Comparison of ACT point-of-care measurements: repeatability and agreement.

Y P J Bosch1, Y M Ganushchak, D S de Jong.   

Abstract

BACKGROUND: Accurate control of heparin anticoagulation is necessary during all stages of cardiopulmonary bypass (CPB). The activated clotting time, first described by Hattersley in 1966, is mostly used for determination of anticoagulation. Either celite or kaolin are used as activators. An ACT value of 480 sec is proposed to be the safe minimum level for anticoagulation during CPB. This study was designed to determine if the activated coagulation time (ACT) values of each analyser separately are repeatable, and to determine whether there exists a significant difference in ACT values measured by three different analysers: the GEM PCL (Instrumentation Laboratory), the Hemochron 801 (International Technidyne Corporation) and the ACT II Automated Coagulation Timer (Medtronic).
METHODS: All patients underwent cardiovascular surgical procedures requiring heparinisation (200-300 IU/kg). Blood samples for the measurement of the ACT were taken from all patients before and after heparinisation, during CPB, and after protamine administration. All samples were measured in duplicate with the three different analysers. To compare the activated clotting time data, the method described by Bland and Altman was used. The Pearson correlation coefficient was used to determine whether the differences were related to the average ACTs. p-Values <0.05 were considered statistically significant.
RESULTS: The results showed that the three tested ACT analysers met the requirements of repeatability. The mean differences and standard deviations of the ACT values measured with the GEM PCL, the Hemochron 801, and the ACT II analyser were, respectively, -8.78 +/- 37.61, -19.77 +/- 68.82, and -6.23 +/- 39.21, with p-values=0.177, 0.081 and 0.384, respectively. The Pearson correlation coefficients were too low (-0.012, -0.221 and -0.241, respectively) to show any correlation between the differences and the means. The ACT values measured with the Hemochron 801 were not significantly different from the ACT values measured with the ACT II analyser: deltaACT =-34.09 +/- 146.68, with p=0.132. However, the GEM PCL did not agree with the Hemochron 801: deltaACT= -80.2 +/- 143.06, with p=0.001, or the ACT II analyser: deltaACT= -119.13 +/- 138.51, with p<0.001. A rather strong correlation was evident between the differences and the means measured with the GEM PCL compared with the Hemochron 801 (r=0.68) and the ACT II analyser (r=0.76).
CONCLUSIONS: All analysers used celite or kaolin as activator. However, it was evident that the ACT measurements depended also on the analyser that had been chosen. A precaution that ACT values could not always be interpreted in the same way seems to be necessary.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16485696     DOI: 10.1191/0267659106pf844oa

Source DB:  PubMed          Journal:  Perfusion        ISSN: 0267-6591            Impact factor:   1.972


  7 in total

1.  On-chip titration of an anticoagulant argatroban and determination of the clotting time within whole blood or plasma using a plug-based microfluidic system.

Authors:  Helen Song; Hung-Wing Li; Matthew S Munson; Thuong G Van Ha; Rustem F Ismagilov
Journal:  Anal Chem       Date:  2006-07-15       Impact factor: 6.986

Review 2.  Point-of-care testing for anticoagulation monitoring in neuroendovascular procedures.

Authors:  H M Hussein; A L Georgiadis; A I Qureshi
Journal:  AJNR Am J Neuroradiol       Date:  2011-09-15       Impact factor: 3.825

3.  STS/SCA/AmSECT Clinical Practice Guidelines: Anticoagulation during Cardiopulmonary Bypass.

Authors:  Linda Shore-Lesserson; Robert A Baker; Victor Ferraris; Philip E Greilich; David Fitzgerald; Philip Roman; John Hammon
Journal:  J Extra Corpor Technol       Date:  2018-03

4.  Heparin therapy during extracorporeal circulation: deriving an optimal activated clotting time during cardiopulmonary bypass for isolated coronary artery bypass grafting.

Authors:  Kenneth Palmer; Tim Ridgway; Omar Al-Rawi; Michael Poullis
Journal:  J Extra Corpor Technol       Date:  2012-09

Review 5.  Uninterrupted DOACs Approach for Catheter Ablation of Atrial Fibrillation: Do DOACs Levels Matter?

Authors:  Michael Hardy; Jonathan Douxfils; Anne-Sophie Dincq; Anne-Laure Sennesael; Olivier Xhaet; Francois Mullier; Sarah Lessire
Journal:  Front Cardiovasc Med       Date:  2022-03-29

6.  Can We Rely on the Activated Clotting Time to Measure Heparin Anticoagulation? A Clinical Evaluation of Two ACT Monitors.

Authors:  Samuel Nilsson; Micael Appelblad; Staffan Svenmarker
Journal:  J Extra Corpor Technol       Date:  2020-09

7.  Point-of-care measurement of activated clotting time for cardiac surgery as measured by the Hemochron signature elite and the Abbott i-STAT: agreement, concordance, and clinical reliability.

Authors:  Daniel Dirkmann; Elisabeth Nagy; Martin W Britten; Jürgen Peters
Journal:  BMC Anesthesiol       Date:  2019-09-06       Impact factor: 2.217

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.