| Literature DB >> 31684057 |
Sergio Curto1, Bassim Aklan2, Tim Mulder3, Oliver Mils4, Manfred Schmidt5, Ulf Lamprecht6, Michael Peller7, Ruediger Wessalowski8, Lars H Lindner9, Rainer Fietkau10, Daniel Zips11, Gennaro G Bellizzi12, Netteke van Holthe13, Martine Franckena14, Margarethus M Paulides15,16, Gerard C van Rhoon17.
Abstract
Clinical outcome of hyperthermia depends on the achieved target temperature, therefore target conformal heating is essential. Currently, invasive temperature probe measurements are the gold standard for temperature monitoring, however, they only provide limited sparse data. In contrast, magnetic resonance thermometry (MRT) provides unique capabilities to non-invasively measure the 3D-temperature. This study investigates MRT accuracy for MR-hyperthermia hybrid systems located at five European institutions while heating a centric or eccentric target in anthropomorphic phantoms with pelvic and spine structures. Scatter plots, root mean square error (RMSE) and Bland-Altman analysis were used to quantify accuracy of MRT compared to high resistance thermistor probe measurements. For all institutions, a linear relation between MRT and thermistor probes measurements was found with R2 (mean ± standard deviation) of 0.97 ± 0.03 and 0.97 ± 0.02, respectively for centric and eccentric heating targets. The RMSE was found to be 0.52 ± 0.31 °C and 0.30 ± 0.20 °C, respectively. The Bland-Altman evaluation showed a mean difference of 0.46 ± 0.20 °C and 0.13 ± 0.08 °C, respectively. This first multi-institutional evaluation of MR-hyperthermia hybrid systems indicates comparable device performance and good agreement between MRT and thermistor probes measurements. This forms the basis to standardize treatments in multi-institution studies of MR-guided hyperthermia and to elucidate thermal dose-effect relations.Entities:
Keywords: anthropomorphic phantom; magnetic resonance imaging-guided hyperthermia; magnetic resonance thermometry; quality assurance; radiofrequency hyperthermia; thermistor probe
Year: 2019 PMID: 31684057 PMCID: PMC6896203 DOI: 10.3390/cancers11111709
Source DB: PubMed Journal: Cancers (Basel) ISSN: 2072-6694 Impact factor: 6.639
Figure 1Pyrexar BSD2000-3D-MR Sigma Eye and Pyrexar BSD2000-3D-MR Universal hyperthermia applicators operating inside GE, Siemens and Philips MR-scanners (A), newly developed anthropomorphic phantom and close view of the catheter positioning (B), 3D schematic of the phantom bone structure and catheter locations (C), schematic cross-section of the positioning of the phantom inside the MR-scanner and monitoring of the central axial slide (D), and sequence of the performed MR scans (E).
MR sequences parameters as used during the clinical routine and during the performed experiments at the five institutions, for High Resolution scans and Magnetic Resonance Thermometry (MRT) scans.
| Institution | Scan Type | TR (ms) | TE1 (ms) | TE2 (ms) | FA (deg) | Acquisition Matrix | Reconstruction Matrix | Scan Time (sec) |
|---|---|---|---|---|---|---|---|---|
| 1 | High Resolution | 120 | 4.8 | 9.60 | 70 | 256/256 | 256/256 | 136 |
| MRT | 620 | 4.8 | 19.1 | 40 | 128/128 | 256/256 | 83 | |
| 2 | High Resolution | 120 | 4.76 | 9.53 | 70 | 256/256 | 256/256 | 124 |
| MRT | 600 | 4.76 | 19.10 | 50 | 128/128 | 128/128 | 78 | |
| 3 | High Resolution | 120 | 4.60 | 9.21 | 70 | 252/250 | 256/256 | 151 |
| MRT | 600 | 4.60 | 18.42 | 50 | 128/128 | 256/256 | 79 |
Figure 2Temporal evolution of temperature increase determined by high resistance thermistor probes and by computed MRT for centric heating target (yellow circle). The position of probes and ROIs in the phantom are indicated by numbers and coloured squares in the MR image showing the cross-section of the phantom and the surrounding water bolus. Measurements were performed at five different institutions.
Temperature increase at 600 sec for centric heating target measured with thermistor probes and computed Magnetic Resonance Thermometry (MRT).
| Institution | Temperature Measurement | Mean Temperature Increase [ΔT (°C)] at Locations 1 to 4 | Mean Temperature Increase± SD (°C) | |||
|---|---|---|---|---|---|---|
| ΔT Location 1 | ΔT Location 2 | ΔT Location 3 | ΔT Location 4 | |||
| 1 | Thermistor probe | 4.7 | 4.6 | 5.2 | 5.3 | 5.0 ± 0.3 |
| MRT | 3.8 | 3.5 | 4.3 | 4.3 | 4.0 ± 0.3 | |
| 2 | Thermistor probe | 6.8 | 4.7 | 5.6 | 5.6 | 5.7 ± 0.7 |
| MRT | 6.2 | 4.9 | 5.3 | 5.1 | 5.4 ± 0.5 | |
| 3 | Thermistor probe | 4.4 | 4.2 | 4.4 | 4.3 | 4.3 ± 0.1 |
| MRT | 3.6 | 3.2 | 3.7 | 3.5 | 3.5 ± 0.2 | |
| 4 | Thermistor probe | 4.1 | 4.1 | 3.9 | 4.0 | 4.0 ± 0.1 |
| MRT | 3.2 | 3.9 | 3.5 | 3.3 | 3.5 ± 0.3 | |
| 5 | Thermistor probe | 3.2 | 3.4 | 4.0 | 4.2 | 3.7 ± 0.4 |
| MRT | 2.3 | 2.4 | 3.0 | 3.3 | 2.7 ± 0.4 | |
Figure 3Scatter plots of MRT data as a function of high resistance thermistor probes measurements for centric heating target, measurements performed at the five institutions. The estimated linear regression equation (solid black line), the 95% confidence interval upper and lower limits (dashed blue lines) and the identity line of both methods (dot-dashed red line) are shown for all institutions. R2 coefficients are reported.
Figure 4Bland-Altman quantitative analysis plots for centric heating target. The plots show the differences in the temperature change obtained by high resistance thermistor probes and MRT against the averages of the two measurements. The dotted blue line shows the mean difference, while the dashed black lines show the mean difference ± 2 × standard deviations. RMSE coefficients are reported.
Figure 5Temporal evolution of temperature increase determined by high resistance thermistor probes and by computed MRT for eccentric heating target (yellow circle). The position of probes and ROIs in the phantom are indicated by numbers and coloured squares in the MR image showing the cross-section of the phantom and the surrounding water bolus. Measurements were performed at five different institutions.
Temperature increases at 600 sec for eccentric heating target measured with thermistor probes and computed Magnetic Resonance Thermometry (MRT).
| Institution | Temp. Measurem. | Mean Temperature Increase | Mean Temp. Increase ± SD (°C) | d(ΔT Location 4- ΔT Location 1) (°C) | |||
|---|---|---|---|---|---|---|---|
| ΔT Location 1 | ΔT Location 2 | ΔT Location 3 | ΔT Location 4 | ||||
| 1 | Thermistor probe | 2.6 | 3.2 | 4.3 | 5.9 | 4.0 ± 1.2 | 3.3 |
| MRT | 2.0 | 2.7 | 4.1 | 5.3 | 3.5 ± 1.3 | 3.3 | |
| 2 | Thermistor probe | 4.5 | 4.7 | 6.1 | 7.2 | 5.6 ± 1.1 | 2.7 |
| MRT | 3.8 | 4.3 | 6.0 | 7.0 | 5.3 ± 1.3 | 3.2 | |
| 3 | Thermistor probe | 2.5 | 3.0 | 3.7 | 4.6 | 3.5 ± 0.8 | 2.1 |
| MRT | 2.3 | 2.5 | 3.9 | 4.4 | 3.3 ± 0.9 | 2.0 | |
| 4 | Thermistor probe | 2.7 | 2.9 | 3.0 | 3.6 | 3.0 ± 0.3 | 0.9 |
| MRT | 2.6 | 2.8 | 3.2 | 3.7 | 3.0 ± 0.4 | 1.1 | |
| 5 | Thermistor probe | 1.3 | 1.7 | 2.8 | 4.3 | 2.5 ± 1.2 | 3.0 |
| MRT | 0.6 | 0.9 | 3.2 | 4.6 | 2.3 ± 1.6 | 3.9 | |
Figure 6Scatter plots of MRT data as a function of high resistance thermistor probes measurements for eccentric heating target, measurements performed at the five institutions. The estimated linear regression equation (solid black line), the 95% confidence interval upper and lower limits (dashed blue lines) and the identity line of both methods (dot-dashed red line) are shown for all institutions. R2 coefficients are reported.
Figure 7Bland-Altman quantitative analysis plots for eccentric heating target. The plots show the differences in the temperature change obtained by high resistance thermistor probes and MRT against the averages of the two measurements. The dotted blue line shows the mean difference, while the dashed black lines show the mean difference ± 2 × standard deviations. RMSE coefficients are reported.