Pediatric chest computed tomography at 100 kVp with tin filtration: comparison of image quality with 70-kVp imaging at comparable radiation dose.

Sophie Vivier1, Valérie Deken2, Younes Arous1, Jean-Baptiste Faivre1, Alain Duhamel2, Antoine Deschildre3, Thomas Flohr4, Jacques Remy1, Martine Remy-Jardin5.   

Abstract

BACKGROUND: Radiation dose reduction is a primary objective in pediatric populations owing to the well-known risks of radiation-induced cancers. Low-energy photons participate in the radiation dose without significantly contributing to image formation. Their suppression by means of tin filtration should decrease the image noise, anticipating a subsequent application to dose saving.
OBJECTIVE: To investigate the level of noise reduction achievable with tin (Sn) filtration at 100 kVp for chest computed tomography (CT) in comparison with a standard scanning mode at 70 kVp with comparable radiation dose.
MATERIALS AND METHODS: Fifty consecutive children (Group 1) underwent non-contrast chest CT examinations on a third-generation dual-source CT system at tin-filtered 100 kVp and pitch 2. The tube-current time product (mAs) was adjusted to maintain the predicted dose length product (DLP) value at 70 kVp for the respective patient. Each child was then paired by weight and age to a child scanned at 70 kVp on the same CT unit (Group 2); Group 2 patients were consecutive patients, retrospectively selected from our database of children prospectively scanned at 70 kVp. Objective and subjective image quality were compared between the two groups of patients to investigate the overall image quality and level of noise reduction that could be subsequently achievable with tin filtration in clinical practice.
RESULTS: The mean image noise was significantly lower in Group 1 compared to Group 2 when measured in the air (P<0.0001) and inside the aorta (P<0.001). The mean noise reduction was 21.6% (standard deviation [SD] 16.1) around the thorax and 12.0% (SD 32.7) inside the thorax. There was no significant difference in subjective image quality of lung and mediastinal images with excellent overall subjective scores in both groups.
CONCLUSION: At comparable radiation dose, the image noise was found to be reduced by 21.6% compared to the 70-kVp protocol, providing basis for dose reduction without altering image quality in further investigations.

Entities:  

Keywords:  Chest; Children; Computed tomography; Radiation dose; Spectral shaping; Tin filtration

Year:  2019        PMID: 31853569     DOI: 10.1007/s00247-019-04543-w

Source DB:  PubMed          Journal:  Pediatr Radiol        ISSN: 0301-0449


  20 in total

1.  Unenhanced third-generation dual-source chest CT using a tin filter for spectral shaping at 100kVp.

Authors:  Holger Haubenreisser; Mathias Meyer; Sonja Sudarski; Thomas Allmendinger; Stefan O Schoenberg; Thomas Henzler
Journal:  Eur J Radiol       Date:  2015-04-28       Impact factor: 3.528

2.  Radiation dose levels in pediatric chest CT: experience in 499 children evaluated with dual-source single-energy CT.

Authors:  Remy-Jardin Martine; Teresa Santangelo; Lucie Colas; Faivre Jean-Baptiste; Alain Duhamel; Antoine Deschildre; Jacques Remy
Journal:  Pediatr Radiol       Date:  2016-11-09

3.  Cohort Profile: the EPI-CT study: a European pooled epidemiological study to quantify the risk of radiation-induced cancer from paediatric CT.

Authors:  Marie-Odile Bernier; Hélène Baysson; Mark S Pearce; Monika Moissonnier; Elisabeth Cardis; Michael Hauptmann; Lara Struelens; Jeremie Dabin; Christoffer Johansen; Neige Journy; Dominique Laurier; Maria Blettner; Lucian Le Cornet; Roman Pokora; Patrycja Gradowska; Johanna M Meulepas; Kristina Kjaerheim; Tore Istad; Hilde Olerud; Aste Sovik; Magda Bosch de Basea; Isabelle Thierry-Chef; Magnus Kaijser; Arvid Nordenskjöld; Amy Berrington de Gonzalez; Richard W Harbron; Ausrele Kesminiene
Journal:  Int J Epidemiol       Date:  2019-04-01       Impact factor: 7.196

4.  Pediatric chest CT at 70 kVp: a feasibility study in 129 children.

Authors:  Tilo Niemann; Simon Henry; Alain Duhamel; Jean-Baptiste Faivre; Antoine Deschildre; Lucie Colas; Teresa Santangelo; Jacques Remy; Martine Remy-Jardin
Journal:  Pediatr Radiol       Date:  2014-06-13

5.  Low-dose high-resolution CT of the chest in children and young adults: dose, cooperation, artifact incidence, and image quality.

Authors:  J Lucaya; J Piqueras; P García-Peña; G Enríquez; M García-Macías; J Sotil
Journal:  AJR Am J Roentgenol       Date:  2000-10       Impact factor: 3.959

6.  Radiation dose reduction with hybrid iterative reconstruction for pediatric CT.

Authors:  Sarabjeet Singh; Mannudeep K Kalra; Anuradha S Shenoy-Bhangle; Aashna Saini; Debra A Gervais; Sjirk J Westra; James H Thrall
Journal:  Radiology       Date:  2012-05       Impact factor: 11.105

7.  Evaluation of a radiation dose reduction strategy for pediatric chest CT.

Authors:  Jee-Eun Kim; Beverley Newman
Journal:  AJR Am J Roentgenol       Date:  2010-05       Impact factor: 3.959

8.  Dose reduction and compliance with pediatric CT protocols adapted to patient size, clinical indication, and number of prior studies.

Authors:  Sarabjeet Singh; Mannudeep K Kalra; Michael A Moore; Randheer Shailam; Bob Liu; Thomas L Toth; Ellen Grant; Sjirk J Westra
Journal:  Radiology       Date:  2009-05-12       Impact factor: 11.105

9.  Effects of automated kilovoltage selection technology on contrast-enhanced pediatric CT and CT angiography.

Authors:  Marilyn J Siegel; Charles Hildebolt; David Bradley
Journal:  Radiology       Date:  2013-04-05       Impact factor: 11.105

10.  Very low-dose (0.15 mGy) chest CT protocols using the COPDGene 2 test object and a third-generation dual-source CT scanner with corresponding third-generation iterative reconstruction software.

Authors:  John D Newell; Matthew K Fuld; Thomas Allmendinger; Jered P Sieren; Kung-Sik Chan; Junfeng Guo; Eric A Hoffman
Journal:  Invest Radiol       Date:  2015-01       Impact factor: 6.016

View more
  1 in total

1.  Low-dose CT with tin filter combined with iterative metal artefact reduction for guiding lung biopsy.

Authors:  Jing Zhang; Meiling Liu; Daihong Liu; Xiaoqin Li; Meng Lin; Yong Tan; Yuesheng Luo; Xiangfei Zeng; Hong Yu; Hesong Shen; Xiaoxia Wang; Leilei Liu; Yuchuan Tan; Jiuquan Zhang
Journal:  Quant Imaging Med Surg       Date:  2022-02
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.