| Literature DB >> 31949913 |
Selina L S Storm1, Adam D Crawshaw1, Nicholas E Devenish1, Rachel Bolton1,2, David R Hall1, Ivo Tews2, Gwyndaf Evans1.
Abstract
With the increasing trend of using microcrystals and intense microbeams at synchrotron X-ray beamlines, radiation damage becomes a more pressing problem. Theoretical calculations show that the photoelectrons that primarily cause damage can escape microcrystals. This effect would become more pronounced with decreasing crystal size as well as at higher energies. To prove this effect, data from cryocooled lysozyme crystals of dimensions 5 × 3 × 3 and 20 × 8 × 8 µm mounted on cryo-transmission electron microscopy (cryo-TEM) grids were collected at 13.5 and 20.1 keV using a PILATUS CdTe 2M detector, which has a similar quantum efficiency at both energies. Accurate absorbed doses were calculated through the direct measurement of individual crystal sizes using scanning electron microscopy after the experiment and characterization of the X-ray microbeam. The crystal lifetime was then quantified based on the D 1/2 metric. In this first systematic study, a longer crystal lifetime for smaller crystals was observed and crystal lifetime increased at higher X-ray energies, supporting the theoretical predictions of photoelectron escape. The use of detector technologies specifically optimized for data collection at energies above 20 keV allows the theoretically predicted photoelectron escape to be quantified and exploited, guiding future beamline-design choices. © Selina L. S. Storm et al. 2020.Entities:
Keywords: CdTe detector; cryo-TEM sample mounts; microcrystals; photoelectron escape
Year: 2020 PMID: 31949913 PMCID: PMC6949606 DOI: 10.1107/S2052252519016178
Source DB: PubMed Journal: IUCrJ ISSN: 2052-2525 Impact factor: 4.769
Beam sizes (FWHM) of the Gaussian-shaped beam and the fluxes measured with a 500 µm thick silicon PIN diode for the three different sessions reported here
| Beamtime session | Small beam (µm) | Large beam (µm) | Flux at 13.5 keV (photons s−1) | Flux at 20.1 keV (photons s−1) |
|---|---|---|---|---|
| A | 9.1 × 8.2 | 21.9 × 18.2 | 2.2 × 1012 | 4.2 × 1011 |
| B | 7.2 × 6.4 | n.a. | 1.8 × 1012 | 4.9 × 1011 |
| C | n.a. | 23.4 × 20.5 | 2.3 × 1012 | 5.5 × 1011 |
Figure 1(a) A cryo-TEM grid held on a tweezer pin. (b) A cryo-TEM grid viewed in the on-axis viewing microscope overlaid with the coordinate system used to identify crystals in the scanning electron microscope (SEM) post X-ray exposure. (c) SEM image of HEWL crystal dimensions being determined.
Figure 2Determination of D 1/2 based on the fitted decay curve of the normalized intensities.
Number of crystals for which the D 1/2 value could be determined
Small crystals refer to an average crystal size of 5 × 3 × 3 µm and large crystals to an average crystal size of 20 × 8 × 8 µm.
| Energy (keV) | Small crystals | Large crystals | |
|---|---|---|---|
| 13.5 | 23 | 35 | |
| 20.1 | 14 | 34 | |
Figure 3(a) D 1/2 values for 22 crystals with an average size of 5.2 × 3.1 × 3.1 µm with a beam size of 6.4 × 7.2 µm collected from the same grid at 13.5 keV (blue) and at 20.1 keV (magenta) plotted with error bars representing measurement errors in flux, crystal size and fit of the decay curve. (b) As in (a) but plotted against the dose corrected for photoelectron escape, referred to as the deposited dose.
Figure 4Average D 1/2 values from all 106 crystals plotted for both crystal sizes and energies. The error bars correspond to the standard deviations of the D 1/2 values.