| Literature DB >> 31701215 |
Nina Huppertz1, Gregory Y H Lip2,3, Deirdre A Lane4,5.
Abstract
AIMS: Undiagnosed atrial fibrillation (AF) accounts for 6% of all strokes, therefore early detection and treatment of the arrhythmia are paramount. Previous research has illustrated that the Microlife WatchBPO3 AFIB, an automated blood pressure (BP) monitor with an inbuilt AF algorithm, accurately detects permanent AF. Currently, limited data exist on whether the modified BP monitor is able to detect paroxysmal AF (PAF). Therefore, this study aims to assess the accuracy of the Microlife WatchBPO3 AFIB monitor to detect PAF against a pacemaker reference standard over a 24-h period. METHODS ANDEntities:
Keywords: Atrial fibrillation; Blood pressure; Detection; Sensitivity; Specificity
Mesh:
Year: 2019 PMID: 31701215 PMCID: PMC7308245 DOI: 10.1007/s00392-019-01567-y
Source DB: PubMed Journal: Clin Res Cardiol ISSN: 1861-0684 Impact factor: 5.460
Clinical characteristics, medications, and pacemaker details of included patients
| Variable | Included patients ( |
|---|---|
| Current age (years) | 71.6 ± 10.9 |
| Age at pacemaker implant (years) | 67.9 ± 11.0 |
| Gender ( | |
| Male | 28 (58.3) |
| Female | 20 (41.7) |
| Height (cm) | 167.9 ± 9.7 |
| Weight (kg) | 76.1 [21.0] |
| Body mass index (kg/m2) | 27.0 [8.4] |
| Ethnicity ( | |
| White | 34 (70.8) |
| Asian (Indian, Pakistani, Bangladeshi) | 10 (20.8) |
| Black/Caribbean | 2 (4.2) |
| Mixed | 2 (4.2) |
| Congestive HF ( | 1 (2.1) |
| Hypertension ( | 32 (66.7) |
| Diabetes mellitus ( | |
| Type I | 0 (0) |
| Type II | 14 (29.2) |
| Previous stroke ( | |
| Ischaemic | 3 (6.3) |
| Haemorrhagic | 2 (4.2) |
| Unspecified | 3 (6.3) |
| Vascular disease ( | |
| Prior myocardial infarction | 5 (10.4) |
| Coronary artery disease | 12 (25.0) |
| Peripheral artery disease | 0 (0) |
| Hyperlipidaemia ( | 19 (39.6) |
| Other cardiovascular disease ( | |
| Prior cardiac surgery | 2 (4.2) |
| Congenital | 0 (0) |
| Valvular | 1 (2.1) |
| Pericarditis | 0 (0) |
| Cardiomyopathy | 0 (0) |
| Medications ( | |
| Anticoagulants | 29 (60.4) |
| Antiplatelets | 16 (33.3) |
| ACE-inhibitors | 15 (31.3) |
| Angiotensin receptor blocker | 8 (16.7) |
| Beta-blockers | 23 (47.9) |
| Calcium-channel blockers | 12 (25.0) |
| Digitalis | 2 (4.2) |
| Diuretics | 17 (35.4) |
| Statins | 31 (64.6) |
| Vasodilators | 8 (16.7) |
| Device model ( | |
| Medtronic Adapta | 8 (16.7) |
| Medtronic Advisa | 14 (29.2) |
| Medtronic Ensura | 6 (12.5) |
| Medtronic Sensia | 8 (16.7) |
| St. Jude Accent | 1 (2.1) |
| St. Jude Assurity | 7 (14.6) |
| St. Jude Azure | 2 (4.2) |
| Sorin Reply | 2 (4.2) |
| Mode ( | |
| DDD | 3 (6.3) |
| DDDR | 16 (33.3) |
| DDI | 1 (2.1) |
| AAI-DDD | 10 (20.8) |
| AAIR-DDDR | 17 (35.4) |
| AAIR | 1 (2.1) |
| Indication for implant ( | |
| Sinus arrest | 19 (39.6) |
| Sinus bradycardia | 18 (37.5) |
| Brady-tachy syndrome | 6 (12.5) |
| Chronotropic incompetence | 1 (2.1) |
| Unspecified SSS | 4 (8.3) |
Mean ± standard deviation given for normally distributed variables, median [interquartile range] given for not normally distributed variables. Number (%) given for categorical variables
Fig. 1Distribution of true paroxysmal AF episodes with detection ability of Microlife WatchBPO3 AFIB
Modified BP monitor diagnoses of paroxysmal AF vs. pacemaker diagnoses of paroxysmal AF
| Positive AF pacemaker diagnosis | Negative AF pacemaker diagnosis | |
|---|---|---|
| Positive AF modified BP monitor diagnosis | 19 | 300 |
| Negative AF modified BP monitor diagnosis | 6 | 1270 |
Sensitivity, specificity, PPV, and NPV with 95% CI of the Microlife WatchBPO3 AFIB to detect paroxysmal AF in various settings
| Sensitivity (%) | Specificity (%) | PPV (%) | NPV (%) | |
|---|---|---|---|---|
| Overall (24-h) AF detection | 76.0 (74–78) | 80.8 (79–83) | 5.9 (5–7) | 99.5 (99–100) |
| Day-time AF detection | 68.4 (66–71) | 79.3 (77–81) | 5.1 (4–6) | 99.4 (99–100) |
| Night-time AF detection | 100 (99a–100) | 84.9 (83–87) | 8.8 (7–10) | 100 (99a–100) |
| AF detection in < 80% successful BP readings | 76.0 (59–96) | 76.9 (70–79) | 15.4 (8–21) | 98.3 (97–100) |
| AF detection in > 80% successful BP readings | 100 (99a-100) | 83.1 (80–86) | 0.5 (0–4) | 100 (99a–100) |
aOne-sided 97.5% confidence interval