| Literature DB >> 31624462 |
Dragana Šegulja1, Danica Matišić1, Karmela Barišić2, Dunja Rogić1.
Abstract
INTRODUCTION: Our aim was to compare analytical specifications of two assays (monoclonal vs. polyclonal) for free light chains (FLCs) quantification optimized for two different analytical platforms, nephelometer ProSpec (Siemens, Erlangen, Germany) and turbidimetric analyser Optilite (The Binding Site, Birmingham, UK).Entities:
Keywords: M components; gammopathy; immunoassay; monoclonal; serum free light chains
Mesh:
Year: 2019 PMID: 31624462 PMCID: PMC6784423 DOI: 10.11613/BM.2019.030709
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Repeatability and reproducibility of FLC kappa and FLC lambda by both analytical systems
| ProSpec | 1.93 | 2.02 | 1.8 | 1.2 | 2.4 | 3.6 |
| Optilite | 2.78 | 1.40 | 3.3 | |||
| ProSpec | 2.15 | 2.61 | 2.2 | |||
| Optilite | 2.31 | 4.15 | 3.0 | |||
| ProSpec | 1.24 | 2.20 | 1.9 | |||
| Optilite | 1.15 | 2.82 | 2.0 | |||
| ProSpec | 3.44 | 3.23 | 3.1 | |||
| Optilite | 1.47 | 1.64 | 2.6 | |||
| CV – coefficient of variation. CVwr – repeability. CVbr – reproducibility. FLC – free light chain. | ||||||
Figure 1Distribution of kappa (A) and lambda (B) free light chaines (FLC) results in serum samples (N = 37) by two assays on ProSpec and Optilite analysers.
Concentrations of FLC kappa and FLC lambda with FLC ratios
| ProSpec | 6.15-2770.00 | 24.05 | 16.80-279.50 |
| Optilite | 5.58-6800.14 | 24.48 | 15.35-721.29 |
| ProSpec | 1.11-12100.40 | 30.80 | 13.87-166.25 |
| Optilite | 1.37-13380.22 | 21.49 | 7.93-108.71 |
| ProSpec | 0.002-432.432 | 1.243 | 0.533-3.244 |
| Optilite | 0.001-654.029 | 1.479 | 0.704-4.596 |
| FLC – free light chain. IQR – interquartile range. | |||
Figure 2Method comparison results with 95% confidence interval (N = 37) of FLC kappa assays on ProSpec and Optilite analysers. The regression equation indicates the existence of constant and proportional error (P = 0.740). FLC - free light chain.
Figure 3Method comparison results with 95% confidence interval (N = 37) of FLC lambda assays on ProSpec and Optilite analysers. The regression equation indicates the existence of proportional error (P = 0.460). Patients 1-3 are samples with FLC lambda values more than twelve times higher by Optilite than by ProSpec analyzer. FLC - free light chain.
Figure 4Concordance analysis according to criteria for progressive disease (FLC kappa or FLC lambda > 100 mg/L; 0.01 > kappa/lambda ratio > 100) and to FLC ratio. Reference ranges for FLC ratio were defined by the manufacturer. FLC – free light chains. M - meets criteria. NM - does not meet the criteria. N - within reference range. P - outside reference range. The grey area indicates the number of samples that gave concordant results.
Results of serum protein electrophoresis and immunofixation
| FLC kappa (mg/L) | 31.9/55.3 | 5.7/14.8 | 22.0/24.2 | |||
| FLC lambda (mg/L) | 302.0/7313.1 | 541.0/8522.0 | 959.0/13380.2 | |||
| FLC ratio | 0.11/< 0.01 | 0.01/< 0.01 | 0.02/< 0.01 | |||
| Electrophoresis | ||||||
| Immunofixation electrophoresis | Monoclonal free lambda chains | Monoclonal free lambda chains | Monoclonal IgG lambda + monoclonal free lambda chains | |||
| FLC – free light chain. Immunofixation is done by free kappa (ĸf), free lambda (λf), gamma (γ), kappa (ĸ) and lambda (λ) antisera. In all three samples is detected M-protein FLC lambda type. In Patient 3 total IgG on Cobas 6000cee (Roche Diagnostics, Rotkreuz, Switzerland) was 10.8 g/L and by densitometric analysis IgG λ monoclonal component was 5.2 g/L and monoclonal FLC λ 3.4 g/L. | ||||||