       Date:  2010-10

Review 10.  Techniques of tumour bed boost irradiation in breast conserving therapy: current evidence and suggested guidelines.

Authors:  Rakesh Jalali; Suruchi Singh; Ashwini Budrukkar
Journal:  Acta Oncol       Date:  2007       Impact factor: 4.089

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.