Healthy knee KT1000 measurements of anterior tibial translation have significant variation.

Antonio Klasan1, Sven Edward Putnis2, Vikram Kandhari2, Takeshi Oshima2, Brett Andrew Fritsch2, David Anthony Parker2.   

Abstract

PURPOSE: The most commonly used arthrometer for measuring and reporting anterior tibial translation after anterior cruciate ligament reconstruction is the KT1000. Reports on its consistency and reproducibility vary in the literature, but it remains the "gold standard". The purpose of this study was to assess agreement of KT1000 measurements in a daily clinical setting.
METHODS: A retrospective analysis of anterior knee translation in the healthy knee of 770 patients over a 17-year time period was performed. In this cohort, a total of 24 investigators performed 1890 measurement sets at 89 Newtons (N), 134N and at maximum manual force (MMax) level. To assess the inter- and intra-observer agreement, the intraclass-correlation coefficient (ICC) was calculated. The "investigator effect" was a difference between two examiners in the same patient and the "device effect'' a difference within one examiner in the same patient. Minimally important difference (MID) was calculated as 0.5 of the standard deviation.
RESULTS: Thirteen investigators were female, performing 1099 measurements and 11 were male, performing 791 measurements. ICC ranged between 0.558 and 0.644. At the MMax level, male investigators had a higher mm reading than female investigators (p < 0.001). Increased experience did not correlate with a higher ICC. MID ranged between 0.85 and 1.65 mm.
CONCLUSION: This study investigated the KT1000 arthrometer in a clinical setting with a large number of investigators. This device delivers moderate agreement of results. Both the device and investigator effect are present. The MMax level has shown the lowest agreement and a dependency on the investigator gender. LEVEL OF EVIDENCE: Level III diagnostic study.

Entities:  

Keywords:  Anterior cruciate ligament; Anterior tibial translation; Arthrometer; Knee

Mesh:

Year:  2019        PMID: 31676998     DOI: 10.1007/s00167-019-05768-w

Source DB:  PubMed          Journal:  Knee Surg Sports Traumatol Arthrosc        ISSN: 0942-2056            Impact factor:   4.342


  4 in total

Review 1.  Interpretation of changes in health-related quality of life: the remarkable universality of half a standard deviation.

Authors:  Geoffrey R Norman; Jeff A Sloan; Kathleen W Wyrwich
Journal:  Med Care       Date:  2003-05       Impact factor: 2.983

2.  Anterior cruciate ligament laxity related to the menstrual cycle: an updated systematic review of the literature.

Authors:  Lesley Belanger; Dawn Burt; Julia Callaghan; Sheena Clifton; Brian J Gleberzon
Journal:  J Can Chiropr Assoc       Date:  2013-03

3.  Joint laxity and graft compliance increase during the first year following ACL reconstruction with short hamstring tendon grafts.

Authors:  Thomas Pouderoux; Bart Muller; Henri Robert
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2019-09-28       Impact factor: 4.342

4.  Degenerative medial meniscus posterior root tear and non-root tear do not show differences in joint survival and clinical outcome after partial meniscectomy.

Authors:  Oh-Jin Kwon; Seong-Il Bin; Jong-Min Kim; Bum-Sik Lee; Sang-Min Lee; Jun-Gu Park; Gi-Woon Yoon
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2019-10-31       Impact factor: 4.342

  4 in total
  1 in total

1.  The evaluation of Rolimeter, KLT, KiRA and KT-1000 arthrometer in healthy individuals shows acceptable intra-rater but poor inter-rater reliability in the measurement of anterior tibial knee translation.

Authors:  Armin Runer; Tommaso Roberti di Sarsina; Vasco Starke; Alessandra Iltchev; Gernot Felmet; Sepp Braun; Christian Fink; Robert Csapo
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2021-03-31       Impact factor: 4.342

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.