So Ron Choi1, Tae Young Lee1, Sung Wan Kim1, Sang Yoong Park1, Chan Jong Chung1, Jeong Ho Kim1.
Abstract
BACKGROUND: The supraglottic airway device is an appropriate alternative to tracheal intubation in laparoscopic surgery. We compared the Baska MaskⓇ with i-gelⓇ by measuring the oropharyngeal leak pressure (OLP) and hemodynamic and respiratory parameters during laparoscopic cholecystectomy.Entities:
Keywords: Cholecystectomy; Hemodynamics; Laparoscopy; Laryngeal masks; Lung compliance; Pneumoperitoneum
Mesh:
Year: 2019 PMID: 31426623 PMCID: PMC6900424 DOI: 10.4097/kja.19195
Source DB: PubMed Journal: Korean J Anesthesiol ISSN: 2005-6419
Fig. 1.Appearances of Baska Mask and i-gel. (A) Flexing the Baska Mask using the hand-tab. (B) Circumferences of size 3 of i-gel and Baska Mask were 11.7 cm and 13.8 cm, respectively. (C) Circumferences of size 4 of i-gel and Baska Mask were 13 cm and 14.8 cm, respectively.
Fig. 2.CONSORT flow diagram for the study patients.
Patients’ Characteristics and Anesthetic Data
| Variable | i-gel (n = 49) | Baska Mask (n = 48) | P value |
|---|---|---|---|
| Age (yr) | 52.4 ± 13.3 | 48.2 ± 11.0 | 0.095 |
| Sex (M/F) | 18/31 | 21/27 | 0.538 |
| ASA (1/2) | 14/35 | 20/28 | 0.206 |
| Height (cm) | 160.9 ± 8.0 | 163.7 ± 8.0 | 0.094 |
| Weight (kg) | 64.6 ± 11.4 | 62.7 ± 10.9 | 0.416 |
| Mallampati score (1/2/3/4) | 19/28/2/0 | 11/33/4/0 | 0.202 |
| Thyromental distance (cm) | 7.5 ± 1.0 | 7.8 ± 0.7 | 0.058 |
| Inter-incisor distance (cm) | 4.5 ± 0.7 | 4.8 ± 0.5 | 0.093 |
| Anesthetic time (min) | 58.1 ± 16.2 | 59.2 ± 11.8 | 0.723 |
| Pneumoperitoneum time (min) | 25.7 ± 14.4 | 21.4 ± 11.3 | 0.109 |
| Total propofol used (mg) | 396.1 ± 127.3 | 382.8 ± 95.5 | 0.565 |
| Total remifentanil used (μg) | 381.67 ± 151.0 | 394.1 ± 133.3 | 0.671 |
| Analgesic method (0/1/2)[ | 20/28/1 | 21/20/7 | 0.054 |
Values are presented as mean ± SD or number. ASA: American Society of Anesthesiologists.
0: none, 1: local infiltration, 2: Transversus abdominis plane block.
ASA: American Society of Anesthesiologists.
Insertion of i-gel and Baska Mask
| Complications | i-gel (n = 49) | Baska Mask (n = 48) | P value |
|---|---|---|---|
| Size (3/4) | 27/22 | 26/22 | 0.926 |
| Attempt no. (1/2) | 48/1 | 48/0 | 0.320 |
| Insert time (min) | 29.0 ± 10.3 | 31.4 ± 6.3 | 0.165 |
| Insert resistance (1/2/3/4)[ | 42/7/0/0 | 44/4/0/0 | 0.355 |
| Fiberoptic view (1/2/3/4)[ | 42/5/2/0 | 41/4/3/0 | 0.855 |
| Manipulation frequency (0/1/2) | 46/3/0 | 42/6/0 | 0.233 |
| OLP (cmH2O) | 26.7 ± 4.5 | 29.6 ± 6.8 | 0.014 |
Values are presented as mean ± SD or number.
1: easy, 2: moderate, 3: difficult, 4: impossible.
1: vocal cord only, 2: vocal cords plus posterior epiglottis visible, 3: vocal cords plus anterior epiglottis visible, 4: vocal cords not seen.
OLP: oropharyngeal leak pressure.
Fig. 3.Changes in the heart rate (A) and mean arterial pressure (B) between the i-gel group and Baska Mask group. Values are presented as mean ± SD. There was no significant difference in the heart rate and mean arterial pressure between the groups. T0: baseline, T1: 1 min after induction, T2: 5 min before CO2 pneumoperitoneum, T3: 5 min after start of CO2 pneumoperitoneum, T4: 5 min before removal of pneumoperitoneum, T5: 5 min after removal of pneumoperitoneum.
Perioperative Complications
| i-gel (n = 49) | Baska Mask (n = 48) | P value | |
|---|---|---|---|
| Audible air leakage | 4 (8.2) | 3 (6.3) | 0.512 |
| Desaturation | 0 (0) | 0 (0) | - |
| Gastric distention | 0 (0) | 0 (0) | - |
| Regurgitation/aspiration | 0 (0) | 0 (0) | - |
| Blood on device | 3 (6.1) | 2 (4.2) | 0.510 |
| Teeth or lip injury | 0 (0) | 0 (0) | - |
| Coughing | 0 (0) | 0 (0) | - |
| Vomiting | 0 (0) | 0 (0) | - |
| Bronchospasm | 0 (0) | 0 (0) | - |
| Sore throat | 10 (20.4) | 19 (39.6) | 0.119 |
| Dysphonia | 2 (4.1) | 1 (2.1) | 0.566 |
| Dysphagia | 2 (4.1) | 7 (14.3) | 0.075 |
Values are presented as number (%).