Evaluation of a new point-of-care celite-activated clotting time analyzer in different clinical settings. The i-STAT celite-activated clotting time test.

Rita Paniccia1, Sandra Fedi, Fiorella Carbonetto, Daniela Noferi, Paolo Conti, Brunella Bandinelli, Betti Giusti, Lucia Evangelisti, Paola Pretelli, Mara F G Palmarini, Rosanna Abbate, Domenico Prisco.   

Abstract

BACKGROUND: Activated clotting time (ACT) is used to monitor heparin therapy during cardiopulmonary bypass, interventional cardiology, and hemodialysis. Traditionally, ACT is performed by use of the Hemochron system. Recently, a new device, the i-STAT system, has been introduced to measure ACT. The aim of this study was to correlate the performances of these two systems and to compare ACT values with heparin levels.
METHODS: One hundred sixty-five samples from 29 patients undergoing cardiopulmonary bypass or hemodialysis were assayed in duplicate with two Hemochron and two i-STAT devices. Heparin levels were determined by anti-factor Xa assay.
RESULTS: The Hemochron ACT ranged from 88 to 1,028 s, and the i-STAT ACT ranged from 80 to 786 s. Heparin plasma levels ranged from 0.01 to 10.8 U/mL. Bland-Altman analysis showed a mean difference between the two methods of 24 +/- 101 s. Strong relationships between anti-factor Xa activity and Hemochron ACTs (r2 = 0.69, P < 0.001) and i-STAT ACTs (r2 = 0.79, P < 0.001) were observed. During cardiac surgery, significant correlations were found: Hemochron, r2 = 0.61, P < 0.001 and i-STAT, r2 = 0.74, P < 0.001. During hemodialysis, relationships between anti-factor Xa activity and ACTs were found: Hemochron, r2 = 0.62, P < 0.001 and i-STAT, r2 = 0.55, P < 0.001.
CONCLUSIONS: During cardiopulmonary bypass procedure and hemodialysis, i-STAT provides measurements of clotting time quite similar to Hemochron ACT, which were significantly correlated with heparin levels.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12826842     DOI: 10.1097/00000542-200307000-00012

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  6 in total

Review 1.  Point-of-care testing for anticoagulation monitoring in neuroendovascular procedures.

Authors:  H M Hussein; A L Georgiadis; A I Qureshi
Journal:  AJNR Am J Neuroradiol       Date:  2011-09-15       Impact factor: 3.825

2.  Clinical Evaluation of Measuring the ACT during Elective Cardiac Surgery with Two Different Devices.

Authors:  Florian Falter; Nabeel Razzaq; Martin John; Jens Fassl; Markus Maurer; Sean Ewing; Ross Hofmeyr
Journal:  J Extra Corpor Technol       Date:  2018-03

3.  Point-of-Care Measurement of Kaolin Activated Clotting Time during Cardiopulmonary Bypass: A Single Sample Comparison between ACT Plus and i-STAT.

Authors:  Nousjka P A Vranken; Amber J A J Theelen; Tamar Orbons; Paul J C Barenbrug; Patrick W Weerwind
Journal:  J Extra Corpor Technol       Date:  2021-03

Review 4.  Biosensors in clinical chemistry - 2011 update.

Authors:  Paul D'Orazio
Journal:  Clin Chim Acta       Date:  2011-06-26       Impact factor: 3.786

5.  Point-of-Care Testing of Hemostasis in Cardiac Surgery.

Authors:  Domenico Prisco; Rita Paniccia
Journal:  Thromb J       Date:  2003-05-06

6.  Point-of-care measurement of activated clotting time for cardiac surgery as measured by the Hemochron signature elite and the Abbott i-STAT: agreement, concordance, and clinical reliability.

Authors:  Daniel Dirkmann; Elisabeth Nagy; Martin W Britten; Jürgen Peters
Journal:  BMC Anesthesiol       Date:  2019-09-06       Impact factor: 2.217

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.