| Literature DB >> 31531292 |
Nainggolan A1,2, Pawiro S A1.
Abstract
BACKGROUND: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated.Entities:
Keywords: Computer-Assisted; Phantoms ; Imaging Radiotherapy Planning ; Radiometry ; Radiotherapy, Intensity-Modulated
Year: 2019 PMID: 31531292 PMCID: PMC6709355 DOI: 10.31661/jbpe.v0i0.839
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Figure1(A) Solid water phantom and CT scan image (B) MatriXX and CT scan image.
Figure2Illustration of Preliminary (A) test P1 and (B) test P2.
Figure3Mock of (a) multi target, (b) prostate, (c) head and neck, and (d) C-shape structures in Eclipse.
Beam parameter setup for VMAT and IMRT.
| Number of beam/arcs | Beam arrangement | Collimator Angle | Prescribe dose (Gy) | Dose per fraction (Gy) | |
|---|---|---|---|---|---|
|
| |||||
| Multitarget | 7 | 50º from anterior | 5º | 50 | 2 |
| Prostate | 7 | 50º from anterior | 5º | 80 | 2 |
| Head and neck | 9 | 40º from anterior | 0º | 50 | 2 |
| C-Shape | 9 | 40º from anterior | 5º | 50 | 2 |
|
| |||||
| Multitarget | 2 | CCW 179-181, CW 181-179 | 15º | 50 | 2 |
| Prostate | 2 | CCW 179-181, CW 181-179 | 30º | 80 | 2 |
| Head and neck | 2 | CCW 179-181, CW 181-179 | 30º | 50 | 2 |
| C-Shape | 2 | CCW 179-181, CW 181-179 | 30º | 50 | 2 |
Technical specifications of the ionization chambers used in the study.
| Ionization chamber | FC65G, S/N: 2311 | CC13, S/N: 9692 | CC01, S/N: 9755 |
|---|---|---|---|
| Active volume | 0.65 cm3 | 0.13 cm3 | 0.01 cm3 |
| Polarizing voltage | 300 V | 300 V | 300 V |
| Wall material | Aluminum + Graphite | Shonka | Steel + Shonka |
Figure4VMAT and IMRT dose distributions for all test commissioning.
VMAT and IMRT treatment plan dose goals results .
| Test/Planning parameters | Volume (cm3) | Plan goal (cGy) | IMRT | VMAT | AAPM TG 119 Mean (cGy) | IMRT/ TG 119 | VMAT/ TG 119 |
|---|---|---|---|---|---|---|---|
| Mean (cGy) | Mean (cGy) | ||||||
|
| |||||||
| Central D99 | 51.6 | ˃5000 | 5023 ± 3 | 5038 ± 23 | 4955 ± 162 | 1.01 | 1.02 |
| Central D10 | ˂5300 | 5286 ± 5 | 5279 ± 4 | 5455 ± 173 | 0.97 | 0.97 | |
| Superior D99 | 51.4 | ˃2500 | 2514 ± 13 | 2527 ± 1 | 2516 ± 85 | 1.00 | 1.00 |
| Superior D10 | ˂3500 | 3431 ± 11 | 3422 ± 44 | 3412 ± 304 | 1.01 | 1.00 | |
| Inferior D99 | 51.5 | ˃1250 | 1374 ± 3 | 1344 ± 57 | 1407 ± 185 | 0.98 | 0.96 |
| Inferior D10 | ˂2500 | 2250 ± 18 | 2445 ± 24 | 2418 ± 272 | 0.93 | 1.01 | |
|
| |||||||
| PTV Prostate D95 | 107.9 | ˃7560 | 7663 ± 36 | 7838 ± 17 | 7566 ± 21 | 1.01 | 1.04 |
| PTV Prostate D5 | ˂8300 | 8129 ± 10 | 8144 ± 38 | 8143 ± 156 | 0.97 | 1.00 | |
| Rectum D30 | 20.71 | ˂7000 | 6266 ± 44 | 6308 ± 49 | 6536 ± 297 | 1.00 | 0.97 |
| Rectum D10 | ˂7500 | 7284 ± 27 | 7331 ± 43 | 7303 ± 15 | 1.01 | 1.00 | |
| Bladder D30 | 63.59 | ˂7000 | 4624 ± 62 | 5103 ± 67 | 4394 ± 878 | 0.98 | 1.16 |
| Bladder D10 | ˂7500 | 5913 ± 64 | 6375 ± 97 | 6269 ± 815 | 0.93 | 1.02 | |
|
| |||||||
| PTV D90 | 721.01 | 5000 | 5072 ± 6 | 5037 ± 9 | 5028 ± 58 | 1.01 | 1.00 |
| PTV D99 | ˃4650 | 4856 ± 7 | 4880 ± 22 | 4704 ± 52 | 1.03 | 1.04 | |
| PTV D20 | ˂5500 | 5247 ± 5 | 5195 ± 5 | 5299 ± 93 | 0.99 | 0.98 | |
| Cord maximum | 19.18 | ˂4000 | 3887 ± 4 | 3877 ± 34 | 3741 ± 250 | 1.04 | 1.04 |
| Parotid LT D50 | 10.21 | ˂2000 | 1927 ± 3 | 1874 ± 29 | 1798 ± 184 | 1.07 | 1.04 |
| Parotid RT D50 | 9.98 | ˂2000 | 1943 ± 2 | 1878 ± 6 | 1798 ± 184 | 1.08 | 1.04 |
|
| |||||||
| PTV D95 | 283.91 | 5000 | 5016 ± 4 | 5012 ± 7 | 5010 ± 17 | 1.00 | 1.00 |
| PTV D10 | ˂5500 | 5351 ± 2 | 5340 ± 5 | 5440 ± 52 | 0.98 | 0.98 | |
| Core D10 | 31.54 | ˂2500 | 2237 ± 22 | 2234 ± 21 | 2200 ± 314 | 1.02 | 1.02 |
|
| |||||||
| PTV D95 | 283.91 | 5000 | 5057 ± 32 | 5024 ± 3 | 5011 ± 165 | 1.01 | 1.00 |
| PTV D10 | ˂5500 | 5596 ± 53 | 5447 ± 4 | 5702 ± 220 | 0.98 | 0.96 | |
| Core D10 | 31.54 | ˂1000 | 1558 ± 19 | 1420 ± 4 | 1630 ± 307 | 0.96 | 0.87 |
The results of treatment plan comparison between VMAT and IMRT.
| Multi Target | Prostate | Head and Neck | C-shape easy | C-shape hard | ||||||
|---|---|---|---|---|---|---|---|---|---|---|
| IMRT | VMAT | IMRT | VMAT | IMRT | VMAT | IMRT | VMAT | IMRT | VMAT | |
|
| 1.357 | 1.372 | 0.980 | 1.160 | 1.198 | 1.128 | 1.193 | 1.349 | 1.355 | 1.400 |
|
| 0.042 | 0.047 | 0.074 | 0.055 | 0.067 | 0.073 | 0.113 | 0.090 | 0.154 | 0.133 |
|
| 7 | 2 | 7 | 2 | 9 | 2 | 9 | 2 | 9 | 2 |
|
| 200 | 200 | 200 | 200 | 200 | 200 | 200 | 200 | 200 | 200 |
|
| 532 | 372 | 469 | 560 | 1374 | 497 | 1174 | 902 | 1464 | 985 |
|
| 1.43 | 1 | 0.84 | 1 | 1.76 | 1 | 1.30 | 1 | 1.49 | 1 |
The point dose measurements for preliminary test P1.
| Test | Ion chamber | Planned dose (cGy) | Measured dose (cGy) | Dose variation |
|---|---|---|---|---|
|
| CC01 | 200 | 201.93 | 0.0096 |
| CC13 | 200 | 202.39 | 0.0120 | |
| FC65G | 200 | 202.78 | 0.0139 |
The point dose measurements for preliminary test P2.
| Test | Location | Planned dose (cGy) | CC01 | CC13 | FC65G | |||
|---|---|---|---|---|---|---|---|---|
| Measured dose (cGy) | Dose variation | Measured dose (cGy) | Dose variation | Measured dose (cGy) | Dose variation | |||
|
| 1st band left of isocenter | 40 | 40.71 | 0.0178 | 40.08 | 0.0019 | 40.09 | 0.0022 |
| 2st band left of isocenter | 80 | 80.78 | 0.0098 | 80.02 | 0.0002 | 80.15 | 0.0019 | |
| Pita 3 di isocenter | 120 | 121.39 | 0.0116 | 120.65 | 0.0054 | 120.83 | 0.0069 | |
| 1st band right of isocenter | 160 | 161.23 | 0.0077 | 159.96 | -0.0003 | 160.19 | 0.0012 | |
| 2nd band right of isocenter | 200 | 202.32 | 0.0116 | 200.72 | 0.0036 | 200.78 | 0.0039 | |
| Mean dose variation | 0.0117 | 0.0022 | 0.0032 | |||||
VMAT and IMRT doses variation, standard deviation, and CL at high dose area.
| Test | Location | IMRT | VMAT | ||||
|---|---|---|---|---|---|---|---|
| CC01 | CC13 | FC65G | CC01 | CC13 | FC65G | ||
| Multi target | Isocenter | 0.00003 | -0.0052 | 0.0006 | -0.0003 | -0.0003 | -0.0137 |
| Prostate | Isocenter | -0.0004 | -0.0060 | -0.0013 | 0.0091 | -0.0123 | -0.0157 |
| Head and neck | Isocenter | 0.0087 | -0.0069 | -0.0030 | 0.0052 | -0.0110 | -0.0125 |
| C shape (easy) | 2.5 cm anterior to isocenter | 0.0047 | -0.0082 | -0.0067 | 0.0103 | -0.0073 | -0.0053 |
| C shape (hard) | 2.5 cm anterior to isocenter | 0.0035 | -0.0110 | -0.0114 | -0.0005 | -0.0017 | -0.0034 |
| Mean of dose variations | 0.0033 | -0.0075 | -0.0044 | 0.0048 | -0.0065 | -0.0101 | |
| Standard deviation σ | 0.0037 | 0.0023 | 0.0048 | 0.0051 | 0.0054 | 0.0054 | |
| CL = |mean| + 1.96 σ | 0.0106 | 0.0119 | 0.0137 | 0.0147 | 0.0171 | 0.0208 | |
VMAT and IMRT doses variation, standard deviation, and CL at low dose area.
| Test | Location | IMRT | VMAT | ||||
|---|---|---|---|---|---|---|---|
| CC01 | CC13 | FC65G | CC01 | CC13 | FC65G | ||
| Multi target | 4 cm inferior to isocenter | 0.0041 | -0.0042 | -0.0110 | 0.0086 | -0.0041 | -0.0206 |
| Prostate | 2.5 cm posterior to isocenter | 0.0072 | -0.0146 | -0.0154 | 0.0006 | 0.0142 | 0.0331 |
| Head and neck | 4 cm posterior to isocenter | 0.0048 | -0.0066 | -0.0264 | 0.0055 | -0.0049 | -0.0082 |
| C shape (easy) | Isocenter | 0.0067 | -0.0030 | -0.0029 | 0.0007 | 0.0089 | 0.0007 |
| C shape (hard) | Isocenter | 0.0043 | -0.0044 | -0.0077 | 0.0072 | -0.0047 | -0.0057 |
| Mean of dose variations | 0.0054 | -0.0065 | -0.0127 | 0.0045 | 0.0019 | -0.0002 | |
| Standard deviation σ | 0.0014 | 0.0047 | 0.0089 | 0.0037 | 0.0090 | 0.0201 | |
| CL = |mean| + 1.96 σ | 0.0082 | 0.0158 | 0.0302 | 0.0117 | 0.0195 | 0.0396 | |
Gamma evaluation for VMAT and IMRT at high dose area.
| Test | Location | 3%, 3 mm gamma pass | 2%, 2 mm gamma pass | ||
|---|---|---|---|---|---|
| IMRT | VMAT | IMRT | VMAT | ||
| Multi target | Isocenter | 99.23 | 99.80 | 97.42 | 98.77 |
| Prostate | Isocenter | 99.04 | 99.62 | 97.31 | 98.42 |
| Head and neck | Isocenter | 99.44 | 99.57 | 96.81 | 96.30 |
| C shape (easy) | 2.5 cm anterior to isocenter | 96.59 | 99.14 | 89.05 | 95.33 |
| C shape (hard) | 2.5 cm anterior to isocenter | 98.65 | 98.77 | 95.46 | 95.50 |
| Overall mean | 98.59 | 99.38 | 95.21 | 96.86 | |
| Overall Standard deviation σ | 1.16 | 0.42 | 3.53 | 1.63 | |
| CL = |mean| + 1.96 σ | 3.68 | 1.44 | 11.71 | 6.33 | |
Gamma evaluation for VMAT and IMRT at low dose area.
| Test | Location | 3%, 3 mm gamma pass | 2%, 2 mm gamma pass | ||
|---|---|---|---|---|---|
| IMRT | VMAT | IMRT | VMAT | ||
| Multi target | 4 cm inferior to isocenter | 95.63 | 98.26 | 89.80 | 94.00 |
| Prostate | 2.5 cm posterior to isocenter | 95.43 | 97.20 | 89.96 | 94.24 |
| Head and neck | 4 cm posterior to isocenter | 98.71 | 98.66 | 88.89 | 94.04 |
| C shape (easy) | Isocenter | 97.87 | 98.76 | 90.42 | 94.92 |
| C shape (hard) | Isocenter | 98.75 | 96.69 | 94.71 | 93.14 |
| Overall mean | 97.28 | 97.91 | 90.75 | 94.07 | |
| Overall Standard deviation σ | 1.63 | 0.92 | 2.28 | 0.64 | |
| CL = |mean| + 1.96 σ | 5.93 | 3.89 | 13.71 | 7.18 | |
Figure5The comparison between the gamma index map of the dose distribution for VMAT and IMRT.