Retinal nerve fiber layer thickness reproducibility using seven different OCT instruments.

Luisa Pierro1, Marco Gagliardi, Lorenzo Iuliano, Alessandro Ambrosi, Francesco Bandello.   

Abstract

PURPOSE: The clinical utility of new optical coherence tomography (OCT) instruments strongly depends on measurements reproducibility. The aim of this study was to assess retinal nerve fiber layer (RNFL) thickness reproducibility using six different spectral-domain OCTs (SD-OCTs) and one time-domain OCT.
METHODS: RNFL thickness (average and four quadrant) from six SD-OCTs (Spectral OCT/SLO OPKO/OTI, 3D-OCT 2000 Topcon, RS-3000 NIDEK, Cirrus HD-OCT Zeiss, RTVue-100 Optovue, and Spectralis Heidelberg) and one time-domain OCT (Stratus OCT Zeiss) was measured twice in 38 right eyes of 38 randomly chosen healthy volunteers by two masked operators. Inter- and intraoperator reproducibility was evaluated by the intraclass correlation coefficient (ICC), coefficient of variation (CV), and Bland-Altman test analysis. Instrument-to-instrument reproducibility was determined by ANOVA for repeated measures. We also tested how the devices disagree in terms of systemic bias and random error using a structural equation model.
RESULTS: Mean RNFL average thickness ranged from 90.08 μm to 106.51 μm. Cirrus and Heidelberg showed the thinnest RNFL values in all measurements, Topcon the highest. ICC, CV, and Bland-Altman plots showed variable inter- and intraoperator agreement depending on the instrument. Heidelberg demonstrated the best interoperator (ICC, 0.92; CV, 1.56%) and intraoperator (ICC, 0.94 and 0.95; CV, 1.28% and 1.26%, respectively, for operator A and operator B) agreement for average RNFL thickness.
CONCLUSIONS: Heidelberg demonstrated the higher agreement in inter- and intraoperator reproducibility, Optovue the worst. In light of our error analysis results, we found that a scale bias among instruments could interfere with a thorough RNFL monitoring, suggesting that best monitoring is obtained with the same operator and the same device.

Entities:  

Mesh:

Year:  2012        PMID: 22871835     DOI: 10.1167/iovs.11-8644

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  55 in total

1.  OCT Glaucoma Staging System: a new method for retinal nerve fiber layer damage classification using spectral-domain OCT.

Authors:  P Brusini
Journal:  Eye (Lond)       Date:  2017-08-04       Impact factor: 3.775

2.  Neuroretinal alterations in the early stages of diabetic retinopathy in patients with type 2 diabetes mellitus.

Authors:  P Carpineto; L Toto; R Aloia; V Ciciarelli; E Borrelli; E Vitacolonna; M Di Nicola; L Di Antonio; R Mastropasqua
Journal:  Eye (Lond)       Date:  2016-02-12       Impact factor: 3.775

3.  Comparison of retinal nerve fiber layer and macular thickness measurements with Stratus OCT and OPKO/OTI OCT devices in healthy subjects.

Authors:  Ahmet Ozkok; Julide Canan Umurhan Akkan; Nevbahar Tamcelik; Mehmet Erdogan; Didar Ucar Comlekoglu; Rengin Yildirim
Journal:  Int J Ophthalmol       Date:  2015-02-18       Impact factor: 1.779

4.  Exploring the role of macular thickness as a potential early biomarker of neurodegeneration in acute schizophrenia.

Authors:  Sukanto Sarkar; A R Rajalakshmi; S Avudaiappan; S Eswaran
Journal:  Int Ophthalmol       Date:  2021-04-15       Impact factor: 2.031

5.  Effectiveness of averaging strategies to reduce variance in retinal nerve fibre layer thickness measurements using spectral-domain optical coherence tomography.

Authors:  Berthold Pemp; Randy H Kardon; Karl Kircher; Elisabeth Pernicka; Ursula Schmidt-Erfurth; Andreas Reitner
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-04-16       Impact factor: 3.117

6.  Visual Outcomes, Visual Fields, and Optical Coherence Tomography in Paediatric Craniopharyngioma.

Authors:  Soraya Mediero; Susana Noval; Luciano Bravo-Ljubetic; Inés Contreras; Fernando Carceller
Journal:  Neuroophthalmology       Date:  2015-06-17

7.  Diagnostic Capability of Peripapillary Three-dimensional Retinal Nerve Fiber Layer Volume for Glaucoma Using Optical Coherence Tomography Volume Scans.

Authors:  Ziad Khoueir; Firas Jassim; Linda Yi-Chieh Poon; Edem Tsikata; Geulah S Ben-David; Yingna Liu; Eric Shieh; Ramon Lee; Rong Guo; Georgia Papadogeorgou; Boy Braaf; Huseyin Simavli; Christian Que; Benjamin J Vakoc; Brett E Bouma; Johannes F de Boer; Teresa C Chen
Journal:  Am J Ophthalmol       Date:  2017-08-12       Impact factor: 5.258

8.  Associations of macular thickness in spectral-domain OCT with ocular and systemic cardiovascular parameters - the MIPH Eye & Health Study.

Authors:  Alexander Karl-Georg Schuster; Joachim Ernst Fischer; Urs Vossmerbaeumer
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-09-26       Impact factor: 3.117

9.  Developing new automated alternation flicker using optic disc photography for the detection of glaucoma progression.

Authors:  J Ahn; I S Yun; H G Yoo; J-J Choi; M Lee
Journal:  Eye (Lond)       Date:  2016-09-23       Impact factor: 3.775

10.  Retinal nerve fibre layer thickness floor and corresponding functional loss in glaucoma.

Authors:  Jean-Claude Mwanza; Donald L Budenz; Joshua L Warren; Aaron D Webel; Courtney E Reynolds; Diego T Barbosa; Shan Lin
Journal:  Br J Ophthalmol       Date:  2014-12-09       Impact factor: 4.638

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.