[Dose comparison between pre and post operation of 125I seeds implantation guided by 3D print tamplate].

H T Zhang1, X M Di, H M Yu, X Z Zhao, L J Zhang, J X Zhao, C Zhang, Z Z Liu, A X Sui, J Wang.   

Abstract

OBJECTIVE: To compare the dose difference between pre and post operation of (125)I seeds implantation guided by 3D print tamplate and investigate the dose accuracy.
METHODS: From August 2015 to November 2015, a total of 9 patients were selected and underwent 3D print tamplate guided (125)I seeds implantation in Hebei General Hospital. All patients had been fixed as the position of operation and then performed CT scan. After preplan was designed, the 3D tamplates were printed. The tumors were punctured through needles holes predesigned. Another CT scan was used to confirm the locations of needles and then seeds were implanted into tumor according to preplan. Postplan was performed after the operation. The D90, V90, V100, V150 and seeds number pre and post operation were collected and compared.
RESULTS: The mean D90, V90, V100, V150 and seeds number preoperation was (7 684±3 279)cGy, 93.2%±0.8%, 89.8%±0.9%, 62.8%±4.7% and 58±17 respectively. The mean D90, V90, V100, V150 and seeds number postoperation was (7 531±3 523) cGy, 92.4%±2.0%, 88.0%±2.5%, 61.9%±5.3% and 56±17 respectively. The difference of all data between pre and post operation was not statistically significant (all P> 0.05).
CONCLUSIONS: The dose parameters in postplan show no difference compared with preplan. For the fixed tumor, 3D print tamplate guided seeds implantation may become an easy repeated and standard procedure.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27055510     DOI: 10.3760/cma.j.issn.0376-2491.2016.09.010

Source DB:  PubMed          Journal:  Zhonghua Yi Xue Za Zhi        ISSN: 0376-2491


  2 in total

1.  Dosimetry Verification of 125I Seeds Implantation With Three-Dimensional Printing Noncoplanar Templates and CT Guidance for Paravertebral/Retroperitoneal Malignant Tumors.

Authors:  Zhe Ji; Yuliang Jiang; Liang Su; Fuxin Guo; Ran Peng; Haitao Sun; Jinghong Fan; Junjie Wang
Journal:  Technol Cancer Res Treat       Date:  2017-08-07

2.  Dosimetry verification of 3D-printed individual template based on CT-MRI fusion for radioactive 125I seed implantation in recurrent high-grade gliomas.

Authors:  Shifeng Liu; Hong Wang; Congxiao Wang; Hao Zhang; Wei Li; Qian Dong; Xiaokun Hu
Journal:  J Contemp Brachytherapy       Date:  2019-06-28
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.