The aim of this study was to evaluate the accuracy of two intraoral scanners used in the dental office. A molar fixed in a typodont was prepared for a ceramic onlay. The preparation was scanned using a high-resolution scanner (reference scanner) and saved as stereolithography (STL) format. The prepared resin molar was scanned again using the intraoral scanners, and all the scans were saved as well in STL format. All STL files were compared using metrology software (Geomagic Control X). Overlapping the meshes allowed the assessment of the scans in terms of trueness and precision. Based on the results of this study, the differences of trueness and precision between the intraoral scanners were minimal.Entities:
Mesh:
Substances:
Year: 2019 PMID: 31741697 PMCID: PMC6854270 DOI: 10.1155/2019/1289570
Source DB: PubMed Journal: Scanning ISSN: 0161-0457 Impact factor: 1.932
Figure 1Reference model.
Figure 2Intraoral scans with PlanScan (b) and Omnicam (a).
Figure 3Alignment process of the meshes.
Figure 4Color map of the deviation on the interest area.
Trueness values (μm) of the intraoral scanners (p value = 0.2).
| M1 | M2 | M3 | M4 | M5 | Mean ± SD | |
|---|---|---|---|---|---|---|
| Planmeca PlanScan | 43 | 53 | 46 | 53 | 48 | 48.6 ± 4.39 |
| CEREC Omnicam | 54 | 53 | 46 | 50 | 62 | 53 ± 5.91 |
Precision values (μm) of the intraoral scanners (p value = 0.08).
| M1 | M2 | M3 | M4 | M5 | M6 | M7 | Mean ± SD | |
|---|---|---|---|---|---|---|---|---|
| Planmeca PlanScan | 28 | 25 | 21 | 28 | 22 | 27 | 23 | 24.86 ± 2.91 |
| CEREC Omnicam | 31 | 31 | 55 | 21 | 39 | 28 | 44 | 35.57 ± 11.34 |
Figure 5Color map of the PlanScan trueness deviation values around the interest area.
Figure 6Color map of the Omnicam trueness deviation values around the interest area.
Figure 7Color map of the PlanScan precision deviation values around the interest area.
Figure 8Color map of the Omnicam precision deviation values around the interest area.