| Literature DB >> 31805699 |
Michael Alexander Wirth1, Gabriella Fischer1,2, Jorge Verdú3, Lisa Reissner1,4, Simone Balocco3,5, Maurizio Calcagni1.
Abstract
This study aims to compare a new inertial measurement unit based system with the highly accurate but complex laboratory gold standard, an optoelectronic motion capture system. Inertial measurement units are sensors based on accelerometers, gyroscopes, and/or magnetometers. Ten healthy subjects were recorded while performing flexion-extension and radial-ulnar deviation movements of their right wrist using inertial sensors and skin markers. Maximum range of motion during these trials and mean absolute difference between the systems were calculated. A difference of 10° ± 5° for flexion-extension and 2° ± 1° for radial-ulnar deviation was found between the two systems with absolute range of motion values of 126° and 50° in the respective axes. A Wilcoxon rank sum test resulted in a no statistical differences between the systems with p-values of 0.24 and 0.62. The observed results are even more precise than reports from previous studies, where differences between 14° and 27° for flexion-extension and differences between 6° and 17° for radial-ulnar deviation were found. Effortless and fast applicability, good precision, and low inter-observer variability make inertial measurement unit based systems applicable to clinical settings.Entities:
Keywords: clinical assessments; inertial measurement units; kinematics; motion analysis; optoelectronic motion capture; range of motion; wrist
Mesh:
Year: 2019 PMID: 31805699 PMCID: PMC6929166 DOI: 10.3390/s19235297
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1Sensor (IMU), skin marker (red arrows, OMC1) and sensor marker (dark blue arrows, OMC2) placement.
Figure A1OMC signals before and after interpolation. (a) Example of interpolation with a low number of missing samples. (b) Example of interpolation with a high number of missing samples.
Summary of the main results. Results in percentage represent the coefficient of variation (CV) using the first element of difference as a reference.
| ROM IMU (°) | ROM OMC1 (°) | ROM OMC2 (°) | MAD IMU-OMC1 (°) | MAD IMU-OMC2 (°) | MAD OMC1-OMC2 (°) | CV IMU-OMC1 (%) | CV IMU-OMC2 (%) | CV OMC1-OMC2 (%) | |
|---|---|---|---|---|---|---|---|---|---|
| Flexion-extension | 126.01 ± 18.85 | 135.96 ± 17.98 | 141.35 ± 20.64 | 10.37 ± 4.22 | 15.48 ± 5.55 | 6.21 ± 4.18 | 8.55 ± 3.59 | 12.42 ± 4.20 | 4.76 ± 2.85 |
| Radial-ulnar deviation | 50.48 ± 9.04 | 51.05 ± 8.44 | 49.91 ± 8.44 | 2.42 ± 1.19 | 2.11 ± 0.85 | 2.67 ± 1.41 | 4.86 ± 2.41 | 4.06 ± 1.06 | 5.42± 2.80 |
Average MAD ROM values between IMU and OMC1, IMU and OMC2 and OMC1 and OMC2 for flexion-extension and radial-ulnar deviation.
| Flexion-Extension | IMU-OMC1 (°) | IMU-OMC2 (°) | OMC1-OMC2 (°) | Radial-Ulnar Deviation | IMU-OMC1 (°) | IMU-OMC2 (°) | OMC1-OMC2 (°) |
|---|---|---|---|---|---|---|---|
| S1 | 3.04 | 12.00 | 11.10 | S1 | 2.87 | 3.94 | 1.26 |
| S2 | 5.07 | 5.66 | 1.46 | S2 | 2.07 | 1.49 | 1.77 |
| S3 | 10.04 | 10.43 | 2.76 | S3 | 1.57 | 2.08 | 1.16 |
| S4 | 10.08 | 23.85 | 13.77 | S4 | 1.50 | 2.06 | 2.13 |
| S5 | 10.02 | 11.42 | 1.85 | S5 | 1.70 | 2.80 | 4.01 |
| S6 | 15.40 | 19.84 | 4.45 | S6 | 4.86 | 1.21 | 4.70 |
| S7 | 7.20 | 17.34 | 10.93 | S7 | 2.80 | 2.58 | 3.87 |
| S8 | 17.36 | 22.36 | 7.27 | S8 | 4.19 | 2.48 | 4.80 |
| S9 | 12.89 | 18.85 | 5.96 | S9 | 1.38 | 0.82 | 1.42 |
| S10 | 12.63 | 13.03 | 2.56 | S10 | 1.25 | 1.68 | 1.63 |
| Mean | 10.37 | 15.48 | 6.21 | Mean | 2.42 | 2.11 | 2.67 |
| SD | 4.22 | 5.55 | 4.18 | SD | 1.19 | 0.85 | 1.41 |
Figure 2(a) Flexion-Extension average ROM values for IMU, OMC1 and OMC2 acquisition systems. (b) Radial-ulnar deviation average ROM values for IMU, OMC1 and OMC2 acquisition systems.
Flexion-Extension average ROM values for each subject with its respective mean and standard deviation. IMU, OMC1, and OMC2 represent Sensor, Vicon skin markers and Vicon sensor markers acquisition systems, respectively.
| Mean ROM IMU (°) | Mean ROM OMC1 (°) | Mean ROM OMC2 (°) | |
|---|---|---|---|
| S1 | 151.99 | 152.88 | 163.97 |
| S2 | 141.04 | 144.81 | 145.32 |
| S3 | 111.87 | 121.90 | 122.29 |
| S4 | 146.46 | 156.54 | 170.31 |
| S5 | 94.75 | 104.95 | 106.35 |
| S6 | 132.96 | 148.37 | 152.80 |
| S7 | 124.01 | 130.42 | 141.35 |
| S8 | 141.47 | 158.83 | 163.83 |
| S9 | 100.77 | 113.65 | 119.61 |
| S10 | 114.50 | 127.13 | 127.53 |
| Mean | 125.98 | 135.95 | 141.34 |
| SD | 19.88 | 18.95 | 21.74 |
Radial-ulnar deviation average ROM values for each subject with its respective mean and standard deviation. IMU, OMC1, and OMC2 represent Sensor, Vicon skin markers and Vicon sensor markers acquisition systems, respectively.
| Mean ROM IMU (°) | Mean ROM OMC1 (°) | Mean ROM OMC2 (°) | |
|---|---|---|---|
| S1 | 70.73 | 68.01 | 66.80 |
| S2 | 40.55 | 38.48 | 39.64 |
| S3 | 57.53 | 56.54 | 55.48 |
| S4 | 50.55 | 50.55 | 52.52 |
| S5 | 54.31 | 53.1 | 57.11 |
| S6 | 47.81 | 52.68 | 47.98 |
| S7 | 50.67 | 52.52 | 48.65 |
| S8 | 50.35 | 54.54 | 49.74 |
| S9 | 35.48 | 36.83 | 35.4 |
| S10 | 46.77 | 47.27 | 45.68 |
| Mean | 50.47 | 51.05 | 49.90 |
| SD | 9.54 | 8.90 | 8.91 |
Average standard deviation of maximum (MAX, wrist flexion) and minimum (MIN, wrist extension) for each subject. IMU, OMC1 and OMC2 depict Sensor, Vicon skin markers and Vicon sensor marker acquisition systems.
| SD Max IMU (°) | SD Min IMU (°) | SD Max OMC1 (°) | SD Min OMC1 (°) | SD Max OMC2 (°) | SD Min OMC2 (°) | |
|---|---|---|---|---|---|---|
| S1 | 1.67 | 2.41 | 1.20 | 2.04 | 1.26 | 2.74 |
| S2 | 1.43 | 1.91 | 1.13 | 1.91 | 1.10 | 1.24 |
| S3 | 2.66 | 2.42 | 2.42 | 2.66 | 2.85 | 2.56 |
| S4 | 1.35 | 1.36 | 0.46 | 1.75 | 0.54 | 1.99 |
| S5 | 3.51 | 2.65 | 3.34 | 2.77 | 3.71 | 3.06 |
| S6 | 2.84 | 2.19 | 3.36 | 2.11 | 3.05 | 2.22 |
| S7 | 2.97 | 2.58 | 2.32 | 1.70 | 2.47 | 1.57 |
| S8 | 3.09 | 2.69 | 2.01 | 2.41 | 1.95 | 1.95 |
| S9 | 1.34 | 2.25 | 0.53 | 2.29 | 0.56 | 2.70 |
| S10 | 1.16 | 1.24 | 1.04 | 1.38 | 1.37 | 1.94 |
| Mean | 2.20 | 2.17 | 1.78 | 2.10 | 1.89 | 2.20 |
| SD | 0.84 | 0.49 | 1.01 | 0.42 | 1.04 | 0.54 |
Average standard deviation of maximum (MAX, wrist radial deviation) and minimum (MIN, wrist ulnar deviation) for each subject. IMU, OMC1 and OMC2 depict Sensor, Vicon skin markers and Vicon senor markers acquisition systems.
| SD Max IMU (°) | SD Min IMU (°) | SD Max OMC1 (°) | SD Min OMC1 (°) | SD Max OMC2 (°) | SD Min OMC2 (°) | |
|---|---|---|---|---|---|---|
| S1 | 0.80 | 0.99 | 0.38 | 0.66 | 0.34 | 0.85 |
| S2 | 0.92 | 1.00 | 0.54 | 0.68 | 0.43 | 0.77 |
| S3 | 0.55 | 0.59 | 0.72 | 0.73 | 0.51 | 0.88 |
| S4 | 0.91 | 1.16 | 0.63 | 0.49 | 0.65 | 0.54 |
| S5 | 0.83 | 1.56 | 0.85 | 1.21 | 0.77 | 1.44 |
| S6 | 1.39 | 1.92 | 0.79 | 1.81 | 0.72 | 1.77 |
| S7 | 1.65 | 1.69 | 1.40 | 1.21 | 1.28 | 1.26 |
| S8 | 1.15 | 3.37 | 0.97 | 3.69 | 0.78 | 3.60 |
| S9 | 0.80 | 1.11 | 0.33 | 0.54 | 0.34 | 0.69 |
| S10 | 1.67 | 0.83 | 0.96 | 0.37 | 0.50 | 0.44 |
| Mean | 1.07 | 1.42 | 0.76 | 1.14 | 0.63 | 1.22 |
| SD | 0.36 | 0.75 | 0.30 | 0.95 | 0.27 | 0.89 |
Figure 3Bland-Altman plot comparing IMU ROM and OMC1 ROM. The x-axis depicts the mean ROM of both systems, while y-axis shows their difference. A confidence interval of ±1.96 * SD is plotted in dashed lines. Results of measurement day 1 are depicted in blue and measurement day 2 in red. Each subject is represented by a different marker (a) Results of the flexion-extension trials. (b) Results of the radial-ulnar deviation trials.
Figure 4Bland-Altman plot comparing IMU ROM and OMC2 ROM. The x-axis depicts the mean ROM of both systems, while y-axis shows their difference. A confidence interval of ±1.96 * SD is plotted in dashed lines. Results of measurement day 1 are depicted in blue and measurement day 2 in red. Each subject is represented by a different marker (a) Results of the flexion-extension trials. (b) Results of the radial-ulnar deviation trials.
Figure 5Bland-Altman plot comparing OMC1 ROM and OMC2 ROM. The x-axis depicts the mean ROM of both systems, while y-axis shows their difference. A confidence interval of ±1.96 * SD is plotted in dashed lines. Results of measurement day 1 are depicted as blue and measurement day 2 as red. Each subject is represented by a different marker (a) Results of the flexion-extension trials. (b) Results of the radial-ulnar deviation trials.
Average standard deviation of the maxima (MAX, corresponding to wrist flexion or radial deviation) and minima (MIN, corresponding to wrist extension or ulnar deviation) within each trial.
| Task | SD Max IMU (°) | SD Min IMU (°) | SD Max OMC1 (°) | SD Min OMC1 (°) | SD Max OMC2 (°) | SD Min OMC2 (°) |
|---|---|---|---|---|---|---|
| Flexion-Extension (mean) | 2.20 | 2.17 | 1.78 | 2.10 | 1.89 | 2.20 |
| Flexion-Extension (SD) | 0.84 | 0.49 | 1.01 | 0.42 | 1.04 | 0.54 |
| Radial-Ulnar (Mean) | 1.07 | 1.42 | 0.76 | 1.14 | 0.63 | 1.22 |
| Radial-Ulnar (SD) | 0.36 | 0.75 | 0.30 | 0.95 | 0.27 | 0.89 |
Average standard deviation of the maxima (MAX, corresponding to wrist flexion or radial deviation) and minima (MIN, corresponding to wrist extension or ulnar deviation) within each session.
| Task | SD Max IMU (°) | SD Min IMU (°) | SD Max OMC1 (°) | SD Min OMC1 (°) | SD Max OMC2 (°) | SD Min OMC2 (°) |
|---|---|---|---|---|---|---|
| Flexion-Extension (Mean) | 2.69 | 2.42 | 2.16 | 1.79 | 2.47 | 2.12 |
| Flexion-Extension (SD) | 1.55 | 1.50 | 1.49 | 1.31 | 1.83 | 1.35 |
| Radial-Ulnar (Mean) | 0.77 | 1.05 | 0.58 | 1.01 | 0.49 | 1.07 |
| Radial-Ulnar (SD) | 0.47 | 0.60 | 0.35 | 0.81 | 0.23 | 0.75 |
Test-Retest Difference of the ROM within each system (IMU, OMC1 and OMC2) for flexion-extension and radial-ulnar deviation.
| Flexion-Extension | IMU (°) | OMC1 (°) | OMC2 (°) | Radial-Ulnar Deviation | IMU (°) | OMC1 (°) | OMC2 (°) |
|---|---|---|---|---|---|---|---|
| S1 | 4.7 | 0.6 | −6.0 | S1 | −2.2 | 1.4 | 1.5 |
| S2 | 9.2 | −0.3 | −1.3 | S2 | 0.1 | −0.8 | 2.1 |
| S3 | 9.2 | 9.5 | 4.4 | S3 | 9.2 | 6.4 | 7.0 |
| S4 | 0.6 | 1.9 | 0.9 | S4 | 0.8 | −2.5 | 0.0 |
| S5 | 20.2 | 24.4 | 22.6 | S5 | 3.7 | 1.2 | 4.4 |
| S6 | −15.2 | −16.8 | −15.2 | S6 | −10.4 | −11.7 | −11.3 |
| S7 | −15.0 | −9.8 | −14.2 | S7 | −1.2 | −6.3 | −4.5 |
| S8 | 6.9 | 1.7 | −11.8 | S8 | 0.1 | 0.7 | 4.5 |
| S9 | 1.8 | −3.4 | −1.9 | S9 | 1.5 | 0.4 | 2.5 |
| S10 | 10.9 | 9.0 | 13.9 | S10 | −1.2 | −0.5 | −0.4 |
| Mean | 3.3 | 1.7 | −0.9 | Mean | 0.0 | −1.2 | 0.6 |
| SD | 10.6 | 10.7 | 11.5 | SD | 4.7 | 4.6 | 5.0 |
| MAD | 9.4 | 7.7 | 9.2 | MAD | 3.0 | 3.2 | 3.8 |