Mark P Christiansen1, Satish K Garg2, Ronald Brazg3, Bruce W Bode4, Timothy S Bailey5, Robert H Slover2, Ashley Sullivan6, Suiying Huang6, John Shin6, Scott W Lee6, Francine R Kaufman6.
Abstract
BACKGROUND: This study evaluated the accuracy and performance of a fourth-generation subcutaneous glucose sensor (Guardian™ Sensor 3) in the abdomen and arm.Entities:
Keywords: Glucose sensor; MARD; Mobile device; Sensor accuracy; Sensor-integrated pump
Mesh:
Substances:
Year: 2017 PMID: 28700272 PMCID: PMC5567873 DOI: 10.1089/dia.2017.0087
Source DB: PubMed Journal: Diabetes Technol Ther ISSN: 1520-9156 Impact factor: 6.118

Study flow and subject disposition.
Characteristics of Eligible Study Subjects
| Female, | 42 (47.7) |
| Male, | 46 (52.3) |
| Age, mean ± SD, years | 42.0 ± 19.08 |
| Weight, mean ± SD, kg | 83.5 ± 24.24 |
| BMI, mean ± SD, kg/m2 | 28.2 ± 7.17 |
| Diabetes classification, | |
| Type 1 | 62 (70.5) |
| Type 2, requiring insulin | 10 (11.4) |
| Type 2, not requiring insulin | 16 (18.2) |
| Prior CGM experience, | |
| Yes | 36 (40.9) |
| No | 52 (59.1) |
| Prior insulin pump experience, | |
| Yes | 47 (53.4) |
| No | 41 (46.6) |
BMI, body mass index; CGM, continuous glucose monitoring; SD, standard deviation.

Guardian Sensor 3 sensor.
Guardian Sensor 3 Sensor Accuracy During Frequent Sample Testing, by Day and Device
| N | N | |||||
|---|---|---|---|---|---|---|
| Day 1 | MiniMed 640G pump | Abdomen | 4294 | 13.0 ± 11.1 | 4136 | 11.7 ± 10.5 |
| Guardian Connect system | Abdomen | 4013 | 12.4 ± 11.1 | 3728 | 11.2 ± 10.8 | |
| GSR[ | Arm | 3390 | 10.8 ± 9.5 | 3591 | 10.3 ± 9.2 | |
| Day 3 | MiniMed 640G pump | Abdomen | 4533 | 8.9 ± 8.0 | 4378 | 8.3 ± 7.5 |
| Guardian Connect system | Abdomen | 4294 | 8.7 ± 8.1 | 4125 | 8.2 ± 7.4 | |
| GSR[ | Arm | 4243 | 8.1 ± 7.2 | 4198 | 7.8 ± 6.9 | |
| Day 7 | MiniMed 640G pump | Abdomen | 3263 | 9.5 ± 9.0 | 3150 | 8.7 ± 8.4 |
| Guardian Connect system | Abdomen | 3312 | 10.1 ± 11.6 | 3084 | 8.9 ± 11.1 | |
| GSR[ | Arm | 2893 | 8.5 ± 8.1 | 2982 | 8.1 ± 7.5 | |
| Overall | MiniMed 640G pump | Abdomen | 12090 | 10.6 ± 9.6 | 11664 | 9.6 ± 9.0 |
| Guardian Connect system | Abdomen | 11619 | 10.4 ± 10.4 | 10937 | 9.4 ± 9.8 | |
| GSR[ | Arm | 10526 | 9.1 ± 8.3 | 10771 | 8.7 ± 8.0 | |
Mean absolute relative difference is shown as Mean ± SD.
N = Number of sensor-YSI paired points.
Calibrating every 12 h.
Calibrating 3–4 times each day.
GSR data were obtained using the Guardian Link 3 transmitter calibration algorithm and calibration values from the MiniMed 640G insulin pump.
FST, frequent sample testing; MARD, Mean absolute relative difference.
Percent Distribution of Overall Mean Absolute Relative Difference Values, by Device
| N | ||||||
|---|---|---|---|---|---|---|
| Minimum calibrations[ | ||||||
| MiniMed 640G pump | 12090 | 59.7 | 26 | 8.9 | 3.9 | 1.5 |
| Guardian Connect system | 11619 | 61.3 | 25.5 | 8.6 | 2.9 | 1.6 |
| GSR[ | 10526 | 66.2 | 24.2 | 6.6 | 2.2 | 0.7 |
| One additional calibration[ | ||||||
| MiniMed 640G pump | 11664 | 63.8 | 24.6 | 7.3 | 3 | 1.3 |
| Guardian Connect system | 10937 | 65.6 | 23.7 | 7.7 | 1.8 | 1.2 |
| GSR[ | 10771 | 68.4 | 23.2 | 6 | 1.8 | 0.6 |
Calibrating every 12 h.
GSR data were obtained using the Guardian Link 3 transmitter calibration algorithm and calibration values from the MiniMed 640G insulin pump.
N = Number of sensor-YSI paired points.
Calibrating 3–4 times each day.

Distributions of overall ARD by device. The percentage of sensor glucose-YSI paired points associated with ranges (0% to >40%) of the overall ARD, according to each device, are shown when calibrating every 12 h (Top) and 3–4 times each day (Bottom). The percent distribution of values when calibrating every 12 h and 3–4 times each day are listed in Table 3. Black: MiniMed 640G insulin pump, Gray: Guardian Connect system, White: Glucose Sensor Recorder. ARD, absolute relative difference; YSI, Yellow Springs Instrument.

Overall ARD, as a function of YSI reference values, by device. The percent ARD distributions relative to YSI reference values (mg/dL) are shown, by device, when calibrating every 12 h (Top) and 3–4 times each day (Bottom). The horizontal line across each graph indicates the mean. GSR, glucose sensor recorder.
Guardian Sensor 3 Sensor-Yellow Springs Instrument Paired %20/20 Agreement by Device, When Calibrating Every 12 Hours
| ≤70 mg/dL[ | 92.5 (1056/1142) | 93.2 (1065/1143) | 93.1 (977/1049) |
| >70–180 mg/dL | 87.0 (6242/7173) | 88.3 (6098/6903) | 91.3 (5738/6282) |
| >180 mg/dL | 89.2 (3367/3775) | 90.7 (3240/3573) | 92.9 (2967/3195) |
| Overall | 88.2 (10665/12090) | 89.5 (10403/11619) | 92.0 (9682/10526) |
GSR data were obtained using the Guardian Link 3 transmitter calibration algorithm and calibration values from the MiniMed 640G insulin pump.
For reference blood glucose ≤80 mg/dL, agreement was based on ±20 mg/dL (Within-agreement paired points/total paired points).
YSI, Yellow Springs Instrument.
Guardian Sensor 3 Sensor-Yellow Springs Instrument Paired %30/30 Agreement by Device, When Calibrating Every 12 Hours
| ≤70 mg/dL[ | 99.3 (1134/1142) | 98.3 (1124/1143) | 99.0 (1038/1049) |
| >70–180 mg/dL | 95.5 (6847/7173) | 95.8 (6614/6903) | 97.5 (6126/6282) |
| >180 mg/dL | 96.3 (3634/3775) | 97.2 (3473/3573) | 98.1 (3134/3195) |
| Overall | 96.1 (11615/12090) | 96.5 (11211/11619) | 97.8 (10298/10526) |
GSR data were obtained using the Guardian Link 3 transmitter calibration algorithm and calibration values from the MiniMed 640G insulin pump.
For reference blood glucose ≤80 mg/dL, agreement was based on ±30 mg/dL (Within-agreement paired points/total paired points).
Guardian Sensor 3 Sensor-Yellow Springs Instrument Paired %40/40 Agreement by Device, When Calibrating Every 12 Hours
| ≤70 mg/dL[ | 99.8 (1140/1142) | 98.9 (1130/1143) | 99.7 (1046/1049) |
| >70–180 mg/dL | 98.7 (7077/7173) | 98.5 (6798/6903) | 99.5 (6248/6282) |
| >180 mg/dL | 99.0 (3738/3775) | 99.4 (3553/3573) | 99.9 (3192/3195) |
| Overall | 98.9 (11955/12090) | 98.8 (11481/11619) | 99.6 (10486/10526) |
GSR data were obtained using the Guardian Link 3 transmitter calibration algorithm and calibration values from the MiniMed 640G insulin pump.
For reference blood glucose ≤80 mg/dL, agreement was based on ±40 mg/dL (Within-agreement paired points/total paired points).
Guardian Sensor 3 Sensor-Yellow Springs Instrument Paired %20/20 Agreement by Device, When Calibrating Three to Four Times Each Day
| ≤70 mg/dL[ | 92.8 (1012/1090) | 93.3 (992/1063) | 93.8 (1004/1070) |
| >70–180 mg/dL | 88.9 (6213/6990) | 90.2 (5875/6516) | 92.3 (5953/6453) |
| >180 mg/dL | 93.5 (3351/3584) | 94.5 (3172/3358) | 94.6 (3072/3248) |
| Overall | 90.7 (10576/11664) | 91.8 (10039/10937) | 93.1 (10029/10771) |
GSR data were obtained using the Guardian Link 3 transmitter calibration algorithm and calibration values from the MiniMed 640G insulin pump.
For reference blood glucose ≤80 mg/dL, agreement was based on ±20 mg/dL (Within-agreement paired points/total paired points).
Guardian Sensor 3 Sensor-Yellow Springs Instrument Paired %30/30 Agreement by Device, When Calibrating Three to Four Times Each Day
| ≤70 mg/dL[ | 99.4 (1083/1090) | 98.2 (1044/1063) | 99.0 (1059/1070) |
| >70–180 mg/dL | 96.2 (6725/6990) | 97.3 (6338/6516) | 97.9 (6316/6453) |
| >180 mg/dL | 97.6 (3497/3584) | 99.0 (3323/3358) | 99.0 (3216/3248) |
| Overall | 96.9 (11305/11664) | 97.9 (10705/10937) | 98.3 (10591/10771) |
GSR data were obtained using the Guardian Link 3 transmitter calibration algorithm and calibration values from the MiniMed 640G insulin pump.
For reference blood glucose ≤80 mg/dL, agreement was based on ±30 mg/dL (Within-agreement paired points/total paired points).
Guardian Sensor 3 Sensor-Yellow Springs Instrument Paired %40/40 Agreement by Device, When Calibrating Three to Four Times Each Day
| ≤70 mg/dL[ | 100.0 (1090/1090) | 98.8 (1050/1063) | 99.7 (1067/1070) |
| >70–180 mg/dL | 98.9 (6913/6990) | 99.0 (6449/6516) | 99.5 (6423/6453) |
| >180 mg/dL | 99.4 (3561/3584) | 99.7 (3349/3358) | 100.0 (3247/3248) |
| Overall | 99.1 (11564/11664) | 99.2 (10848/10937) | 99.7 (10737/10771) |
GSR data were obtained using the Guardian Link 3 transmitter calibration algorithm and calibration values from the MiniMed 640G insulin pump.
For reference blood glucose ≤80 mg/dL, agreement was based on ±40 mg/dL.
(Within-agreement paired points/total paired points).
Consensus Error Grid Analyses of Sensors by Device, When Calibrating Every 12 Hours
| N | N | N | ||||
|---|---|---|---|---|---|---|
| Zone A + B | 12052 | 99.9 | 11577 | 99.8 | 10499 | 99.9 |
| Zone A | 10581 | 87.7 | 10362 | 89.4 | 9652 | 91.8 |
| Zone B | 1471 | 12.2 | 1215 | 10.5 | 847 | 8.1 |
| Zone C | 11 | 0.1 | 12 | 0.1 | 10 | 0.1 |
| Zone D | 0 | 0 | 6 | 0.1 | 0 | 0 |
| Zone E | 0 | 0 | 0 | 0 | 0 | 0 |
Values are from overall ranges (40–400 mg/dL).
GSR data were obtained using the Guardian Link 3 transmitter calibration algorithm and calibration values from the MiniMed 640G insulin pump.
N = Number of sensor-YSI paired points.
Clarke Error Grid Analyses of Sensors by Device, When Calibrating Every 12 Hours
| N | N | N | ||||
|---|---|---|---|---|---|---|
| Zone A + B | 11951 | 99.1 | 11496 | 99.1 | 10420 | 99.2 |
| Zone A | 10572 | 87.6 | 10292 | 88.8 | 9619 | 91.5 |
| Zone B | 1379 | 11.4 | 1204 | 10.4 | 801 | 7.6 |
| Zone C | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 |
| Zone D | 111 | 0.9 | 99 | 0.9 | 89 | 0.8 |
| Zone E | 1 | 0.0 | 0 | 0.0 | 0 | 0.0 |
Values are from overall ranges (40–400 mg/dL).
GSR data were obtained using the Guardian Link 3 transmitter calibration algorithm and calibration values from the MiniMed 640G insulin pump.
N = Number of sensor-YSI paired points.
Consensus Error Grid Analyses of Sensors by Device, When Calibrating Three to Four Times Each Day
| N | N | N | ||||
|---|---|---|---|---|---|---|
| Zone A + B | 11631 | 99.9 | 10897 | 99.8 | 10745 | 99.9 |
| Zone A | 10504 | 90.3 | 10026 | 91.9 | 10001 | 93.0 |
| Zone B | 1127 | 9.7 | 871 | 8.0 | 744 | 6.9 |
| Zone C | 7 | 0.1 | 12 | 0.1 | 9 | 0.1 |
| Zone D | 0 | 0.0 | 6 | 0.1 | 0 | 0.0 |
| Zone E | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 |
Values are from overall ranges (40–400 mg/dL).
GSR data were obtained using the Guardian Link 3 transmitter calibration algorithm and calibration values from the MiniMed 640G insulin pump.
N = Number of sensor-YSI paired points.
Clarke Error Grid Analyses of Sensors by Device, When Calibrating Three to Four Times Each Day
| N | N | N | ||||
|---|---|---|---|---|---|---|
| Zone A + B | 11541 | 99.2 | 10833 | 99.2 | 10675 | 99.3 |
| Zone A | 10490 | 90.1 | 9938 | 91.0 | 9964 | 92.7 |
| Zone B | 1051 | 9.0 | 895 | 8.2 | 711 | 6.6 |
| Zone C | 0 | 0.0 | 0 | 0.0 | 0 | 0.0 |
| Zone D | 96 | 0.8 | 82 | 0.8 | 79 | 0.7 |
| Zone E | 1 | 0.0 | 0 | 0.0 | 0 | 0.0 |
Values are from overall ranges (40–400 mg/dL).
GSR data were obtained using the Guardian Link 3 transmitter calibration algorithm and calibration values from the MiniMed 640G insulin pump.
N = Number of sensor-YSI paired points.