Ketan R Parmar1, Christine Dickinson2, Bruce J W Evans3.
Abstract
BACKGROUND: Various instruments have been developed to measure aligning prism, the prism that eliminates a fixation disparity (associated heterophoria). This includes the established Mallett near vision unit and recently developed Thomson Vision Toolbox on the iPad. With no previous research investigating the agreement between these instruments, practitioners may question if they can be used interchangeably.Entities:
Keywords: Aligning prism; Associated phoria; Disparidad de fijación; Fixation disparity; Foria asociada; Mallett unit; Marcadores Nonius; Nonius markers; Prisma de alineamiento; Unidad de Mallett
Mesh:
Year: 2019 PMID: 31501055 PMCID: PMC6978586 DOI: 10.1016/j.optom.2019.03.002
Source DB: PubMed Journal: J Optom ISSN: 1989-1342
Figure 1(a and b) The Mallett near fixation disparity test indicating the location of the individual components.
Figure 2(a and b) An image of the Near Thomson Fixation Disparity test indicating the location of individual components.
The proportion of participants with a horizontal fixation disparity (FD) on each instrument.
| Horizontal FD on Mallett unit | No Horizontal FD on Mallett unit | |
|---|---|---|
| Horizontal FD on iPad | 19 (23.75%) | 7 (8.75%) |
| No Horizontal FD on iPad | 17 (21.25%) | 37 (46.25%) |
The proportion of participants with a vertical fixation disparity (FD) on each instrument.
| Vertical FD on Mallett unit | No Vertical FD on Mallett unit | |
|---|---|---|
| Vertical FD on iPad | 2 (2.5%) | 6 (7.5%) |
| No Vertical FD on iPad | 6 (7.5%) | 66 (82.5%) |
The proportion of participants with a significant horizontal fixation disparity (FD) on each instrument.
| Significant horizontal FD on Mallett unit | Non-significant horizontal FD on Mallett unit | |
|---|---|---|
| Significant horizontal FD on iPad | 7 (8.75%) | 7 (8.75%) |
| Non-significant horizontal FD on iPad | 10 (12.5%) | 56 (70%) |
The proportion of participants with a significant vertical fixation disparity (FD) on each instrument.
| Significant vertical FD on Mallett unit | Non-significant vertical FD on Mallett unit | |
|---|---|---|
| Significant vertical FD on iPad | 0 (0%) | 0 (0%) |
| Non-significant vertical FD on iPad | 2 (2.5%) | 78 (97.5%) |
Figure 3Bland and Altman plot for horizontal aligning prism data. The number of subjects at each data point is represented by the accompanying number.
Figure 4Bland and Altman plot for vertical aligning prism data. The number of subjects at each data point is represented by the accompanying number.