Xinyi Li1,2, Jackie Wu1,2, Manisha Palta1, You Zhang3, Yang Sheng1, Jiahan Zhang1, Chunhao Wang1.
Abstract
PURPOSE: To optimize collimator setting to improve dosimetric quality of pancreas volumetric modulated arc therapy plan for stereotactic body radiation therapy.Entities:
Keywords: IMRT; SBRT; VMAT; collimator settings; optimization; pancreas cancer; simulated annealing; treatment planning
Mesh:
Year: 2019 PMID: 31426721 PMCID: PMC6702773 DOI: 10.1177/1533033819870767
Source DB: PubMed Journal: Technol Cancer Res Treat ISSN: 1533-0338
Figure 1.Flowchart of the proposed collimator angle optimization algorithm. Red blocks are inputs, and green block is output. Xd indicates the larger jaw opening value of the dual-arc pair in the X direction (multi-leaf collimator [MLC] motion direction).
Figure 2.Dose distribution comparison of Plan (left) and Plan (right) from a selected case. Yellow arrows in (A) and (B) indicate the comparison of dose conformity, and blue arrows in (C) and (D) indicate the comparison of organs at risk (OAR) sparing. Primary planned target volume (PTV25 Gy) is shown in cyan, and PTV33 Gy is shown in red. Proximal OARs are shown in yellow (stomach), dark blue (duodenum), and dark green (large bowel). Isodose lines are shown in yellow (3300 cGy) and white (3135 cGy). Plan indicates proposed algorithm; Plan, plan with the template setting.
Figure 3.Comparison of the dose statistics (normalized to prescription dose value) in Plan and Plan.*Statistical significance. Plan indicates proposed algorithm; Plan, plan with the template setting
Dose Statistic Results of Planconv and Planopt, as shown in Figure 3.a
| OAR | Parameters |
|
| |
|---|---|---|---|---|
| Stomach | 642.0 | 605.5 | .01b | |
| 3078.0 | 3037.5 | .11 | ||
| 3108.0 | 3058.5 | .03b | ||
| Duodenum | 1276.5 | 1256.5 | .13 | |
| 2904.5 | 2838.0 | <.01b | ||
| 2970.0 | 2906.5 | <.01b | ||
| Bowel | 705.0 | 669.5 | .06 | |
| 2776.0 | 2713.0 | .03b | ||
| 2792.0 | 2729.5 | .01b | ||
| Kidney | 449.5 | 458.5 | .01b | |
| Liver | 316.0 | 301.0 | <.01b |
Abbreviations: OAR, organs at risk; Plan, proposed algorithm; Plan, plan with the template setting.
a The reported dose was the median of the 16 studied cases.
b Statistical significance.
Figure 4.Comparisons of the modulation complexity score, weighed multi-leaf collimator (MLC) opening, and total MU between Plan and Plan. Both tests were of statistical significance. Plan indicates proposed algorithm; Plan, plan with the template setting.
Figure 5.Illustration of the initial collimator settings in the dual collimator angles plane. Each pixel value in the figure is the occurrence count of the desirable collimator angle pairs of 39 library cases. Please note that each case could have multiple desirable collimator angle pairs with similar Xd. Red dots represent the selected initial CSa,i.
Figure 6.Collimator angle pair results from (A) simulated annealing optimization and (B) exhausting search (red dots) overlapped to the results from simulated annealing optimization. These 2 methods show good coincidence with each other.