Ping Fu1,2, Sushil Devare3, Jian-Fang Liu4, Xiao-Ting Lv4, Peng Yin4, Bing-Ting Wu1,2, Ling Ke1,2, Yu Liu1,2, Hua Shan5.
Abstract
BACKGROUND AND OBJECTIVES: To prevent Treponema Pallidum (TP) transmission from blood transfusion, enzyme-linked immunosorbent assay (EIA) for anti-TP has been widely used in routine blood donation screening in China for many years. The aim of this study was to evaluate the performance of the Abbott CMIA assay for detection of anti-TP in Chinese blood donors.Entities:
Keywords: CMIA; China; blood donors; syphilis
Mesh:
Year: 2019 PMID: 31617243 PMCID: PMC6977309 DOI: 10.1002/jcla.23033
Source DB: PubMed Journal: J Clin Lab Anal ISSN: 0887-8013 Impact factor: 2.352
Routine screening results of samplesa
| Routine anti‐syphilis screening results | Number of the samples (%) |
|---|---|
| Reactive by two anti‐TP EIAs | 299 (12.36%) |
| Reactive by one anti‐TP EIA | 126 (5.20%) |
| Gray zone | 2 (0.1%) |
| Gray zone | 8 (0.33%) |
| Anti‐TP nonreactive with a screening reactive result in EIA for HBsAg, anti‐HCV or anti‐HIV | 809 (33.43%) |
| Nonreactive in donor screening | 1176 (48.60%) |
| Total | 2420 |
All samples were freshly frozen after collection before shipped to the Institute of Blood Transfusion (IBT) and restored at −20℃.
Gray zone: samples tested with borderline results (0.5 < S/CO < 1) by EIA.
Figure 1The algorithm of the screening and confirmatory testing. CMIA, Chemiluminescent microparticle immunoassay; CT, Confirmatory testing; EIA, Enzyme‐linked immunosorbent assay; TP, Treponema Pallidum
Results of the samples tested nonreactive for anti‐TP by two ELAs
| Anti‐TP EIA results | Total | Abbott CMIA | CT results | |||||
|---|---|---|---|---|---|---|---|---|
| + | − | Positive | Negative | IND | NT | Total | ||
| Negative | 1176 | 3 | 1173 | 0 | 1 | 1 | 1 | 3 |
| Gray zone | 2 | 0 | 2 | 0 | 0 | 1 | 1 | 2 |
| Gray zone | 8 | 0 | 8 | 0 | 8 | 0 | 0 | 8 |
|
anti‐TP nonreactive but HBsAg, anti‐HCV, or anti‐HIV reactive | 809 | 15 | 794 | 4 | 4 | 0 | 7 | 15 |
| Total | 1995 | 18 | 1977 | 4 | 13 | 2 | 9 | 28 |
Abbreviations: −, Nonreactive; +, Reactive; IND, Indeterminate; NT, not tested.
Gray zone: samples with border line results (0.5 < S/CO < 1).
Results of the samples tested reactive for anti‐TP by EIA
| CT | Total | Abbott CMIA | ELISA | |||
|---|---|---|---|---|---|---|
| + | − | R1+R2+ | R1+R2− | R1−R2+ | ||
| Positive | 246 | 243 | 3 | 236 | 3 | 7 |
| Negative | 143 | 4 | 139 | 45 | 35 | 63 |
| Indeterminate | 36 | 16 | 20 | 18 | 2 | 16 |
| Total | 425 | 263 | 162 | 299 | 40 | 86 |
Abbreviations: −, Nonreactive; +, Reactive; R: EIA reagent.
Performance of different assays for anti‐TP screening
| Different assays correctly identified | Confirmed positive n = 250 (%) | 95% CI | Confirmed negative n = 2123 (%) | 95% CI | Consistency |
|
|---|---|---|---|---|---|---|
| Abbott CMIA | 247 (98.80) | 97.44‐100.16 | 2114 (99.58) | 99.30‐99.85 | 99.49 | 0.97 |
| EIA R1 | 239 (95.60) | 93.04‐98.16 | 2043 (96.23) | 95.42‐97.04 | 96.17 | 0.82 |
| EIA R2 | 243 (97.20) | 95.14‐99.26 | 2015 (94.91) | 93.98‐95.85 | 95.15 | 0.78 |
| Two EIAs | 246 (98.40) | 96.83‐99.97 | 1980 (93.26) | 92.20‐94.33 | 93.81 | 0.74 |