| Literature DB >> 31600932 |
Qiqi Sun1,2, Wei Zheng3, Chao Lin4, Dongxuan Shen5.
Abstract
Immunoassays have been widely used in scientific research and clinical diagnosis due to their versatile detection capability and high specificity. Immunoagglutination assays are kinds of immunoassay, which can simply and rapidly measure the concentration of analytes. In this work, we developed a low-cost micro-volume nephelometric system for quantitative immunoagglutination assays. We used off-the-shelf components to build the system, and the total cost of key components is only about 20 US dollars. The total detection volume in our system was as low as 3 µL, which could significantly reduce the reagent cost and required sample volume. We further evaluated the system performance via the immunoagglutination assay to measure the concentration of C-reactive protein, a plasma protein with levels rising in response to inflammation. The results demonstrated that our system could measure the concentration of analytes with relatively high sensitivity and precision within four minutes, and has high potential to be applied for clinical diagnostic tests.Entities:
Keywords: C-reactive protein; immunoagglutination assay; immunoassay; light scattering; nephelometry
Mesh:
Substances:
Year: 2019 PMID: 31600932 PMCID: PMC6832725 DOI: 10.3390/s19204359
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1The schematic of nephelometric system. Red arrows indicate light path.
Figure 2The 3D image of key components of nephelometric system. Only parts of scattered light are drawn. Inserted photograph shows the real image of glass capillary and its mount.
Figure 3Polar graph of scattering patterns of spheres with diameters of 100 nm and 200 nm. Light is incident from the left on a sphere located at the center of the polar plot. The data are normalized so that the integral of it over 4π steradians is unity.
Figure 4Intensity change of scattered light over time. (a) Intensity change of light in two experiments using a same CRP sample (2.5 mg/L) but different capillaries. (b) Intensity change of light in experiments using CRP samples with different concentrations. The curves were biased to have the same start point.
Figure 5Plot of rate of intensity change versus CRP concentration. The inserted image is a zoomed-in image on low levels of CRP concentration. Error bars represent SD from n = 3 measurements.