Literature DB >> 21371288

Is quantitative spectral-domain superior to time-domain optical coherence tomography (OCT) in eyes with age-related macular degeneration?

Urban Eriksson1, Albert Alm, Eva Larsson.   

Abstract

PURPOSE: The aim of this study was to determine the variability of macular map measurements, for two generations of optical coherence tomography (OCT) instruments, in eyes with wet age related macular degeneration (AMD) and low visual acuity.
METHODS: Patients were examined with Stratus OCT and Cirrus HD-OCT. The macular thickness was assessed with the 'macular thickness map scan' and 'fast protocol' in Stratus and with the 512 × 128 and 200 × 200 cube protocols in Cirrus OCT. Two measurements were taken one directly after the other, at the first visit to analyse repeatability. Approximately 1 week later, a third measurement was taken to analyse reproducibility. In Cirrus OCT, a manual correction of foveal location was also performed. Repeatability and reproducibility were calculated as a coefficient of variance (CoV) and a coefficient of repeatability/reproducibility.
RESULTS: Repeatability for central macular thickness (expressed as CoV) was about three per cent for all protocols, and the coefficient of repeatability between 34 and 54 μm. Reproducibility (also expressed as CoV) was between four to seven per cent and coefficient of repeatability between 64 and 89 μm. After manual adjustment of foveal location in Cirrus OCT, the coefficient of repeatability improved to 12-18 μm, and the coefficient of reproducibility to 44-47 μm.
CONCLUSIONS: In eyes affected by wet AMD, there were small differences in repeatability and reproducibility when comparing quantitative maps in Stratus and Cirrus OCT. However, when the software for manual correction of foveal position in Cirrus OCT was used, the variability decreased markedly, and the repeatability was close to what had been reported in normal eyes, demonstrating a significant, potential advantage of spectral-domain over time-domain OCT.
© 2011 The Authors. Acta Ophthalmologica © 2011 Acta Ophthalmologica Scandinavica Foundation.

Entities:  

Mesh:

Year:  2011        PMID: 21371288     DOI: 10.1111/j.1755-3768.2011.02112.x

Source DB:  PubMed          Journal:  Acta Ophthalmol        ISSN: 1755-375X            Impact factor:   3.761


  9 in total

1.  Effect of optical coherence tomography scan decentration on macular center subfield thickness measurements.

Authors:  Jeong W Pak; Ashwini Narkar; Sapna Gangaputra; Ronald Klein; Barbara Klein; Stacy Meuer; Yijun Huang; Ronald P Danis
Journal:  Invest Ophthalmol Vis Sci       Date:  2013-07-02       Impact factor: 4.799

2.  Therapeutic efficacy of bevacizumab for age-related macular degeneration: what are the implications of CATT for routine management?

Authors:  Focke Ziemssen; Bianka Sobolewska
Journal:  Drugs Aging       Date:  2011-11-01       Impact factor: 3.923

3.  COMPARISON OF SPECTRAL DOMAIN OPTICAL COHERENCE TOMOGRAPHY SCAN PATTERNS AND CLINICAL REVIEW STRATEGIES IN NEOVASCULAR AGE-RELATED MACULAR DEGENERATION.

Authors:  Robert J Courtney; Jedediah I McClintic; Justis P Ehlers
Journal:  Retina       Date:  2015-07       Impact factor: 4.256

4.  Manual Segmentation of 12 Layers of the Retina and Choroid through SD-OCT in Intermediate AMD: Repeatability and Reproducibility.

Authors:  Pedro Camacho; Marco Dutra-Medeiros; Luís Salgueiro; Sílvia Sadio; Paulo C Rosa
Journal:  J Ophthalmic Vis Res       Date:  2021-07-29

5.  Glaucoma Diagnosis and Monitoring Using Advanced Imaging Technologies.

Authors:  Mitra Sehi; Shawn M Iverson
Journal:  US Ophthalmic Rev       Date:  2013

6.  Comparison of optical coherence tomography assessments in the comparison of age-related macular degeneration treatments trials.

Authors:  Francisco A Folgar; Glenn J Jaffe; Gui-Shuang Ying; Maureen G Maguire; Cynthia A Toth
Journal:  Ophthalmology       Date:  2014-05-15       Impact factor: 12.079

7.  Spectral domain OCT versus time domain OCT in the evaluation of macular features related to wet age-related macular degeneration.

Authors:  Luisa Pierro; Elena Zampedri; Paolo Milani; Marco Gagliardi; Vincenzo Isola; Alfredo Pece
Journal:  Clin Ophthalmol       Date:  2012-02-09

8.  Repeatability of quantitative measurements of retinal layers with SD-OCT and agreement between vertical and horizontal scan protocols in healthy eyes.

Authors:  Alberto Domínguez-Vicent; Rune Brautaset; Abinaya Priya Venkataraman
Journal:  PLoS One       Date:  2019-08-22       Impact factor: 3.240

9.  Is spectral-domain optical coherence tomography essential for flexible treatment regimens with ranibizumab for neovascular age-related macular degeneration?

Authors:  Abdullah Ozkaya; Zeynep Alkin; Hande Mefkure Ozkaya; Alper Agca; Engin Bilge Ozgurhan; Yalcin Karakucuk; Ahmet Taylan Yazici; Ahmet Demirok
Journal:  J Ophthalmol       Date:  2013-11-11       Impact factor: 1.909

  9 in total