Armin Runer1, Tommaso Roberti di Sarsina2, Vasco Starke3, Alessandra Iltchev3, Gernot Felmet4, Sepp Braun5,6, Christian Fink5,6, Robert Csapo6.
Abstract
PURPOSE: To assess measurement equivalence, inter- and intra-rater reliability, standard error of measurements (SEM) and false positive measurements (FPM) of four different knee arthrometers (KLT,Karl Storz; KiRA, I + ; KT-1000 MEDmetric Corp; Rolimeter, Aircast) in healthy patients.Entities:
Keywords: ACL anterior tibial translation; Anterior cruciate ligament; Equivalence testing; Lachman Test; KLT; KT1000; KiRA; Knee arthrometer; Rolimeter; Side-to-side difference
Mesh:
Year: 2021 PMID: 33791824 PMCID: PMC8298217 DOI: 10.1007/s00167-021-06540-9
Source DB: PubMed Journal: Knee Surg Sports Traumatol Arthrosc ISSN: 0942-2056 Impact factor: 4.342
Fig. 1Test setup and patient positioning with all four arthrometers tested: KT-1000 (MEDmetric Corp, San Diego, Calif., USA), Rolimeter (Aircast, Europe), KLT (Karl Storz, Tuttlingen, Germany) and KiRA (I + Italy)
Inter-rater equivalence and reliability for Anterior Tibial Translation (ATT) and Side-to-Side Differences (SSD) measurements between advanced raters and beginners
| Anterior Tibial Translation (ATT) | Side-to-Side Difference (SSD) | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Examiner 1 | Examiner 2 | ICC (95% CI) | SEM | Examiner 1 | Examiner 2 | ICC (95% CI) | SEM | |||
| Rolimeter | 6.2 ± 1.5 | 6.5 ± 1.2 | 0.01 | 0.76* (0.37–0.92) | 0.68 | − 0.5 ± 1.1 | − 0.1 ± 0.8 | 0.03 | NR | 1.15 |
| KLT | 6.9 ± 1.9 | 6.5 ± 1.5 | n.s | 0.72* (0.31–0.91) | 0.90 | − 0.5 ± 1.6 | − 0.2 ± 0.8 | 0.09 | 0.25 (0–0.71) | 1.14 |
| KiRA | 11.1 ± 1.4 | 11.3 ± 2.2 | n.s | 0.18 (0–0.68) | 1.70 | − 2.1 ± 1.6 | − 0.2 ± 2.8 | n.s | 0.08 (0–0.55) | 2.15 |
| KT1000 | 8.5 ± 1.8 | 8.0 ± 2.2 | n.s | 0.70* (0.27–0.90) | 1.08 | − 1.3 ± 1.0 | 0.6 ± 1.3 | n.s | 0.16 (0–0.56) | 0.89 |
All values obtained from retest. Mean ATTs ± standard deviation displayed for the left knee. SSD’s obtained by subtracting the value of the left from the right leg. ¢, significance level reported for the equivalence test with ± 1 mm equivalence boundary
*Significant ICCs (p < 0.05) are flagged with *. Standard errors of measurement (SEM) are given in mm
NR Not reliable; ICC (95% CI) Intraclass correlation (95% Confidence Interval
Examiner 1–2 = advanced rater; Examiner 3–4 = beginner
Fig. 2Inter-rater equivalence testing for Anterior Tibial Translation (ATT) and Side-to-Side Difference (SSD) in healthy individuals for all arthrometers between advanced (a) and beginner (b) raters. Equivalence boundariesy’s are set to ± 1 mm. * Please note the different scaling (± 6 mm) for KiRA arthrometer
Intra-rater equivalence and reliability for Anterior Tibial Translation (ATT) and Side-to-Side Differences (SSD) measurements for each rater
| Anterior Tibial Translation (ATT) | ||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Rolimeter | KLT | KiRA | KT-1000 | |||||||||||||
| Mean ± SD | ICC (95% CI) | SEM | Mean ± SD | ICC (95% CI) | SEM | Mean ± SD | ICC (95% CI) | SEM | Mean ± SD | ICC (95% CI) | SEM | |||||
| Ex 1Test | 5.8 ± 1.4 | < 0.01 | 0.94 (0.25–0.99) | 0.23 | 6.9 ± 1.9 | < 0.01 | 0.87 (0.60–0.96) | 0.73 | 11.1 ± 1.9 | 0.02 | 0.63 (0.10–0.88) | 1.04 | 8.5 ± 2.0 | < 0.01 | 0.91 (0.73–0.97) | 0.57 |
| Ex 1 Retest | 6.2 ± 1.5 | 6.9 ± 1.9 | 11.1 ± 1.4 | 8.5 ± 1.8 | ||||||||||||
| Ex 2 Test | 6.3 ± 1.1 | < 0.01 | 0.92 (0.76–0.98) | 0.31 | 6.6 ± 1.3 | < 0.01 | 0.90 (0.70–0.97) | 0.45 | 11.5 ± 2.0 | 0.03 | 0.79 (0.41–0.93) | 1.00 | 8.2 ± 2.1 | < 0.01 | 0.94 (0.80–0.98) | 0.53 |
| Ex 2 Retest | 6.5 ± 1.2 | 6.5 ± 1.5 | 11.3 ± 2.2 | 8.0 ± 2.2 | ||||||||||||
| Ex 3 Test | 5.1 ± 1.4 | < 0.01 | 0.89 (0.67–0.97) | 0.38 | 5.6 ± 1.6 | < 0.01 | 0.89 (0.66–0.97) | 0.73 | 12.3 ± 4.0 | n.s | 0.68 (0.18–0.90) | 2.27 | 7.9 ± 1.6 | 0.03 | 0.77 (0.40–0.93) | 0.83 |
| Ex 3 Retest | 4.8 ± 1.1 | 5.5 ± 1.5 | 12.2 ± 3.8 | 7.6 ± 1.9 | ||||||||||||
| Ex 4 Test | 5.5 ± 0.9 | < 0.01 | 0.72 (0.27–0.91) | 0.55 | 5.0 ± 1.2 | < 0.01 | 0.83 (0.50–0.95) | 0.54 | 15.0 ± 2.8 | n.s | 0.36$ (0–0.76) | 2.05 | 6.7 ± 1.7 | < 0.01 | 0.89 (0.66–0.97) | 0.56 |
| Ex 4 Retest | 5.6 ± 1.1 | 4.9 ± 1.3 | 14.5 ± 2.3 | 6.8 ± 1.6 | ||||||||||||
All SSD measurements obtained during retest. ATT measurements obtained from the left knee. Measurements displayed as mean ± standard deviation (SD),
¢Significance level reported for the equivalence test with ± 1 mm equivalence boundary. All ICCs are significant (p < 0.05) if not flagged with § (p > 0.05)
Standard errors of measurement (SEM) are given in millimeter (mm)
NR not reliable; Ex Examiner; ICC (95% CI) Intraclass correlation (95% Confidence Interval)
Examiner 1–2 = advanced rater; Examiner 3–4 = beginners
Measures of intra-rater reliability by arthrometer based on pooled data from all four examiners
| Anterior tibial translation (ATT) | Side–to–side difference (SSD) | |||||
|---|---|---|---|---|---|---|
| Mean ± SD | ICC | SEM | Mean ± SD | ICC | SEM | |
| Rolimeter | 5.8 ± 1.4* | 0.9 | 0.41 | 0.8 ± 0.6* | 0.7 | 0.54 |
| KLT | 6.0 ± 1.7* | 0.9 | 0.56 | 1.1 ± 0.9* | 0.73 | 0.74 |
| KiRA | 12.3 ± 2.8 | 0.7 | 1.64 | 2.4 ± 1.5 | 0.11 | 3.03 |
| KT-1000 | 7.7 ± 1.9 | 0.89 | 0.63 | 1.1 ± 0.9* | 0.66 | 0.81 |
All SSD measurements obtained during retest. ATT measurements obtained from the left knee during retest
Standard errors of measurement (SEM) are given in mm
Ex Examiner
*Equivalent measurements between arthrometers
Fig. 3Intra-rater equivalence testing for Anterior Tibial Translation (ATT) and Side-to-Side Difference (SSD) in healthy individuals for all arthrometers between all four raters (A1, A2 = advanced raters; B1, B2 = beginner) Equivalence boundary’s are set to ± 1 mm. * Please note the different scaling (± 8 mm) for KiRA arthrometer