| Literature DB >> 31842191 |
Mads Forslund Jacobsen1,2, Lars Konge2, Mark Alberti1, Morten la Cour1, Yoon Soo Park3, Ann Sofia Skou Thomsen1,2.
Abstract
PURPOSE: To compare manual and robot-assisted vitreoretinal surgery using a virtual-reality surgical simulator.Entities:
Mesh:
Year: 2020 PMID: 31842191 PMCID: PMC7575030 DOI: 10.1097/IAE.0000000000002720
Source DB: PubMed Journal: Retina ISSN: 0275-004X Impact factor: 3.975
Fig. 1.Overview of the crossover study design.
Fig. 2.The robot assisted surgery setup. A. EyeSi simulator controls display. B. EyeSi simulator phantom head. C. Robot instrument manipulator. D. Robot motion controller. E. Robot controls display.
Overview of Included Eyesi Simulator Modules
| Module No. | Module Name | Type | Task Difficulty Level | Score |
| 1 | Bimanual | Abstract task | Level 5 out of 5 | 0–100 |
| 2 | ILM peeling | Procedural task | Level 5 out of 6 | 0–100 |
| 3 | ERM peeling | Procedural task | Level 4 out of 6 | 0–100 |
Level 1 being easiest and difficulty increasing with higher levels.
Descriptive Data on Study Participants (Mean Values)
| Age (Range) | Right Handed | Stereopsis TNO Test (Range) | Vitrectomies Performed | Retinal Detachment Surgeries Performed | Macular Surgeries Performed | |
| Novices | 36 years (31–42) | 60% | 60 seconds arc (60–60) | 0 | 0 | 0 |
| Vitreoretinal surgeons | 49 years (36–66) | 60% | 60 seconds arc (60–60) | 2,339 (100–8,000) | 995 (50–5000) | 733 (30–2000) |
Impact of Robot-Assisted Surgery Compared With That of Manual Surgery, Overall and for Each Primary Outcome for all Participants
| Score | Standardized Effect Size (Coef.) | SE | 95% Confidence Interval | |||
| Novices and vitreoretinal surgeons | Overall | −0.18 | 0.06 | 0.003 | −0.299 | −0.063 |
| Time | 0.49 | 0.08 | <0.001 | 0.335 | 0.638 | |
| Movement | −0.71 | 0.09 | <0.001 | −0.896 | −0.532 | |
| Tissue treatment | −0.32 | 0.12 | 0.007 | −0.545 | −0.085 | |
Impact of Robot-Assisted Surgery Compared With That of Manual for Each Primary Outcome for the Two Groups of Participants
| Score | Standardized Effect Size (Coef.) | SE | 95% Confidence Interval | |||
| Novices (n = 10) | Time | 0.11 | 0.032 | 0.456 | ||
| Movement | 0.17 | −1.294 | −0.628 | |||
| Tissue treatment | 0.23 | −1.032 | −0.144 | |||
| Vitreoretinal surgeons (n = 10) | Time | 0.10 | 0.525 | 0.932 | ||
| Movement | 0.06 | −0.595 | −0.340 | |||
| Tissue treatment | −0.042 | 0.04 | 0.333 | −0.126 | 0.043 | |
Bold entries mark statistical significance on a 0.05 level.
Fig. 3.Participant's impression of the ease of use of robot-assisted compared with that of manual surgery.
Fig. 4.Participant's impression of the safety of robot-assisted compared with that of manual surgery.