Dosimetric impact of setup accuracy for an electron breast boost technique.

Scott Davidson1, Steven Kirsner2, Bryan Mason2, Kelly Kisling2, Renee D Barrett3, Anthony Bonetati3, Matthew T Ballo4.   

Abstract

PURPOSE: To determine the setup error on an electron breast boost technique using daily cone beam computed tomography (CBCT). Patient and setup attributes were studied as contributing factors to the accuracy. METHODS AND MATERIALS: Reproducibility of a modified lateral decubitus position breast boost setup was verified for 33 patients using CBCT. Three-dimensional matching was performed between the CBCT and the initial planning CT for each boost fraction by matching the tumor bed and/or surgical clips. The dosimetric impact of the daily positioning error was achieved by rerunning the initial treatment plans incorporating the recorded shifts to study the dose differences. Breast compression, decubitus angle, tumor bed location and volume, and cup size were studied for their contribution to setup error.
RESULTS: The range of setup errors was: 1.5 cm anterior to 9 mm posterior, 1.3 cm superior to 2.3 cm inferior, and 3.2 cm medial to 2.4 cm lateral. Seven patients had setup errors that were ≥2-cm margin placed on the tumor bed and scar. Four of those 7 patients had unacceptable coverage as defined by the volume of the tumor bed plus scar that is covered by the 90% isodose line (V90) compared with the original plan. All other patients had no discernible difference in the coverage (V90). The use of compression, tumor bed location, or volumes >20 mL showed no effect on coverage.
CONCLUSIONS: In general, this study supported that a 2-cm margin was adequate (29 of 33 patients) when patients are treated under typical conditions. Care should be taken when high electron energies are selected because the coverage at depth is more difficult to maintain in the clinical environment.
Copyright © 2015 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2015        PMID: 25858772     DOI: 10.1016/j.prro.2015.02.011

Source DB:  PubMed          Journal:  Pract Radiat Oncol        ISSN: 1879-8500


  1 in total

1.  Characterization of irregular electron beam for boost dose after whole breast irradiation.

Authors:  Ayat M Saadeldin; Azhar M Elwan
Journal:  Rep Pract Oncol Radiother       Date:  2020-01-22
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.