Literature DB >> 20410401

Evaluation of a radiation dose reduction strategy for pediatric chest CT.

Jee-Eun Kim1, Beverley Newman.   

Abstract

OBJECTIVE: The purpose of our study was to quantify the effect of changes made to the CT chest protocol on patient dose, image quality, and image noise when using a kilovoltage (kVp)-lowering strategy.
MATERIALS AND METHODS: We retrospectively selected 120 children who underwent chest CT: 60 in 2006 and 60 in 2008. In each group there were 30 children weighing less than 15 kg and 30 between 15 and 60 kg. In 2006 the CT protocol was 120 kVp and the reference current (mAs) was 65. In 2008, the kVp was 80 for < 15 kg and 100 for 15-60 kg, with reference mAs of 55. For each examination, the volume CT dose index (CTDI(vol)) and dose-length product (DLP) were recorded. Effective dose (ED) was estimated using the DLP method. Image noise was measured. Overall image quality was subjectively evaluated.
RESULTS: For a weight < 15.0 kg, the CTDI(vol), DLP, and ED were reduced by 73%, 75%, and 73%, respectively (p < 0.05). For the weight range 15-60 kg, the CTDI(vol), DLP, and ED were reduced by 45%, 44%, and 48%, respectively (p < 0.05). Measured noise increased by 55% in the younger children and 41% in the older group (p < 0.05). All studies were considered diagnostically adequate.
CONCLUSION: Significant radiation dose reduction can be achieved for routine pediatric chest CT by weight-based decreases in kVp in addition to low mAs. Increased noise was considered an acceptable trade-off for decreased dose, and image quality was acceptable.

Entities:  

Mesh:

Year:  2010        PMID: 20410401     DOI: 10.2214/AJR.09.3726

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  28 in total

1.  Reply to commentary--'CT radiation dose reduction: can we do harm by doing good?'.

Authors:  Beverley Newman; Michael J Callahan
Journal:  Pediatr Radiol       Date:  2012-04

2.  ATS Core Curriculum 2015: Part III. Pediatric Pulmonary Medicine.

Authors:  Debra Boyer; Mary Nevin; Carey C Thomson; Don B Sanders; Stamatia Alexiou; Samuel B Goldfarb; Jennifer L Nicholas; Paul G Thacker; Andrea M Coverstone; Albert Faro; George Cheng; Adnan Majid; Paul E Moore
Journal:  Ann Am Thorac Soc       Date:  2015-11

3.  Image quality and dose optimisation for infant CT using a paediatric phantom.

Authors:  Jack W Lambert; Andrew S Phelps; Jesse L Courtier; Robert G Gould; John D MacKenzie
Journal:  Eur Radiol       Date:  2015-08-26       Impact factor: 5.315

4.  Third-generation dual-source 70-kVp chest CT angiography with advanced iterative reconstruction in young children: image quality and radiation dose reduction.

Authors:  Oliver Rompel; Martin Glöckler; Rolf Janka; Sven Dittrich; Robert Cesnjevar; Michael M Lell; Michael Uder; Matthias Hammon
Journal:  Pediatr Radiol       Date:  2016-01-06

5.  Accuracy, image quality and radiation dose comparison of high-pitch spiral and sequential acquisition on 128-slice dual-source CT angiography in children with congenital heart disease.

Authors:  Pei Nie; Ximing Wang; Zhaoping Cheng; Xiaopeng Ji; Yanhua Duan; Jiuhong Chen
Journal:  Eur Radiol       Date:  2012-05-18       Impact factor: 5.315

6.  Chest CT in children: anesthesia and atelectasis.

Authors:  Beverley Newman; Elliot J Krane; Rakhee Gawande; Tyson H Holmes; Terry E Robinson
Journal:  Pediatr Radiol       Date:  2013-11-08

7.  Rebalancing the risks of Computed Tomography and Magnetic Resonance imaging.

Authors:  S Bruce Greenberg
Journal:  Pediatr Radiol       Date:  2011-05-28

Review 8.  CT dose reduction in practice.

Authors:  Michael J Callahan
Journal:  Pediatr Radiol       Date:  2011-08-17

9.  Interruption of the aortic arch: diagnosis with multidetector computed tomography.

Authors:  Shuhei Sato; Noriaki Akagi; Mayu Uka; Katsuya Kato; Yoshihiro Okumura; Susumu Kanazawa
Journal:  Jpn J Radiol       Date:  2011-01-26       Impact factor: 2.374

10.  Radiation dose levels in pediatric chest CT: experience in 499 children evaluated with dual-source single-energy CT.

Authors:  Remy-Jardin Martine; Teresa Santangelo; Lucie Colas; Faivre Jean-Baptiste; Alain Duhamel; Antoine Deschildre; Jacques Remy
Journal:  Pediatr Radiol       Date:  2016-11-09