Experimental demonstration of direct L-shell x-ray fluorescence imaging of gold nanoparticles using a benchtop x-ray source.

Nivedh Manohar, Francisco J Reynoso, Sang Hyun Cho.   

Abstract

PURPOSE: To develop a proof-of-principle L-shell x-ray fluorescence (XRF) imaging system that locates and quantifies sparse concentrations of gold nanoparticles (GNPs) using a benchtop polychromatic x-ray source and a silicon (Si)-PIN diode x-ray detector system.
METHODS: 12-mm-diameter water-filled cylindrical tubes with GNP concentrations of 20, 10, 5, 0.5, 0.05, 0.005, and 0 mg∕cm3 served as calibration phantoms. An imaging phantom was created using the same cylindrical tube but filled with tissue-equivalent gel containing structures mimicking a GNP-loaded blood vessel and approximately 1 cm3 tumor. Phantoms were irradiated by a 3-mm-diameter pencil-beam of 62 kVp x-rays filtered by 1 mm aluminum. Fluorescence∕scatter photons from phantoms were detected at 90° with respect to the beam direction using a Si-PIN detector placed behind a 2.5-mm-diameter lead collimator. The imaging phantom was translated horizontally and vertically in 0.3-mm steps to image a 6 mm×15 mm region of interest (ROI). For each phantom, the net L-shell XRF signal from GNPs was extracted from background, and then corrected for detection efficiency and in-phantom attenuation using a fluorescence-to-scatter normalization algorithm.
RESULTS: XRF measurements with calibration phantoms provided a calibration curve showing a linear relationship between corrected XRF signal and GNP mass per imaged voxel. Using the calibration curve, the detection limit (at the 95% confidence level) of the current experimental setup was estimated to be a GNP mass of 0.35 μg per imaged voxel (1.73×10(-2) cm3). A 2D XRF map of the ROI was also successfully generated, reasonably matching the known spatial distribution as well as showing the local variation of GNP concentrations.
CONCLUSIONS: L-shell XRF imaging can be a highly sensitive tool that has the capability of simultaneously imaging the spatial distribution and determining the local concentration of GNPs presented on the order of parts-per-million level within subcentimeter-sized ex vivo samples and superficial tumors during preclinical animal studies.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23927295      PMCID: PMC3745523          DOI: 10.1118/1.4816297

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  18 in total

1.  Estimation of strontium in animal bone using x-ray fluorescence analysis.

Authors:  W M ROBERTS
Journal:  Nature       Date:  1959-03-28       Impact factor: 49.962

2.  Investigation of X-ray fluorescence computed tomography (XFCT) and K-edge imaging.

Authors:  Magdalena Bazalova; Yu Kuang; Guillem Pratx; Lei Xing
Journal:  IEEE Trans Med Imaging       Date:  2012-06-05       Impact factor: 10.048

3.  The feasibility of polychromatic cone-beam x-ray fluorescence computed tomography (XFCT) imaging of gold nanoparticle-loaded objects: a Monte Carlo study.

Authors:  Bernard L Jones; Sang Hyun Cho
Journal:  Phys Med Biol       Date:  2011-05-31       Impact factor: 3.609

4.  A method of measuring gold nanoparticle concentrations by X-ray fluorescence for biomedical applications.

Authors:  Di Wu; Yuhua Li; Molly D Wong; Hong Liu
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

5.  Use of portable X-ray fluorescence (PXRF) in vivo as an alternative technique for the assessment of iron levels in patients with thalassemia and hemochromatosis.

Authors:  Marcelo Estevam; Carlos Roberto Appoloni
Journal:  Health Phys       Date:  2013-02       Impact factor: 1.316

6.  A quantitative x-ray detection system for gold nanoparticle tumour biomarkers.

Authors:  K Ricketts; A Castoldi; C Guazzoni; C Ozkan; C Christodoulou; A P Gibson; G J Royle
Journal:  Phys Med Biol       Date:  2012-08-08       Impact factor: 3.609

7.  The use of gold nanoparticles to enhance radiotherapy in mice.

Authors:  James F Hainfeld; Daniel N Slatkin; Henry M Smilowitz
Journal:  Phys Med Biol       Date:  2004-09-21       Impact factor: 3.609

8.  X-ray fluorescence camera for imaging of iodine media in vivo.

Authors:  Hiroshi Matsukiyo; Manabu Watanabe; Eiichi Sato; Akihiro Osawa; Toshiyuki Enomoto; Jiro Nagao; Purkhet Abderyim; Katsuo Aizawa; Etsuro Tanaka; Hidezo Mori; Toshiaki Kawai; Shigeru Ehara; Shigehiro Sato; Akira Ogawa; Jun Onagawa
Journal:  Radiol Phys Technol       Date:  2008-11-11

9.  In vivo tumor targeting and spectroscopic detection with surface-enhanced Raman nanoparticle tags.

Authors:  Ximei Qian; Xiang-Hong Peng; Dominic O Ansari; Qiqin Yin-Goen; Georgia Z Chen; Dong M Shin; Lily Yang; Andrew N Young; May D Wang; Shuming Nie
Journal:  Nat Biotechnol       Date:  2007-12-23       Impact factor: 54.908

10.  First demonstration of multiplexed X-ray fluorescence computed tomography (XFCT) imaging.

Authors:  Yu Kuang; Guillem Pratx; Magdalena Bazalova; Bowen Meng; Jianguo Qian; Lei Xing
Journal:  IEEE Trans Med Imaging       Date:  2012-10-11       Impact factor: 10.048

View more
  12 in total

1.  Experimental validation of L-shell x-ray fluorescence computed tomography imaging: phantom study.

Authors:  Magdalena Bazalova-Carter; Moiz Ahmad; Lei Xing; Rebecca Fahrig
Journal:  J Med Imaging (Bellingham)       Date:  2015-10-08

Review 2.  The potential of L-shell X-ray fluorescence CT (XFCT) for molecular imaging.

Authors:  Magdalena Bazalova-Carter
Journal:  Br J Radiol       Date:  2015-07-24       Impact factor: 3.039

3.  Development of bimetallic (Zn@Au) nanoparticles as potential PET-imageable radiosensitizers.

Authors:  Jongmin Cho; Min Wang; Carlos Gonzalez-Lepera; Osama Mawlawi; Sang Hyun Cho
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

4.  L-shell x-ray fluorescence computed tomography (XFCT) imaging of Cisplatin.

Authors:  Magdalena Bazalova; Moiz Ahmad; Guillem Pratx; Lei Xing
Journal:  Phys Med Biol       Date:  2013-12-13       Impact factor: 3.609

5.  Feasibility study of Compton cameras for x-ray fluorescence computed tomography with humans.

Authors:  Don Vernekohl; Moiz Ahmad; Garry Chinn; Lei Xing
Journal:  Phys Med Biol       Date:  2016-11-15       Impact factor: 3.609

6.  Technical Note: A benchtop cone-beam x-ray fluorescence computed tomography (XFCT) system with a high-power x-ray source and transmission CT imaging capability.

Authors:  Nivedh Manohar; Francisco J Reynoso; Sang Hyun Cho
Journal:  Med Phys       Date:  2018-09-14       Impact factor: 4.071

Review 7.  Roadmap to Clinical Use of Gold Nanoparticles for Radiation Sensitization.

Authors:  Jan Schuemann; Ross Berbeco; Devika B Chithrani; Sang Hyun Cho; Rajiv Kumar; Stephen J McMahon; Srinivas Sridhar; Sunil Krishnan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-09-30       Impact factor: 7.038

8.  Development of an attenuation correction method for direct x-ray fluorescence (XRF) imaging utilizing gold L-shell XRF photons.

Authors:  Md Foiez Ahmed; Selcuk Yasar; Sang Hyun Cho
Journal:  Med Phys       Date:  2018-11-08       Impact factor: 4.071

9.  High-sensitivity imaging and quantification of intratumoral distributions of gold nanoparticles using a benchtop x-ray fluorescence imaging system.

Authors:  Nivedh Manohar; Francisco Reynoso; Sandun Jayarathna; Hem Moktan; Md Foiez Ahmed; Parmeswaran Diagaradjane; Sunil Krishnan; Sang Hyun Cho
Journal:  Opt Lett       Date:  2019-11-01       Impact factor: 3.776

10.  Quantitative X-ray fluorescence imaging of gold nanoparticles using joint L1 and total variation regularized reconstruction.

Authors:  Junwei Shi; Blaine Granger; Keying Xu; Yidong Yang
Journal:  Quant Imaging Med Surg       Date:  2020-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.