| Literature DB >> 34796120 |
Bin Feng1,2,3, Lei Yu1,2,3, Enwei Mo1,2,3, Liyuan Chen1,2,3, Jun Zhao1,2,3, Jiazhou Wang1,2,3, Weigang Hu1,2,3.
Abstract
PURPOSE: The difference in anatomical structure and positioning between planning and treatment may lead to bias in electronic portal image device (EPID)-based in vivo dosimetry calculations. The purpose of this study was to use daily CT instead of planning CT as a reference for EPID-based in vivo dosimetry calculations and to analyze the necessity of using daily CT for EPID-based in vivo dosimetry calculations in terms of patient quality assurance.Entities:
Keywords: EPID; IMRT verification; Monte Carlo method; portal dosimetry; quality assurance
Year: 2021 PMID: 34796120 PMCID: PMC8592931 DOI: 10.3389/fonc.2021.782263
Source DB: PubMed Journal: Front Oncol ISSN: 2234-943X Impact factor: 6.244
Figure 1Anatomical variations between planning CT (pCT) and daily CT (dCT). (A) pCT of a rectal patient; (B) dCT of a rectal patient; (C) pCT of an NPC patient; and (D) dCT of an NPC patient. NPC, Nasopharyngeal Carcinoma.
Figure 2The study workflow.
Figure 3The CT-Linac 506c.
Figure 4The quality assurance (QA) phantom, the RW3 slab phantom (PTW, Freiburg).
Patient cohort characteristics.
| Disease site | Beam number | Number of fractions | Prescription dose/cGy | Technology | Daily CT number | |
|---|---|---|---|---|---|---|
| Patient 1 | Prostate | 7 | 25 | 4,500 | sIMRT | 2 |
| Patient 2 | Nasopharyngeal | 6 | 30 | 6,000 | sIMRT | 2 |
| Patient 3 | Cervical | 7 | 25 | 5,000 | sIMRT | 2 |
| Patient 4 | Rectum | 7 | 25 | 5,000 | sIMRT | 2 |
| Patient 5 | Esophagus | 9 | 34 | 6,120 | dIMRT | 2 |
| Patient 6 | Oral cavity | 9 | 33 | 6,600 | dIMRT | 1 |
| Patient 7 | Rectum | 7 | 25 | 5,000 | sIMRT | 2 |
| Patient 8 | Rectum | 7 | 25 | 5,000 | sIMRT | 2 |
| Patient 9 | Oral cavity | 9 | 30 | 6,000 | dIMRT | 2 |
| Patient 10 | Rectum | 7 | 25 | 5,000 | sIMRT | 2 |
| Patient 11 | Rectum | 7 | 25 | 5,000 | sIMRT | 2 |
| Patient 12 | Rectum | 7 | 25 | 5,000 | sIMRT | 2 |
| Patient 13 | Rectum | 7 | 25 | 5,000 | sIMRT | 2 |
| Patient 14 | Lung | 6 | 28 | 5,040 | dIMRT | 2 |
| Patient 15 | Parotid gland | 6 | 30 | 6,600 | dIMRT | 2 |
| Patient 16 | Parotid gland | 6 | 30 | 6,000 | dIMRT | 2 |
| Patient 17 | Rectum | 7 | 25 | 5,000 | sIMRT | 2 |
| Patient 18 | Rectum | 7 | 25 | 5,000 | sIMRT | 2 |
| Patient 19 | Rectum | 7 | 25 | 5,000 | sIMRT | 2 |
| Patient 20 | Rectum | 7 | 25 | 5,100 | sIMRT | 1 |
dIMRT, dynamic intensity-modulated radiation therapy; sIMRT, static intensity-modulated radiation therapy.
Figure 5The cases of pretreatment quality assurance (QA). (A) An electronic portal image device (EPID)-measured image. (B) A phantom model-based calculated image. (C) The corresponding Gamma analysis results. (D) Profiles representing inlines and crosslines are shown in panels (A, B).
Figure 6The Gamma analysis results of planning CT (pCT) and daily CT (dCT). The red box diagram represents the pCT results. The blue box represents the dCT results.
Figure 7Gamma analysis results of the daily CT-based vs. planning CT-based methods. The daily CT-based results indicate a higher gamma passing rate; (A) 3 mm 3%; (B) 2 mm 3%; and (C) 2 mm 2%.
Figure 8The cases of electronic portal image device (EPID)-based in vivo dosimetry during treatment. (A) An EPID-measured image. (B) A planning CT (pCT)-based calculated image. (C) A daily CT (dCT)-based calculated image. (D) The corresponding Gamma analysis results between the pCT and measurement. (E) The corresponding Gamma analysis results between the dCT and measurement. (F) Profiles representing inlines and crosslines are shown in panels (A–C).
Mean and standard deviation of gamma passing rate for different error introduction and best threshold in varying variations.
| Failure mode | dCTerror vs. measurement | dCTerror vs. measurement | Best threshold (2 mm 2%) | |
|---|---|---|---|---|
| 2 mm 2% (γmean% ± STD) | 3 mm 3% (γmean% ± STD) | |||
| Gold criterion | 82.8 ± 14.7 | 93.6 ± 8.2 | – | |
| MLC systematic shift | ± 1 mm | 72.90 ± 12.36 | 90.34 ± 10.31 | 76.80 |
| ± 2 mm | 60.64 ± 11.59 | 81.60 ± 11.19 | 76.88 | |
| ± 5 mm | 42.66 ± 11.89 | 57.44 ± 12.61 | 66.25 | |
| MU scaling | 3% | 76.38 ± 14.94 | 90.94 ± 9.46 | 69.71 |
| 5% | 62.97 ± 11.07 | 84.96 ± 9.00 | 70.59 |
dCT, daily CT; MLC, multileaf collimator; MU, monitor unit.
Figure 9Distribution of gamma passing rates for daily CT (dCT)-based plan and a plan with a 3% and 5% scaling in monitor units (MUs) and a 1-, 2-, and 5-mm shift in multileaf collimator (MLC) leaf (left). Dilstribution of gamma passing rates for planning CT (pCT)-based plan and a plan with a 3% and 5% scaling in MU and a 1-, 2-, and 5-mm shift in MLC leaf (right).
Figure 10Receiver operating characteristic (ROC) curve for a systematic shift to multileaf collimator (MLC) leaves of ±1, ± 2, and ±5 mm (left). ROC curve for scaling of the monitor unit (MU) by 3% and 5% (right).
Figure 11An example of introducing error on electronic portal image device (EPID)-based in vivo dosimetry during treatment.
Figure 12(A) The gamma passing rate (3 mm, 3%) between the planning CT (pCT)-based calculated dose and measured dose for a patient with 10 treatment fractions. (B) The gamma passing rate (3 mm, 3%) between the daily CT (dCT)-based calculated dose and measured dose for a patient.