| Literature DB >> 31435430 |
Shifeng Liu1, Hong Wang2, Congxiao Wang1, Hao Zhang1, Wei Li1, Qian Dong3, Xiaokun Hu1.
Abstract
PURPOSE: To verify the accuracy and efficacy of three-dimensional printing individual template (3D-PIT) with computed tomography-magnetic resonance imaging (CT-MRI) fusion for radioactive iodine-125 (125I) seed implantation in high-grade brain gliomas.Entities:
Keywords: 3D printing template; brachytherapy; computed tomography; dosimetry; glioma; magnetic resonance
Year: 2019 PMID: 31435430 PMCID: PMC6701383 DOI: 10.5114/jcb.2019.85729
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
General status of the 16 patients in our study
| Characteristics | General |
|---|---|
| Sex | |
| Male | 11 |
| Female | 5 |
| Age (years) | 28-69 |
| Karnofsky performance status | Median score, 80 (70-90) |
| Primary treatment | |
| Surgery | 1 |
| Radiotherapy | 5 |
| Combination therapy | 10 |
| Initial staging | |
| III | 6 |
| VI | 10 |
| Prescribed dose (Gy) | 120-140 |
| Seed air-kerma strength (μGy.m2.h-1) | Median, 0.76 (0.60-0.94) |
Fig. 1Preoperative planning design. A) Determining delineation of the target area, needle tract, calculating the dose distribution of target volume based on CT-MRI fusion images by BTPS; B) Dose volume histogram of preoperative plan
Fig. 23D-PIT design with CT-MRI guidance. A) Needle tract avoidance of large blood vessels; B) 3D-PIT appearance structure; C) Possible distribution of seeds in the tumor; D) Actual printed planning 3D-PIT
Fig. 3Seed implantation with the planning 3D-PIT guidance. A) inserting a needle according to the needle tract of 3D-PIT; B) Seed implantation into the tumor
Fig. 4Postoperative dosimetry verification. A) Actual seed distribution of postoperation; B) Dose volume histogram of postoperative plan
Preoperative and postoperative dosimetry parameters for the 16 patients
| Pt | D90 (Gy) | V100 (%) | V200 (%) | CI | EI (%) | |||||
|---|---|---|---|---|---|---|---|---|---|---|
| Pre | Post | Pre | Post | Pre | Post | Pre | Post | Pre | Post | |
| 1 | 157.6 | 157.2 | 97.0 | 96.9 | 51.3 | 51.2 | 0.83 | 0.81 | 13.2 | 15.8 |
| 2 | 148.0 | 148.2 | 95.0 | 94.1 | 42.3 | 41.9 | 0.87 | 0.90 | 9.6 | 12.3 |
| 3 | 144.0 | 142.6 | 97.6 | 98.0 | 51.6 | 51.4 | 0.75 | 0.69 | 15.2 | 18.6 |
| 4 | 136.7 | 136.1 | 96.5 | 96.4 | 49.8 | 52.1 | 0.86 | 0.81 | 10.8 | 9.5 |
| 5 | 168.1 | 162.4 | 97.6 | 97.8 | 54.4 | 54.0 | 0.79 | 0.72 | 18.3 | 15.2 |
| 6 | 162.4 | 161.8 | 97.8 | 98.3 | 46.9 | 49.5 | 0.80 | 0.78 | 13.5 | 11.8 |
| 7 | 161.6 | 161.9 | 97.5 | 97.6 | 50.2 | 52.5 | 0.83 | 0.85 | 12.1 | 11.9 |
| 8 | 139.2 | 139.8 | 94.7 | 95.1 | 37.0 | 40.1 | 0.65 | 0.75 | 17.4 | 15.7 |
| 9 | 148.5 | 150.2 | 97.5 | 97.9 | 48.4 | 48.7 | 0.79 | 0.85 | 13.2 | 16.3 |
| 10 | 157.8 | 159.3 | 96.8 | 96.7 | 44.2 | 43.5 | 0.81 | 0.85 | 20.8 | 16.7 |
| 11 | 146.8 | 146.5 | 95.6 | 96.3 | 43.9 | 43.2 | 0.76 | 0.70 | 12.7 | 11.5 |
| 12 | 153.7 | 153.9 | 97.0 | 97.1 | 46.2 | 45.6 | 0.75 | 0.80 | 11.4 | 12.3 |
| 13 | 158.9 | 158.7 | 96.9 | 96.5 | 50.5 | 50.4 | 0.83 | 0.80 | 18.7 | 15.9 |
| 14 | 149.6 | 150.0 | 95.2 | 95.5 | 46.3 | 46.5 | 0.78 | 0.82 | 13.9 | 12.6 |
| 15 | 159.9 | 159.6 | 97.2 | 97.3 | 49.4 | 48.2 | 0.70 | 0.90 | 12.9 | 12.0 |
| 16 | 148.2 | 148.5 | 96.8 | 96.2 | 50.3 | 50.6 | 0.85 | 0.83 | 8.5 | 9.8 |
Comparison of preoperative and postoperative dosimetry parameters (16 cases)
| Parameter | Preoperative plan | Postoperative plan | |||||
|---|---|---|---|---|---|---|---|
| Range | Median | M (SD) | Range | Median | M (SD) | ||
| D90 (Gy) | 136.7-168.1 | 151.65 | 152.56 (8.83) | 136.1-162.4 | 152.05 | 152.29 (8.33) | 0.815 |
| V100 (%) | 94.7-97.8 | 96.95 | 96.69 (1.00) | 94.1-98.3 | 96.80 | 96.73 (1.14) | 0.376 |
| V200 (%) | 37.0-54.4 | 48.90 | 47.67 (4.30) | 40.1-54.0 | 49.10 | 48.09 (4.17) | 0.816 |
| CI | 0.65-0.87 | 0.80 | 0.79 (0.06) | 0.69-0.90 | 0.81 | 0.80 (0.06) | 0.697 |
| EI (%) | 8.5-20.8 | 13.20 | 13.89 (3.41) | 9.5-18.6 | 12.45 | 13.62 (2.68) | 0.605 |
V100%, V200% – volume of the GTV receiving 100%, 200% of the prescribed dose, D90 – dose covering 90% of the GTV, CI – conformity index, EI – external index of the target volume, M – mean, SD – standard deviation