| Literature DB >> 31616496 |
Chelsea Zimmerman1, Anastasia Albanese-O'Neill1, Michael J Haller1.
Abstract
The past 10 years have witnessed rapid advances in the technology used to treat patients with type 1 diabetes (T1D). While the disease burden is still high, these advances have contributed to improvements in both glycaemic control and quality of life for many of those affected. New technologies allow for individualisation of care, as patients are able to work with their providers to determine which systems best fit their lifestyle and needs. In addition, thanks to improved glucose monitoring technologies, patients can now simultaneously improve glycaemic control and reduce hypoglycaemia, thereby mitigating risk for acute and chronic complications. Technological advances in T1D care are rapidly moving us toward increasingly automated devices, which offer the promise of reduced disease burden. In this article, we review advances in glucose monitoring, insulin and glucagon delivery, and the applications and algorithms seeking to integrate novel technologies.Entities:
Keywords: Type 1 diabetes; advances; continuous glucose monitor; insulin pump; technology
Year: 2019 PMID: 31616496 PMCID: PMC6785958 DOI: 10.17925/EE.2019.15.2.70
Source DB: PubMed Journal: Eur Endocrinol ISSN: 1758-3772
Currently available continuous glucose monitors15
| CGM | Receiver | Size and weight | Warm-up/calibration | Dosing decisions | Duration | Alerts | Share feature | Age | MARD | Software |
|---|---|---|---|---|---|---|---|---|---|---|
| Abbot Freestyle Libre | Reader; Sensor stores 8 hours of data if not read | 1.38 in. diameter x 0.2 in; | 1-hour warm up; no calibration required | Yes | 14 days | No | No | ≥18 years and older | 10% | LibreView |
| Dexcom G5® | Receiver or Dexcom phone app | 1.5 x 0.9 x 0.5 in; 0.4 oz. with sensor | 2-hour warm up; calibration required every 12 hours | Yes | 7 days | Yes | Yes | ≥2 years and older | 9% | Clarity® |
| Dexcom G6® | Receiver or Dexcom phone app (Apple and Android devices) or integrated with t:slim X2™ pump | 4.02 x 2.44 x 0.46 in; 3.3 oz. | 2-hour warm up; no calibration required | Yes | 10 days | Yes | Yes | ≥2 years and older | 9% | Clarity® |
| Medtronic Guardian™ Sensor 3 | Integrated with Medtronic MiniMed 670G insulin pumps | 1.4 x 1.12 x 0.37 in; 0.19 oz. without sensor | 2-hour warm up; calibration required every 12 hours | No | 7 days | Yes | No | ≥7 years and older | 9.1% (8.7% with 3-4 calibrations a day | Carelink™ |
| Medtronic Guardian™ Connect | Guardian™ Connect smart phone app (Apple devices only) | 1.41 x 1.13 x 0.38 in; 0.04 oz. | 2-hour warm up; calibration required every 12 hours | No | 7 days | Yes | Yes | ≥14 years and older | 9.1% (8.7% with 3-4 calibrations a day | Sugar.lQ™ |
| Senseonics Eversense® | Smart phone app (Apple and Android devices) | 1.48 x 1.89 x 0.35 in; 0.39 oz. | 24-hour warm up; calibration required every 12 hours | No | 90 days in USA; 180 days in Europe | Yes, and transmitter vibrates alerts when away from phone | Yes | ≥18 years and older | 8.8% | Eversense® DMS |
CGM = continuous glucose monitor; in = inch; MARD = Mean absolute relative difference; oz = ounces.
Currently available insulin delivery systems34
| System | Reservoir Size | Basal | Bolus | Age | Meter/CGM integration | Additional Features | Waterproof |
|---|---|---|---|---|---|---|---|
| Insulet Corporation OmniPod® System | 200 unit | 0.05-30 units per hour in 0.05 unit-increments | 0.05-30 units in increments of 0.05 units. Whole unit insulin-to-carb ratios. | All ages | PDM with built-in glucose meter | Tubeless. PDM with database of 1,000 common foods with nutrition information. | Up to 7.6 m for 60 minutes |
| Insulet Corporation OmniPod DASH™ System | 200 unit | 0-30 units per hour in 0.05 unit-increments | 0.05-30 units in increments of 0.05 units. | All stages | Contour™ Next One meter | Tubeless. PDM with CalorieKing database. OmniPod DISPLAY™ app allows data to be viewed on a smart device and shared with up to 12 others. Omnipod VIEW™ app allows caregivers to view insulin delivery data remotely on their smartphone. | Up to 7.6 m for 60 minutes |
| Medtronic MiniMed™ 630G | 300 unit | 0.025-35 units per hour in 0.025 unit-increments | 0.025-25 units in increments of 0.025 units. Insulin-to-carb ratio with fraction of grams. | ≥16 years and older | Contour™ Next 2.4 meter | SmartGuard™ Technology with PLGS feature. | 3.7 m up to 24 hours |
| Medtronic MiniMed™ 670G | 300 unit | 0.025-35 units per hour in 0.025 unit-increments | 0.025-25 units in increments of 0.025 units. Insulin-to-carb ratio with fraction of grams. | ≥14 years and older | Contour™ Next 2.4 meter | Hybrid closed-loop pump with SmartGuard™ Technology. Auto Mode adjusts basal insulin based on CGM readings and recent insulin requirements. Users input boluses. Manual Mode with PLGS feature. | 3.7 m up to 24 hours |
| Tandem Diabetes Care t:slim X2™ | 300 unit | 0.1-15 units per hour in 0.001 unit-increments | 0.05-25 units in 0.01 unit increments. Insulin-to-carb ratio with fractions of grams. | ≥6 years and older | Dexcom G6 CGM | Tandem Device Updater remotely updates software without purchasing a new device. | 0.9 m for 30 minutes |
CGM = continuous glucose monitor; PDM = personal diabetes manager; PLGS = predictive low glucose suspend.