Literature DB >> 31799344

Data on hydroxychloroquine interference with urine laboratory testing.

Jennie M Kingery1, Joshua B Radke2, Jon Maakestad1, Matthew D Krasowski1.   

Abstract

Hydroxychloroquine is a medication used to treat rheumatoid arthritis, systemic lupus erythematosus, and other autoimmune disorders. Previous studies have shown that hydroxychloroquine and the structurally related drug chloroquine have the potential to interfere with some common urine chemistry tests, especially at high concentrations. In the related research article, we observed suspected interference with urine drug of abuse testing in a patient who ingested approximately 12 g of hydroxychloroquine in an acute overdose, with urine hydroxychloroquine concentrations exceeding 500 mg/L. This case prompted a more detailed investigation of the effects of hydroxychloroquine spiked into pooled de-identified urine specimens from a hospital clinical laboratory. The data in this article provides the raw data for 24 urine assays that were investigated. The analyzed data is provided in the tables included in this article. The dataset reported is related to the research article entitled "Diagnostic Pitfalls and Laboratory Test Interference After Hydroxychloroquine Intoxication: A Case Report" [1].
© 2019 The Author(s).

Entities:  

Keywords:  Absorbance error; Assay interference; Clinical chemistry tests; Drug of abuse testing; Hydroxychloroquine; Photometry

Year:  2019        PMID: 31799344      PMCID: PMC6881646          DOI: 10.1016/j.dib.2019.104781

Source DB:  PubMed          Journal:  Data Brief        ISSN: 2352-3409


Specifications Table The data provided is of value as there is currently only limited published data demonstrating interfering effects of hydroxychloroquine on urine laboratory assays. Other researchers or personnel in clinical laboratories might find this data useful as a reference for comparison. Our data set would serve as a starting point for researchers interested in future investigations studying the effects of hydroxychloroquine on urine chemistry assays marketed by vendors. The data is of value as previous studies have not examined a wide range of urine laboratory assays commonly performed in clinical laboratories. The data provide information for 24 urine assays tested up to 1000 mg/L hydroxychloroquine, a concentration achievable in large overdose.

Data

We investigated the effect on urine assays of hydroxychloroquine at concentration up to 1000 mg/L spiked into pools of de-identified urine specimens from the university medical center central clinical laboratory. This followed from observation that a patient with a large overdose of hydroxychloroquine showed suspected interference for some urine laboratory tests obtained for clinical care [1]. There is limited published data that hydroxychloroquine can interfere with urine protein dipstick methods [2,3] and with some urine drug screening tests [4]. A detailed review of package inserts for drug of abuse and therapeutic drug monitoring assays did not find hydroxychloroquine reported as an interferent in any of the assays [5,6]. The 24 specific urine assays analyzed are as follows: amphetamines drug screen, amylase, benzodiazepines drug screen, buprenorphine drug screen, calcium, chloride, cocaine metabolite screen, cotinine screen, creatinine, glucose, human chorionic gonadotropin (hCG), magnesium, microalbumin, myoglobin, NGAL, opiates drug screen, oxycodone drug screen, pH, phosphorus, potassium, protein, sodium, tetrahydrocannabinol (THC) drugs screen, and urea nitrogen. Technical details on these 24 urine assays are available elsewhere [1]. The assays were initially screened in triplicate for a single de-identified pooled urine sample. Thirteen of the assays showed no absorbance or other alarms/errors and also did not have any results that differed by more than 15% from the control without hydroxychloroquine. The raw data for these 13 assays is in Table 1. Raw data for the remaining 11 assays is in Table 2. These were all tested in triplicate for a total of 4 separate pooled urine samples.
Table 1

Raw data for 13 urine assays on a single pooled urine samples.

AssayUnitsHydroxychoroquine concentration
0 mg/L1 mg/L10 mg/L100 mg/L500 mg/L1000 mg/L
AmylaseU/L121121123122121122
Calciummg/dL10.410.410.610.89.910.0
Chloridemmol/L69.070.768.070.372.776.7
Creatininemg/dL76.979.379.377.882.380.7
Glucosemg/dL6.05.55.45.36.06.0
hCGmIU/mL1.01.11.11.11.01.1
Magnesiummg/dL4.84.74.74.74.34.2
NGALng/mL12.011.011.512.011.011.5
pHpH units76.86.96.96.96.9
Phosphorusmg/dL20.421.322.021.619.620.1
Potassiummmol/L353534353939
Proteinmg/dL555757585449
Sodiummmol/L606261626566
Table 2

Raw data for 13 urine assays on four separate pooled urine samples.

AssayUnitsSample #Hydroxychloroquine concentration
0 mg/L1 mg/L10 mg/L100 mg/L500 mg/L1000 mg/L

Amphetamines Drug ScreenRelative1−268−286−276−281−271−250
absorbance units2−302−323−309−314−298−291
(positive: >0)3−273−283−284−279−281−264
4−299−299−295−298−283−265
Benzodiazepines Drug ScreenRelative1−187−178−176−174−171−164
absorbance units2−263−204−209−201−185−58
(positive: >0)3−210−181−183−176−176−172
4102108106105105107
Buprenorphine Drug ScreenRelative1−339−441−441−472−534Absorbance error
absorbance units2−283−421−410−429−472Absorbance error
(positive: >0)3−283−393−398−420−418Absorbance error
4−258−389−395−405−451Absorbance error
Cocaine Drug ScreenRelative1−548−538−554−557−565−540
absorbance units2−575−587−570−581−578−558
(positive: >0)3−553−543−546−546−537−536
4−558−560−568−570−551−552
Cotinine ScreenRelative1>2000>2000>2000>2000>2000Absorbance error
absorbance units21294786810784705Absorbance error
(positive: >0)3−250−194−182−202−268Absorbance error
4229398397348119Absorbance error
Microalbuminmcg/mg creatinine1293273273281315329
2155105105112144189
3242023242630
4215225224231249274
Myoglobinng/mL1<21<21<21<21<21<21
21008899900876752599
3<21<21<21<21<21<21
4<21<21<21<21<21<21
Opiates Drug ScreenRelative1881031011019689
absorbance units2−520−497−496−496−469−436
(positive: >0)3−483−477−455−475−458−459
4−479−446−453−447−426−392
Oxycodone Drug ScreenRelative1−158−159−160−158−158Absorbance error
absorbance units22831323127Absorbance error
(positive: >0)3−154−159−159−157−156Absorbance error
4−155−155−156−155−153Absorbance error
Urea nitrogenmg/dL1859862884877866863
2684696692694673647
3951927930926930923
4541529531519519518
THC Drug ScreenRelative1−178−165−162−163−155−126
absorbance units2−221−191−186−180−167−152
(positive: >0)3−225−206−206−207−195−201
4−225−210−211−212−194−180
Raw data for 13 urine assays on a single pooled urine samples. Raw data for 13 urine assays on four separate pooled urine samples. Fig. 1 shows data for 4 of the assays (amphetamine screen, benzodiazepine screen, buprenorphine screen, and cocaine screen). An absorbance alarm was evident for all 4 samples containing 1000 mg/L hydroxychloroquine for the buprenorphine screen (Fig. 1B). Fig. 2 shows data for another 4 of the assays (cotinine screen, microalbumin, myoglobin, and opiates screen). Positive bias and absorbance errors were evident for the cotinine assay for all 4 specimens (Fig. 2A). Biphasic effects of hydroxychloroquine were evident for the microalbumin assay, especially for samples 1, 2, and 4 (Fig. 2B). A negative bias was evident for the myoglobin assay for sample 2 (Fig. 2C). The remaining urine samples had <21 ng/mL myoglobin and did not show any evident effect of hydroxychloroquine, although a negative bias would not be detectable in these 3 samples with the myoglobin concentrations below the lower limit of quantitation. Fig. 3 shows data for the remaining 3 assays analyzed in detail. Absorbance errors were evident for the oxycodone for all 4 samples spiked with 1000 mg/L (Fig. 3A). A positive bias from hydroxychloroquine was evident for all 4 samples analyzed for tetrahydrocannabinol (THC; Fig. 3B), although all 4 of these samples would still be negative in this qualitative assay.
Fig. 1

Analysis of hydroxychloroquine effects on urine assays for (A) amphetamine screen, (B) benzodiazepine screen, (C) buprenorphine screen, and (D) cocaine screen. Four separate de-identified pooled urine samples were tested in triplicate at the indicated hydroxychloroquine concentrations.

Fig. 2

Analysis of hydroxychloroquine effects on urine assays for (A) cotinine screen, (B) microalbumin screen, (C) myoglobin assay, and (D) opiates screen. Four separate de-identified pooled urine samples were tested in triplicate at the indicated hydroxychloroquine concentrations.

Fig. 3

Analysis of hydroxychloroquine effects on urine assays for (A) oxycodone screen, (B) tetahydrocannabinol (THC) screen, and (C) urea nitrogen. Four separate de-identified pooled urine samples were tested in triplicate at the indicated hydroxychloroquine concentrations.

Analysis of hydroxychloroquine effects on urine assays for (A) amphetamine screen, (B) benzodiazepine screen, (C) buprenorphine screen, and (D) cocaine screen. Four separate de-identified pooled urine samples were tested in triplicate at the indicated hydroxychloroquine concentrations. Analysis of hydroxychloroquine effects on urine assays for (A) cotinine screen, (B) microalbumin screen, (C) myoglobin assay, and (D) opiates screen. Four separate de-identified pooled urine samples were tested in triplicate at the indicated hydroxychloroquine concentrations. Analysis of hydroxychloroquine effects on urine assays for (A) oxycodone screen, (B) tetahydrocannabinol (THC) screen, and (C) urea nitrogen. Four separate de-identified pooled urine samples were tested in triplicate at the indicated hydroxychloroquine concentrations.

Experimental design, materials, and methods

All analyses were performed on Roche Diagnostics cobas 8000 analyzers (c501, c502, c602, and c701). The complete list of assays with vendor name, methodology, and assay version are summarized elsewhere [1]. All assays were run in accordance with package insert instructions. A total of 4 pooled urine specimens were prepared using de-identified specimens from the clinical laboratory. All 24 assays were tested in triplicate for 1 of the urine pools. As described above, 11 of the assays were then tested and triplicate for the remaining 3 urine pools. The raw data consists of concentration or absorbance units for the specimens. It should be noted that the urine total protein method used for the data in this report uses a biuret complex method using divalent copper in alikaline solution [7]. This is in contrast to previous studies using tetrabromophenol blue [3] and pyrogallol red-molybdate methods [2] that showed interference by hydroxychloroquine.

Specifications Table

SubjectMedicine and Dentistry
Specific subject areaPathology and Medical Technology
Type of dataTableFigure
How data were acquiredUrine chemistry assays run on Roche Diagnostics cobas c501, c502, e602, and c701 clinical chemistry analyzers
Data formatRaw and Analyzed
Parameters for data collectionPools of de-identified urine specimens from the hospital clinical laboratory were analyzed without hydroxychloroquine and with hydroxychloroquine (Sigma-Aldrich, St. Louis, MO) spiked at concentrations of 1, 10, 100, 500, and 1000 mg/L.
Description of data collectionData for 13 of the assays (amylase, calcium, chloride, creatinine, glucose, human chorionic gonadotropin, magnesium, NGAL, pH, phosphorus, potassium, protein, and sodium) were analyzed by analysis of a single urine pool. Data for 11 of the assays (amphetamines screen, benzodiazepines screen, buprenorphine screen, cocaine metabolite screen, cotinine screen, microalbumin, myoglobin, opiates screen, oxycodone screen, urea nitrogen, and tetrahydrocannabinol screen) were analyzed by analysis of four separate urine pools.
Data source locationIowa City, Iowa, United States of America
Data accessibilityWith the article
Related research articleAuthor's name Joshua B. Radke, Jennie M. Kingery, Jon Maakestad, Matthew D. KrasowskiTitle Diagnostic Pitfalls and Laboratory Test Interference After Hydroxychloroquine Intoxication: A Case ReportJournal Toxicology Reports [1]DOI: 10.1016/j.toxrep.2019.10.006
Value of the Data

The data provided is of value as there is currently only limited published data demonstrating interfering effects of hydroxychloroquine on urine laboratory assays.

Other researchers or personnel in clinical laboratories might find this data useful as a reference for comparison.

Our data set would serve as a starting point for researchers interested in future investigations studying the effects of hydroxychloroquine on urine chemistry assays marketed by vendors.

The data is of value as previous studies have not examined a wide range of urine laboratory assays commonly performed in clinical laboratories.

The data provide information for 24 urine assays tested up to 1000 mg/L hydroxychloroquine, a concentration achievable in large overdose.

  7 in total

1.  Analytical interference of quinolone antibiotics and quinine derived drugs on urinary protein determined by reagent strips and the pyrogallol red-molybdate protein assay.

Authors:  Adriana Scotti da Silva; Miriam Falkenberg
Journal:  Clin Biochem       Date:  2011-05-24       Impact factor: 3.281

2.  An accurate and rapid method for the determination of proteins in small amounts of blood serum and plasma.

Authors:  T E WEICHSELBAUM
Journal:  Am J Clin Pathol       Date:  1946-03       Impact factor: 2.493

3.  Method development and validation for rapid quantification of hydroxychloroquine in human blood using liquid chromatography-tandem mass spectrometry.

Authors:  Ling-Zhi Wang; Rina Yue-Ling Ong; Tan-Min Chin; Win-Lwin Thuya; Seow-Ching Wan; Andrea Li-Ann Wong; Sui-Yung Chan; Paul C Ho; Boon-Cher Goh
Journal:  J Pharm Biomed Anal       Date:  2011-12-08       Impact factor: 3.935

4.  A new highly specific buprenorphine immunoassay for monitoring buprenorphine compliance and abuse.

Authors:  Stacy E F Melanson; Marion L Snyder; Petr Jarolim; James G Flood
Journal:  J Anal Toxicol       Date:  2012-04       Impact factor: 3.367

5.  Chemoinformatic methods for predicting interference in drug of abuse/toxicology immunoassays.

Authors:  Matthew D Krasowski; Mohamed G Siam; Manisha Iyer; Anthony F Pizon; Spiros Giannoutsos; Sean Ekins
Journal:  Clin Chem       Date:  2009-04-02       Impact factor: 8.327

6.  Molecular similarity methods for predicting cross-reactivity with therapeutic drug monitoring immunoassays.

Authors:  Matthew D Krasowski; Mohamed G Siam; Manisha Iyer; Sean Ekins
Journal:  Ther Drug Monit       Date:  2009-06       Impact factor: 3.681

7.  Diagnostic pitfalls and laboratory test interference after hydroxychloroquine intoxication: A case report.

Authors:  Joshua B Radke; Jennie M Kingery; Jon Maakestad; Matthew D Krasowski
Journal:  Toxicol Rep       Date:  2019-10-07
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.