Repeatability of quantitative measurements of retinal layers with SD-OCT and agreement between vertical and horizontal scan protocols in healthy eyes.

Alberto Domínguez-Vicent1, Rune Brautaset1, Abinaya Priya Venkataraman1.   

Abstract

PURPOSE: To evaluate the repeatability of the new spectral domain optical coherence tomography (HOCT-1F), and also to evaluate the agreement between vertical and horizontal scan protocols. In addition, we also evaluated the relation between the repeatability and age.
METHODS: Three consecutive measurements of the inner limiting membrane-retinal pigment epithelium (ILM-RPE), inner limiting membrane-inner plexiform layer (ILM-IPL) from macular horizontal and vertical scans, and inner limiting membrane-retinal nerve fiber layer (ILM-RNFL) from optic disc horizontal scan. 159 subjects were included in the analysis. The within subject standard deviation (Sw) and the repeatability limits (Rlimit) are used to represent the repeatability of the parameters for the different sectors.
RESULTS: The Sw for the ILM-RPE thickness was less than 3.5 μm for each sector and scan direction. The Sw values varied within the sectors and scan modes, with horizontal scan modes resulting in better values for the horizontal sectors, and vice versa. The Sw for the GCL-IPL thickness was less than 2 μm, and was similar between the vertical and horizontal scan modes for each sector map. For the optic disc scan, the Sw was not symmetric along the clock-hour map sectors, the largest Sw values were seen in the vertical sectors (8.6 μm). The mean difference between the vertical and horizontal scans was less than 2 μm for each retinal thickness sector map. Significant but weak correlation between the Sw and the subject's age was seen in both macular and optic disc scans.
CONCLUSIONS: The repeatability of the HOCT-1F to measure the ILM-RPE-, ILM-IPL- and ILM-RNFL-thickness is good. The repeatability of the ILM-RPE thickness is dependent on the scan direction, which should be taken into account when calculating retinal thickness. There is a weak correlation between the repeatability and the subject's age.

Entities:  

Mesh:

Year:  2019        PMID: 31437222      PMCID: PMC6705867          DOI: 10.1371/journal.pone.0221466

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Optical coherence tomography (OCT) has an irreplaceable role in the field of ophthalmology. Both clinicians and researchers use OCT to envisage and understand retinal and choroidal pathologies[1-3]. Different types of OCT are available commercially and each type comes with its own automated segmentation algorithm. The segmentation provides useful quantitative information about each individual layer of the retina. The objective measurements of macular thickness, ganglion cell-inner plexiform layer (GCL-IPL) thickness and peripapillary retinal nerve fiber layer (pRNFL) thickness provides valuable information in the diagnosis and follow up of patients with glaucoma, macular diseases and neurological disorders[4-8]. In order to know how consistent the quantitative measurements of the retinal layers are, it is important to assess the repeatability of the measurements. The repeatability of the measurement is dependent on scan resolution, acquisition time, segmentation algorithm and the retinal condition[9-13]. Current generation OCT instruments provide better resolution, faster measurements and better segmentation algorithms which in turn provides us with reliable quantitative data. The HOCT-1F (Huvitz) is a new spectral domain OCT, which has an axial resolution of around 6 to 7 microns, and acquisition rate of 68,000 A-scans per second. The three-dimensional scan can be performed with various scan resolutions and with both vertical and horizontal scan lines, both with auto- and semi-auto scanning protocols. This study evaluates the repeatability of the HOCT-1F. We evaluated the repeatability of macular thickness, GCL-IPL thickness and pRNFL thickness. In addition, the agreement between vertical and horizontal scan protocols was also evaluated for macular thickness and GCL-IPL thickness. The relation between repeatability values with age was also evaluated.

Materials and methods

Study participants

The study participants were Caucasian volunteers above 18 years of age. The study adhered to the tenets of the Declaration of Helsinki and was approved by the Regional Ethical Committee (Regionala etikpröningsnämnden, Stockholm). Written informed consent was obtained after the nature and purpose of the study had been clearly explained to the participants. The study exclusion criteria were spherical refractive error more than ±5.0 D, astigmatic refractive error more than ±3.0 D, intraocular pressures more than 21 mmHg, best-corrected visual acuity worse than 0.0 logMAR for subjects under 70 years and 0.1 logMAR for subjects over 70 years of age, significant media opacities that cause poor imaging during OCT measurements, retinal disorders, and prior ocular surgeries except for cataract extraction. Participants under 70 years were excluded if they had known systemic disorders whereas no such criterion was set for participants above 70 years of age. The subjects underwent a complete ophthalmic examination, including BCVA, intraocular pressure with non-contact tonometry, axial length measurement with optical biometer (Lenstar 900, Haag-Streit), slit lamp biomicroscopy, undilated fundus photography and OCT measurements, along with a complete ocular and medical history.

OCT measurements

All participants underwent OCT imaging with the HOCT-1F. Only one eye (right eye) per subject was included in the analysis. A three-dimensional scan protocols composed of 512 A-scans for each of 96 B-scans was used for both macular and optic nerve head measurements. Macular scans were performed with both horizontal and vertical scanning protocols covering a 9x9 mm area centered on the fovea. For the optic nerve head scans, a horizontal scanning protocol covering 6x6 mm area centered on the optic nerve head was used. All scans were repeated 3 times with adequate breaks in between. The duration of the 9 consecutive measurements were less than 10 minutes. The scans were repeated if the fixation was poor or if the subject blinked during the measurements. Scans with signal strength less than 6 were excluded. All OCT measurements were performed by two experienced examiners. 179 subjects were recruited, and 159 subjects met the inclusion criteria and were included in the analysis. All the thickness measures were obtained using the automated segmentation algorithms from the OCT instrument. No manual adjustment of the segmentation was performed. For the macular full thickness (from inner limiting membrane to retinal pigment epithelium, ILM-RPE) evaluations, 9 sectors based on Early Treatment Diabetic Retinopathy Study (ETDRS) protocol with the central circle, inner ring (Superior, Nasal, Inferior and temporal), and outer ring (Superior, Nasal, Inferior and temporal) using 1, 3, and 6 mm diameters were used. GCL-IPL layer thickness (from inner limiting membrane to inner plexiform layer, ILM-IPL) was evaluated in 6 sectors (Superior, superior nasal, inferior nasal, inferior, inferior temporal and superior temporal) with an inner and outer circle of diameter 1, and 4 mm centered on the fovea. The pRNFL thickness (from inner limiting membrane to retinal nerve fiber layer, ILM-RNFL) was evaluated in twelve clock-hour sectors in a circle of 3.45 mm diameter centered at the optic disc. The measurement locations for the macular and the optic disc scan is shown in Fig 1.
Fig 1

Schematic representation of the sectors for the retinal layer thickness.

In A ILM-RPE represents inner limiting membrane to retinal pigment epithelium thickness and numbers 1 to 9 indicate Center, inner superior, inner nasal, inner inferior, inner temporal, outer superior, outer nasal, outer inferior and outer temporal sectors. In B, ILM-IPL represents inner limiting membrane to inner plexiform layer thickness and numbers 1to 6 indicate superior, superior nasal, inferior nasal, inferior, inferior temporal and superior temporal sectors. In C, ILM-RNFL represents inner limiting membrane to retinal nerve fiber layer thickness and the numbers represent the clock hour position of the sectors. A and B are from the macular scans and C is from the optic disc scan.

Schematic representation of the sectors for the retinal layer thickness.

In A ILM-RPE represents inner limiting membrane to retinal pigment epithelium thickness and numbers 1 to 9 indicate Center, inner superior, inner nasal, inner inferior, inner temporal, outer superior, outer nasal, outer inferior and outer temporal sectors. In B, ILM-IPL represents inner limiting membrane to inner plexiform layer thickness and numbers 1to 6 indicate superior, superior nasal, inferior nasal, inferior, inferior temporal and superior temporal sectors. In C, ILM-RNFL represents inner limiting membrane to retinal nerve fiber layer thickness and the numbers represent the clock hour position of the sectors. A and B are from the macular scans and C is from the optic disc scan.

Statistical analysis

Descriptive statistics were used to summarize the baseline demographics of the observations and participants included. The within subject standard deviation (Sw) and repeatability limits (Rlimit) were used to describe the repeatability of the HOCT-1F to measure the ILM-RPE thickness and the ILM-IPL thickness from the macular scans, and the ILM-RNFL thickness from the optic disc scan. The Sw, which represents the repeatability of the measurements, was calculated with a one-way analysis of variance with the subject as a factor[14]. The repeatability limit was calculated as , and it represents the expected limits that 95% of the measurements should be within. The agreement between the horizontal and vertical scans was done with the Bland-Altman test[15] for repeated measurements for the ILM-RPE thickness and the ILM-IPL thickness from the macular scans. Finally, the Pearson correlation coefficient was used to calculate the correlation between the repeatability of the parameters and the participant’s age. The statistical significance limit was set to a p-value < 0.05.

Results

The mean age of the 159 participants was 49.58 ± 17.01 years old [18–83 years], and the number of men and women were almost the same (77 men and 82 women). The mean axial length and spherical equivalent of the participants were 23.89 ± 0.98 mm and -0.31 ± 1.90 D, respectively. Table 1 shows the average thickness values for the ILM-RPE and GCL-IPL thickness for each sector from horizontal and vertical macular scan modalities. Within the ILM-RPE thicknesses, the thinnest value was obtained at the central sector (about 258 μm), and the thickest thicknesses were measured at the inner circle (about 320 μm) in both scan modalities. Regarding the ILM-IPL thicknesses, the temporal sectors showed about 13-μm thinner values than the other four quadrants in both scan modalities. Table 2 shows the average thickness values for the ILM-RNFL for 12 sectors from the optic disc horizontal scan. At the clock-hour map, the thicknesses among the superior and inferior sectors (clock position 11–1 and 5–7) were about 30 μm thicker than at the nasal and temporal sectors (clock position 2–4 and 8–10).
Table 1

Thickness values for different sectors from horizontal and vertical macular scans.

Horizontal scanVertical scan
Average ± STDMinimumMaximumAverage ± STDMinimumMaximum
ILM–RPECentral257.5 ± 18.8212312259.2 ± 19.2212306
Inner circleSuperior322.6 ± 16.2277364323.8 ± 16.0277364
Nasal324.9 ± 16.2270363324.6 ± 16.9276363
Inferior320.3 ± 15.5281364321.5 ± 15.8274365
Temporal311.9 ± 14.8271353311.9 ± 14.9264353
Outer circleSuperior273.4 ± 13.4240311274.5 ± 13.6243318
Nasal295.6 ± 16.6258338294.6 ± 16.1255336
Inferior265.2 ± 14.3229311267.0 ± 13.8231310
Temporal259.5 ± 13.5221301259.3 ± 14.0223302
ILM–IPLSuperior116.5 ± 9.783143117.7 ± 9.781144
Nasal superior119.5 ± 9.785144119.5 ± 9.882144
Nasal inferior119.0 ± 10.481143119.2 ± 10.482142
Inferior116.3 ± 9.589142117.7 ± 9.789144
Temporal inferior106.3 ± 8.380129107.1 ± 8.579129
Temporal superior103.3 ± 8.075126104.0 ± 8.374126

ILM-RPE: Inner limiting membrane to retinal pigment epithelium ILM-IPL: Inner limiting membrane to inner plexiform layer, Ganglion cell-inner plexiform layer.

STD: Standard deviation.

All values are expressed in microns.

Table 2

Retinal nerve fibre layer thickness values for different sectors from horizontal optic disc scans.

SectorAverage ± STDMinimumMaximum
12110.9 ± 23.439184
1101.6 ± 20.538173
297.2 ± 21.338145
365.9 ± 14.833109
477.6 ± 15.728126
5100.5 ± 19.228154
6127.3 ± 21.559184
7133.3 ± 18.183180
875.7 ± 14.340119
959.6 ± 8.8137102
1085.6 ± 14.543164
11130.4 ± 19.656172

STD: Standard deviation.

All values are expressed in microns.

ILM-RPE: Inner limiting membrane to retinal pigment epithelium ILM-IPL: Inner limiting membrane to inner plexiform layer, Ganglion cell-inner plexiform layer. STD: Standard deviation. All values are expressed in microns. STD: Standard deviation. All values are expressed in microns.

Intra-device repeatability

The intra-device repeatability of three consecutive measurements of the ILM-RPE thickness is shown in Fig 2. Overall, the repeatability of the measurements for the ILM-RPE thickness map was less than 3.5 μm for each sector and scan direction. The Sw for the Nasal and Temporal sectors was about 0.8 and 1.5 μm less with the horizontal scan mode than with the vertical one. Furthermore, the smallest Rlimits were obtained with the horizontal scan mode, which was about 2.5 μm less than the vertical scan mode. The opposite tendency was obtained for the Superior and Inferior sectors, in which the Sw and Rlimit values with the Horizontal scan mode were about 1- and 2 μm more than the Vertical mode, respectively. Finally, the Sw for the central sector was 2.8 and 2.4 μm with the Horizontal and Vertical scan modes, respectively. The Rlimit for this sector was about 7 μm for both scan modes.
Fig 2

Intra-device repeatability for the inner limiting membrane to retinal pigment epithelium thickness in different sectors for horizontal and vertical scans.

OT: outer temporal, IT: inner temporal, IN: inner nasal, ON: outer nasal, C: central circle, OS: outer superior, IS: inner superior, II: inner inferior, OI: outer inferior sectors.

Intra-device repeatability for the inner limiting membrane to retinal pigment epithelium thickness in different sectors for horizontal and vertical scans.

OT: outer temporal, IT: inner temporal, IN: inner nasal, ON: outer nasal, C: central circle, OS: outer superior, IS: inner superior, II: inner inferior, OI: outer inferior sectors. The intra-device repeatability of three consecutive measurements of the GCL-IPL thickness is shown in Fig 3. The Sw and Rlimit were similar between the vertical and horizontal scan modes for each sector map. The Sw ranged from 1.4 to 1.9 μm, and the Rlimit ranged from 3.9 to 5.3 μm.
Fig 3

Intra-device repeatability for the inner limiting membrane to inner plexiform layer thickness in different sectors for horizontal and vertical scans.

S: superior, SN: superior nasal, IN: inferior nasal, I: inferior, IT: inferior temporal and ST: superior temporal sectors.

Intra-device repeatability for the inner limiting membrane to inner plexiform layer thickness in different sectors for horizontal and vertical scans.

S: superior, SN: superior nasal, IN: inferior nasal, I: inferior, IT: inferior temporal and ST: superior temporal sectors. Fig 4 shows a polar plot with the repeatability of the three measurements obtained for the RNFL thickness for each clock-hour sector. In this plot, the central small black points represent the Sw, and the grey circles surrounding each black point represents the Rlimit of each Sw. The size of these circles was scaled for visualization purpose to facilitate the comparison of the Rlimit among sectors. Overall, the repeatability was not symmetric along the clock-hour map sectors. As it can be seen, the repeatability was low for the vertical sectors, and the best repeatability values were obtained for the horizontal sectors. Concretely, the Sw at the 12- and 6 clock-hour sectors were above 6.5 μm, whereas it is as low as 2.3 μm on the 9 clock-hour sector. Similarly, the Rlimit was also not symmetric, with minimum values at the horizontal sectors and maximum values at the vertical ones. Concretely, the maximum Rlimit was obtained at the 12 clock-hour sector (23.3 μm), and the minimum Rlimit was obtained at the 9 clock-hour sector (6.5 μm).
Fig 4

Intra-device repeatability for the inner limiting membrane to retinal nerve fiber layer thickness in different clock hour sectors for horizontal scans.

The small black dots represent the within subject standard deviation and the grey circles surrounding each black point represent the repeatability limits, where the areas are scaled by a factor of 30 points for visualization purposes.

Intra-device repeatability for the inner limiting membrane to retinal nerve fiber layer thickness in different clock hour sectors for horizontal scans.

The small black dots represent the within subject standard deviation and the grey circles surrounding each black point represent the repeatability limits, where the areas are scaled by a factor of 30 points for visualization purposes.

Agreement between horizontal and vertical scans

Table 3 shows the agreement between the horizontal and vertical scans for the ILM-RPE and ILM-IPL thicknesses. The mean difference between the vertical and horizontal scans was less than 2 μm for each retinal thickness sector map. The mean difference for the horizontal sectors (Nasal and Temporal) was about 1 μm less than for the vertical ones (Superior and Inferior). Nevertheless, the maximum variability between the horizontal and vertical scan directions was between 9 μm and 17 μm for the inner and outer circle sectors. The vertical scan measured on average 1.67 μm thinner than the horizontal scan in the central sector, and the maximum variability between both scan directions was 20.13 μm.
Table 3

Comparison between the horizontal and vertical macular scan.

Map sectorsMean difference [Limits of Agreement]
ILM–RPECentral-1.67 [-11.77–8.43]
InnercircleSuperior-1.24 [-6.27–3.78]
Nasal0.25 [-7.76–8.26]
Inferior-1.23 [-9.09–6.63]
Temporal-0.03 [-4.81–4.75]
OutercircleSuperior-1.07 [-8.25–6.10]
Nasal0.97 [-5.78–7.73]
Inferior-1.83 [-8.65–4.98]
Temporal0.24 [-8.40–8.88]
ILM–IPLSuperior-1.27 [-6.92–4.38]
Nasal superior0.00 [-5.65–5.64]
Nasal inferior-0.20 [-6.33–5.94]
Inferior-1.39 [-6.61–3.83]
Temporal inferior-0.81 [-5.92–4.31]
Temporal superior-0.64 [-5.83–4.55]

ILM-RPE: Inner limiting membrane to retinal pigment epithelium.

ILM-IPL: Inner limiting membrane to inner plexiform layer.

All values are expressed in microns.

ILM-RPE: Inner limiting membrane to retinal pigment epithelium. ILM-IPL: Inner limiting membrane to inner plexiform layer. All values are expressed in microns. With regard to the ILM-IPL thickness, the mean difference between the vertical and horizontal scans was less 2 μm for each sector. The mean differences for the horizontal sectors (Nasal superior, Nasal inferior, Temporal superior, and Temporal inferior) was less than for the vertical sectors (Superior and Inferior). Considering the limits of agreement, the maximum difference between both scan directions could be up to 12.3 μm.

Repeatability as a function of the subject’s age

Tables 4 and 5 show the correlation values between the Sw and the participant’s age for the macular and optic disc maps, respectively. Although the correlations between the Sw and the age were statistically significant (P < 0.05) for some sectors, the correlation was weak in all sectors for both macular and optic disc maps.
Table 4

Correlation values between the repeatability and age for the macular scans.

Horizontal scanVertical scan
ILM–RPECentral0.251*0.010
Inner circleSuperior0.127-0.015
Nasal0.0290.004
Inferior0.0680.059
Temporal0.0560.153
Outer circleSuperior0.138-0.016
Nasal0.1870.158*
Inferior0.1450.012
Temporal0.1620.174*
ILM–IPLSuperior0.250*0.033
Nasal superior0.216*0.082
Nasal inferior0.0610.007
Inferior0.243*0.178*
Temporal inferior0.0760.052
Temporal superior0.1350.059

ILM-RPE: Inner limiting membrane to retinal pigment epithelium

ILM-IPL: Inner limiting membrane to inner plexiform layer, Ganglion cell-inner plexiform layer.

* Indicates p<0.05

Table 5

Correlation values between the retinal nerve fibre layer thickness repeatability and age for the optic disc scans.

SectorsHorizontal scan
120.161*
10.316
20.178*
30.121
40.340*
50.192*
60.325*
70.222*
80.093
90.078
100.301*
110.300*

* Indicates p<0.05

ILM-RPE: Inner limiting membrane to retinal pigment epithelium ILM-IPL: Inner limiting membrane to inner plexiform layer, Ganglion cell-inner plexiform layer. * Indicates p<0.05 * Indicates p<0.05

Discussion

The aim of this study was to assess the repeatability of the Huvitz HOCT-1F system to measure the retinal-, GCL/IPL, and RNFL thicknesses. Based on the results of this study, clinicians will be able to set the protocols for OCT scans. The intra-device repeatability of each ILM-RPE sector map (Fig 2) varied with the scan direction (Horizontal or Vertical). The best repeatability outcomes for the horizontal sectors (Nasal and Temporal) were obtained with the horizontal scan direction. On the other hand, the best repeatability values for the vertical sectors (Superior and Inferior) were obtained with the vertical scan direction. Whereas, the repeatability for the central sector was similar with both scan directions. The maximum Sw and Rlimit (3.339 and 9.248 μm, respectively) were seen for the temporal sector in the outer circle from the vertical scan. Nevertheless, the maximum Sw reported here is less than the axial resolution (6–7 μm) of the HOCT-1F instrument. The Rlimit was also less than the axial resolution of the instrument in 5 out of 9 sectors for the horizontal scan, and 6 out of 9 sectors for the vertical scan. The intra-device repeatability of each ILM-IPL sector map (Fig 3) was similar for both scan direction (Horizontal or Vertical). Both Sw and Rlimit were less than the axial resolution of the instrument for the same measurement conditions used this study. Several previous studies have also reported good repeatability for both ILM-RPE and ILM-IPL [11,13,16-19]. All in all, most of these studies assessed the repeatability of different SD-OCT to measure the GCL/IPL and RNFL thicknesses. Some of the studies reported the repeatability using the intraclass correlation coefficient, which values were greater than 0.9 for the GCL/IPL and greater than 0.8 for the RNFL in each sector thickness map [13,17,18]. Another study described the repeatability of six SD-OCTs to measure the central retinal thickness using the coefficient of variation, and this value ranged from 0.46 to 3.50 [16]. Finally, Ctori et al. assessed the repeatability of a SD-OCT to measure each retinal layer, and the coefficient of repeatability ranged from 0.31 (retinal thickness) to 1.73 (outer nuclear layer) [19]. Since different metrics have been used to describe the repeatability of a specific SD-OCT, direct comparisons among studies are difficult to make. Besides all this, it should be taken into account that these results evidence that the repeatability of the automatic segmentation algorithms available in the SD-OCT is good. Fig 5 show a colour coded map of the intra-device repeatability for the macula scans, where Rlimit values more than the instrument’s axial resolution are marked in red. A previous study reported that there is no variation in the repeatability values between horizontal and vertical scans based on analyses from 20 healthy eyes and 20 eyes of patients with multiple sclerosis[20]. However, that study did not evaluate sector wise the repeatability as it is done in the present study. We believe that the variation in the repeatability can be seen in a sector wise analysis rather than the analysis of the whole scanned area. This variation could be a result of the influence of the retinal blood vessel in the segmentation algorithm. The B-scan used for the calculation of Nasal and Temporal sectors could pass through less number of blood vessels in horizontal scan mode compared to the B-scan from the vertical scan mode, and vice versa for the vertical sectors. From our results, it can be recommended to use vertical scans for vertical sectors, and horizontal scans for horizontal sectors. In other words, the accuracy of the quantitative measurements with OCT can be improved by performing both horizontal and vertical scans consecutively, and combining the information from both scans in the calculations of the retinal thickness.
Fig 5

Colour coded map of the intra-device repeatability for the macula scans in different sectors.

Green: repeatability limit values within the instrument’s axial resolution, Red: repeatability limit values above the instrument’s axial resolution. ILM-RPE: inner limiting membrane to retinal pigment epithelium thickness ILM-IPL: inner limiting membrane to inner plexiform layer thickness.

Colour coded map of the intra-device repeatability for the macula scans in different sectors.

Green: repeatability limit values within the instrument’s axial resolution, Red: repeatability limit values above the instrument’s axial resolution. ILM-RPE: inner limiting membrane to retinal pigment epithelium thickness ILM-IPL: inner limiting membrane to inner plexiform layer thickness. The intra-device repeatability of the ILM-RNFL from the optic disc scan varied with the clock-hour sector (Fig 4). The repeatability was worse for the vertical sectors compared to the horizontal sectors. In comparison to the repeatability of the parameters from the macular scans, the optic disc scan showed lower repeatability as shown by larger Sw. This finding is not surprising as the influence of the blood vessels are more in vertical clock-hour positions compared to the horizontal clock-hour positions[21-23]. Previous studies have also reported the same trend. With the agreement analysis, it can be seen that on average, the difference between horizontal and vertical scans is minimal for both ILM-RPE and ILM-IPL thicknesses (Table 3). However, large variabilities in the differences are seen in the ILM-RPE thickness in the outer circle and centre. The variabilities in the differences are similar between the ILM-RPE thickness in the inner-circle and the ILM-IPL thickness. It is almost the same retinal location that it is used in the analysis of the inner circle of the ILM-RPE and ILM-IPL. This area does not have that many blood vessels as the outer circle of the ILM-RPE analysis. This reasoning does not explain the large variability in the difference of the central ILM-RPE thickness as there are no blood vessels in the centre. The possible reason for this could be that the central values are critically dependent on the scan centration[24]. The retina undergoes significant changes as a part of the normal aging process[25,26]. It would be interesting to know how the repeatability of OCT parameters varies with age. There is a significant correlation between the repeatability and the subject’s age in some of the sectors of the macular and optic disc scans (Tables 4 and 5), however the correlation coefficient values were weak (between 0.15 and 0.34). Liu et al. showed that age is not one of the factors that increase the prevalence of artefacts in OCT imaging[27]. However, it could be difficult for older subjects to fixate during OCT measurements, but the faster acquisition time and better scan algorithms make it possible to have good repeatability independent of the subject’s age. According to the maximum Sw that we got for the ILM-RPE, ILM-IPL, and ILM-RNFL thicknesses, we calculated the measurement tolerance, MT (MT = (1.96∙Sw)/√N) for N number of measurements according to the ISO standards[28,29]. For 1, 2, and 3 number of measurements, the measurement tolerance for the ILM-RPE is 6.54, 4.63, and 3.78 μm, respectively. For the ILM-IPL, the measurement tolerances were 3.72, 2.63, and 2.15 μm, respectively. For the ILM-RNFL from the optic disc scan, the measurement tolerances were 16.77, 11.86, and 9.68 μm, respectively. For the macula thickness measurements in retinas with no morphological changes, one three-dimensional scan protocol composed of 512 A-scans for each of 96 B-scans is enough to have a tolerance within the axial resolution of the instrument. However, when it comes to the RNFL thickness from the peripapillary scans, more than 3 scans with the above set parameters are needed to ensure a tolerance within the axial resolution of the instrument. However, the number of scans required to achieve these tolerance limits can vary for retinas with morphological changes. Further studies can also consider the repeatability analyses of different scans protocols, in both healthy as well as pathological retinas. A previous study has shown that the repeatability of automatic segmentation algorithms in pathologic eyes is worse than in normal eyes [30]. However, the repeatability improved up to the level of healthy eyes when manual corrections of the segmentation were performed. As a consequence, the intrinsic variability in manual measurements might affect the reproducibility of the measurements in pathologic eyes. The present study included only healthy eyes to assess the performance of HOCT-1F but further studies are recommended to assess the performance in pathologic eyes. On the other hand, several studies reported a positive relationship between the signal strength and the pRNFL thickness [31,32]. This relationship indicates that low signal strength is associated with thin pRNFL thickness. As the present study included scans with a good signal strength only, further studies can be performed to assess the correlation between the signal strength and the repeatability. In conclusion, the repeatability of the Huvitz HOCT-1F to measure the ILM-RPE thickness, ILM-IPL thickness and ILM-RNFL thickness is good. The repeatability of the ILM-RPE thickness is dependent on the scan direction, and this factor should be taken into account while choosing scan protocols. A significant but weak correlation is seen between the repeatability of the HOCT-1F and the subject’s age.

RawData.

(XLS) Click here for additional data file.
  30 in total

Review 1.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

2.  Retinal thickness decreases with age: an OCT study.

Authors:  B Alamouti; J Funk
Journal:  Br J Ophthalmol       Date:  2003-07       Impact factor: 4.638

3.  Comparison of retinal nerve fiber layer thickness in normal eyes using time-domain and spectral-domain optical coherence tomography.

Authors:  Leonard K Seibold; Naresh Mandava; Malik Y Kahook
Journal:  Am J Ophthalmol       Date:  2010-12       Impact factor: 5.258

4.  Relationship between retinal nerve fiber layer measurement and signal strength in optical coherence tomography.

Authors:  Carol Yim Lui Cheung; Christopher Kai Shun Leung; Dusheung Lin; Chi-Pui Pang; Dennis Shun Chiu Lam
Journal:  Ophthalmology       Date:  2008-02-21       Impact factor: 12.079

5.  Is quantitative spectral-domain superior to time-domain optical coherence tomography (OCT) in eyes with age-related macular degeneration?

Authors:  Urban Eriksson; Albert Alm; Eva Larsson
Journal:  Acta Ophthalmol       Date:  2011-03-04       Impact factor: 3.761

6.  Reproducibility of macular thickness and volume using Zeiss optical coherence tomography in patients with diabetic macular edema.

Authors:  Magdalena G Krzystolik; Samara F Strauber; Lloyd Paul Aiello; Roy W Beck; Brian B Berger; Neil M Bressler; David J Browning; Robert B Chambers; Ronald P Danis; Matthew D Davis; Adam R Glassman; Victor H Gonzalez; Paul B Greenberg; Jeffrey G Gross; Judy E Kim; Craig Kollman
Journal:  Ophthalmology       Date:  2007-03-13       Impact factor: 12.079

7.  Macular thickness measurements in healthy eyes using six different optical coherence tomography instruments.

Authors:  Ute E K Wolf-Schnurrbusch; Lala Ceklic; Christian K Brinkmann; Milko E Iliev; Manuel Frey; Simon P Rothenbuehler; Volker Enzmann; Sebastian Wolf
Journal:  Invest Ophthalmol Vis Sci       Date:  2009-02-21       Impact factor: 4.799

8.  Blood vessel contributions to retinal nerve fiber layer thickness profiles measured with optical coherence tomography.

Authors:  Donald C Hood; Brad Fortune; Stella N Arthur; Danli Xing; Jennifer A Salant; Robert Ritch; Jeffrey M Liebmann
Journal:  J Glaucoma       Date:  2008 Oct-Nov       Impact factor: 2.503

9.  Factors associated with variability in retinal nerve fiber layer thickness measurements obtained by optical coherence tomography.

Authors:  Ziqiang Wu; Matti Vazeen; Rohit Varma; Vikas Chopra; Alexander C Walsh; Laurie D LaBree; Srinivas R Sadda
Journal:  Ophthalmology       Date:  2007-03-23       Impact factor: 12.079

10.  Retinal nerve fiber layer imaging with spectral-domain optical coherence tomography: a variability and diagnostic performance study.

Authors:  Christopher Kai-Shun Leung; Carol Yim-Lui Cheung; Robert N Weinreb; Quanliang Qiu; Shu Liu; Haitao Li; Guihua Xu; Ning Fan; Lina Huang; Chi-Pui Pang; Dennis Shun Chiu Lam
Journal:  Ophthalmology       Date:  2009-05-22       Impact factor: 12.079

View more
  6 in total

1.  Central Macular Thickness in a Healthy Mexican Population Using Huvitz Optical Coherence Tomography.

Authors:  Renata García-Franco; Dalia Méndez-Marín; Marlon García-Roa; Paulina Ramirez-Neria; Diego Valera-Cornejo; Van Charles Lansingh
Journal:  Clin Ophthalmol       Date:  2020-11-17

2.  OCT Variability Prevents Their Use as Robust Biomarkers in Multiple Sclerosis.

Authors:  Marta Para-Prieto; Raul Martin; Sara Crespo; Laura Mena-Garcia; Andres Valisena; Lisandro Cordero; Gloria Gonzalez Fernandez; Juan F Arenillas; Nieves Tellez; Jose Carlos Pastor
Journal:  Clin Ophthalmol       Date:  2021-05-14

3.  Agreement of different OCT scan directions for individual retinal-layer thickness measurements in multiple sclerosis subjects with prior unilateral optic neuritis.

Authors:  Alberto Domínguez-Vicent; Maria Nilsson; Rune Brautaset; Abinaya Priya Venkataraman
Journal:  Sci Rep       Date:  2022-01-12       Impact factor: 4.379

4.  Precision and Agreement of Individual and Simultaneous Macular and Optic Disc Volumetric Measurements With Spectral Domain Optical Coherence Tomography.

Authors:  Alberto Domínguez-Vicent; Jesper Kensén; Marika Wahlberg Ramsay; Rune Brautaset; Abinaya Priya Venkataraman
Journal:  Front Med (Lausanne)       Date:  2021-11-25

5.  Impact of optical coherence tomography scan direction on the reliability of peripapillary retinal nerve fiber layer measurements.

Authors:  Abinaya Priya Venkataraman; Josefine Andersson; Lina Fivelsdal; Maria Nilsson; Alberto Domínguez-Vicent
Journal:  PLoS One       Date:  2021-02-22       Impact factor: 3.240

6.  Effects of uncomplicated Descemet membrane endothelial keratoplasty on the central retinal thickness.

Authors:  Tibor Lohmann; Sabine Baumgarten; Niklas Plange; Peter Walter; Matthias Fuest
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2021-05-11       Impact factor: 3.117

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.