A sector-integration method for calculating the output factors of irregularly shaped electron fields.

P A Jursinic1, R Mueller.   

Abstract

An empirical model is presented that uses a sector-integration method for calculating the output factors of irregularly shaped electron fields. The sector-integration method accounts for changes in electron fluence, lateral scatter equilibrium, and scatter from the edge of a cutout shield. This method is tested for elliptical and rectangular fields with a ratio of the long-to-short axis as great as 4 to 1. Differences between measured and calculated values for output factors were less than +/- 1%. Comparisons were also carried out for a large number of cutout shields that were used in the clinic and similar levels of accuracy were obtained.

Mesh:

Year:  1997        PMID: 9394283     DOI: 10.1118/1.597962

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

Review 1.  Some computer graphical user interfaces in radiation therapy.

Authors:  James C L Chow
Journal:  World J Radiol       Date:  2016-03-28

2.  Characterization of irregular electron beam for boost dose after whole breast irradiation.

Authors:  Ayat M Saadeldin; Azhar M Elwan
Journal:  Rep Pract Oncol Radiother       Date:  2020-01-22

3.  Comparison of two programs in calculating electron output factors at extended source-surface distances.

Authors:  Yunfei Hu; Jonathan Andrew Lambert; Yang Wang
Journal:  J Med Phys       Date:  2016 Jan-Mar

4.  A graphical user interface for an electron monitor unit calculator using a sector-integration algorithm and exponential curve-fitting method.

Authors:  James C L Chow; Grigor N Grigorov; Christopher MacGregor
Journal:  J Appl Clin Med Phys       Date:  2006-02-15       Impact factor: 2.102

5.  Convolution-based modified Clarkson integration (CMCI) for electron cutout factor calculation.

Authors:  Jina Chang; Mu-Han Lin; Weiguo Lu; Mingli Chen; Steve Jiang
Journal:  J Appl Clin Med Phys       Date:  2018-02-03       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.