Daniel Dirkmann1, Elisabeth Nagy2, Martin W Britten2, Jürgen Peters2.
Abstract
BACKGROUND: Since inadequate heparin anticoagulation and insufficient reversal can result in complications during cardiopulmonary bypass (CPB) surgery, heparin anticoagulation monitoring by point-of-care (POC) activated clotting time (ACT) measurements is essential for CPB initiation, maintainance, and anticoagulant reversal. However, concerns exist regarding reproducibility of ACT assays and comparability of devices.Entities:
Keywords: ACT; Activated clotting time; Anticoagulation; Cardiopulmonary bypass; Heparin; Method comparison; Protamine
Mesh:
Substances:
Year: 2019 PMID: 31492108 PMCID: PMC6728977 DOI: 10.1186/s12871-019-0846-z
Source DB: PubMed Journal: BMC Anesthesiol ISSN: 1471-2253 Impact factor: 2.217
Demographics of the study population
| Gender (number; male/female) | 19 / 11 |
|---|---|
| Age (years; mean ± SD) | 70,5 ± 11 |
| Height (cm; mean ± SD) | 170,2 ± 8,8 |
| Weight (kg; mean ± SD) | 80,3 ± 17,3 |
| Type of surgery | (n) |
| Coronary artery bypass surgery alone | 13 |
| Coronary artery bypass + aortic valve surgery | 5 |
| Coronary artery bypass + mitral valve surgery | 1 |
| Coronary artery bypass + ascending aortic surgery | 1 |
| Aortic valve surgery | 3 |
| Aortic valve + ascending aortic surgery | 4 |
| Mitral valve surgery | 3 |
Data are numbers or means ± standard deviation
Fig. 1Linear correlation and Bland-Altman-Plots for comparison of results from the two i-STAT devices (pannel a and b), the two Haemochron signature elite devices (panels c and d), and for results from i-STATs and Haemochron assays (pannel e and f)
Fig. 2Linear correlation for the ACT measurement below 200 s (panels a, c, and e), for the whole measurement range up to 1000 s (panels b, d, and f), for comparison of the ACT measurements from the two i-STAT devices, the two Haemochron devices, and for the results from i-STAT and Haemochron assays respectively. Dotted lines indicate the respective clinical target values (i.e.,120 s; pannels a, c, and e; and 430 s; panels b, d, and f), respectively. Disconcordant, clinically relevant discrepant resulsts are depicted in the light grey fields