Dongye Yang1, Liting Lin1, Penghao Guo2, Weina Zhang1, Xingfang He1, Qiuyu Huang1, Longhui Lai1, Weiqing Long2. 1. Guanghua School of Stomatology, Hospital of Stomatology, Sun Yat-Sen University, Guangdong Provincial Key Laboratory of Stomatology, Guangzhou, China. 2. Department of Clinical Laboratory, The First Affiliated Hospital of Sun Yat-Sen University, Guangzhou, China.
Abstract
AIM: This study aims to assess the infection risks of flashlight contamination in a stomatology hospital and compare the disinfection effectiveness of alcohol (75%) and disinfecting wipes. BACKGROUND: The flashlight is a basic non-critical medical device in oral and maxillofacial surgery wards. Wounds are mostly found in oral cavities; therefore, reusable flashlights may be a potential source of nosocomial infections (NIs). However, the microbial flora present in flashlights used in hospitals has not yet been explored. METHODS: This study investigated the microbial contamination of 41 flashlights used in a stomatology hospital in Guangzhou in March 2016. RESULTS: Results indicated that 75.6%(31/41) of the flashlights had microbial contamination. Gram-positive bacteria accounted for 72.7%(24/33)of the microbial groups contaminating the flashlights, and Gram-negative bacteria (21.2%, 7/33), and fungi (6.1%, 2/33) constituted the remaining contaminants. The predominantly isolated species was Staphyloccus (66.7%, 22/33), especially Staphylococcus aureus (24.2%, 8/33). Approximately 77.3% (17/22) of the types of bacteria detected in the hands were same as those in the corresponding flashlights. Both the bacterial overstandard and S. aureus detection rates of doctors' flashlights were higher than those of nurses' flashlights (16/17 vs. 14/23, 7/17 vs. 1/23, respectively) (P < 0.05). Moreover, both disinfectants performed excellently, and their eligibility rates were not significantly different (17/17 vs. 14/14) (P > 0.05). CONCLUSION: Flashlights are potential causes of NIs. Disinfecting flashlights could be an effective and practical infection control method.
AIM: This study aims to assess the infection risks of flashlight contamination in a stomatology hospital and compare the disinfection effectiveness of alcohol (75%) and disinfecting wipes. BACKGROUND: The flashlight is a basic non-critical medical device in oral and maxillofacial surgery wards. Wounds are mostly found in oral cavities; therefore, reusable flashlights may be a potential source of nosocomial infections (NIs). However, the microbial flora present in flashlights used in hospitals has not yet been explored. METHODS: This study investigated the microbial contamination of 41 flashlights used in a stomatology hospital in Guangzhou in March 2016. RESULTS: Results indicated that 75.6%(31/41) of the flashlights had microbial contamination. Gram-positive bacteria accounted for 72.7%(24/33)of the microbial groups contaminating the flashlights, and Gram-negative bacteria (21.2%, 7/33), and fungi (6.1%, 2/33) constituted the remaining contaminants. The predominantly isolated species was Staphyloccus (66.7%, 22/33), especially Staphylococcus aureus (24.2%, 8/33). Approximately 77.3% (17/22) of the types of bacteria detected in the hands were same as those in the corresponding flashlights. Both the bacterial overstandard and S. aureus detection rates of doctors' flashlights were higher than those of nurses' flashlights (16/17 vs. 14/23, 7/17 vs. 1/23, respectively) (P < 0.05). Moreover, both disinfectants performed excellently, and their eligibility rates were not significantly different (17/17 vs. 14/14) (P > 0.05). CONCLUSION: Flashlights are potential causes of NIs. Disinfecting flashlights could be an effective and practical infection control method.
Entities:
Keywords:
Cross-contamination; Disinfectant; Flashlights; Infection control
Nosocomial infections (NIs) are rising worldwide [1]. Disease transmission in hospitals occurs mainly through the air, droplets, or by contact, directly (touching with the hands) or indirectly (fomites) [2]. Microorganisms on a fomite, which may remain viable for days or months, could be transmitted among patients and hospital personnel [3]. Several studies have demonstrated the bacterial contamination of non-critical medical devices (NCMDs), such as otoscopes, white coats, pens, goggles, and other surfaces [3], [4], [5], [6], [7], [8]. In China, the flashlight is a basic NCMD for healthcare workers. This device is usually used for examinations, especially for oral care in oral and maxillofacial surgery wards, because patients' partial wounds are in oral cavities. Microorganisms are commonly transmitted through the hands of healthcare personnel, but hospital materials and articles could also carry microorganisms [1]. Thus, reusable flashlights may be a potential source of NIs. However, the microbial flora present in flashlights used in hospitals has not yet been explored.
Materials and methods
Design
A study was conducted in March 2016 in the wards of a stomatology hospital in Guangzhou.
Methods
A total of 41 flashlights of the same type were included in the study. They were used by doctors and nurses, except for one exclusively used by an isolated patient. The flashlights were small and cylindrical, with a length of 9.5 cm, a radius of 1.25 cm, and a total surface area of approximately 28.66 cm2. According to Chinese standards, if the surface area of an object is below 100 cm2, the whole object needs to be swabbed [9]. Therefore, the entire flashlight surface was swabbed before disinfection. These swab samples were divided into the following: A, swabs from the flashlights used by doctors (n = 17); B, swabs from the flashlights used by nurses (n = 23); and C, swabs from the flashlight used by the isolated patient (n = 1). The hands of doctors and nurses were swabbed, and the swabs were marked with the same numbers in accordance with the flashlight samples. After disinfection, additional swabs from the entire flashlight surface were obtained. These swab samples were divided into the following: A, swabs from the flashlights disinfected using alcohol (75%) (n = 20) and B, swabs from the flashlights disinfected using CaviWipes (Metre, Orange, CA; 17.2% isopropanol and 0.28% diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride) (n = 21). The disinfection procedure lasted for at least 30 s, and the swabs were collected within 5 min. During sampling before disinfection, the swabs were placed in 10 mL tubes of common broth at the end of sampling. The common broth was used as a neutralizing agent in the alcohol group, whereas common broth supplemented with 0.3% Tween 80 and 0.3% lecithinin as surfactants was used as the neutralizing agent in the disinfectant wipes group [10], [11]. Each sample was placed in a tube and shaken 80 times in an oscillator before the solution (1 mL) was extracted and inoculated on a general nutrient agar plate, which was then incubated for 48 h at 37 °C. The colony morphology on the culture plate was then observed. According to the standard protocols for contamination, an object is not contaminated if the total number of bacterial colonies based on the culture result is 0 CFU/cm2. If the number is greater than 10 CFU/cm2, then the object exceeds the standard; if the number is less than 10 CFU/cm2 after disinfection, then it can be considered as an eligible contamination [9].
Data analysis
Frequency of positive data was analyzed using a test. The MecNemar test was used to evaluate the differences before and after disinfection. All statistical procedures were performed on SPSS software (version 13.0; Chicago, IL, USA). Statistical significance was considered at p < 0.05.
Results
Bacterial colony contamination of flashlights
Up to 75.6% (31/41)of the flashlights were contaminated with bacteria and/or fungi, 93.5%(29/31) of which were colonized by one organism and 6.5%(2/31) were colonized by more than one organism. In total, Gram-positive bacteria accounted for 72.7%(24/33) of the microbial group contaminating the flashlights, and Gram-negative bacteria (21.2%, 7/33) and fungi (6.1%, 2/33) constituted the remaining contaminants. The distribution of microorganisms isolated from the flashlights is elaborated in Fig. 1. The leading isolated species was Staphyloccus (66.7%, 22/33), with Staphylococcus aureus being the most common (24.2%,8/33). Moreover, Pseudomonas aeruginosa was isolated from the isolated patient. Both bacterial overstandard and S. aureus detection rates of the doctors' flashlights were higher than those of the nurses' flashlights (16/17 vs. 14/23, 7/17 vs. 1/23, respectively) (P < 0.05) (Table 1).
Fig. 1
Distribution of the microorganisms isolated from the flashlights.
Table 1
Positive rates of the flashlights between the two groups before disinfection.
Group
Bacterial overstandard
Staphylococcus aureus
Doctors' flashlights (n = 17)
16/17
7/17
Nurses' flashlights (n = 23)
14/23
1/23
χ2
5.763
6.145
P value
0.016
0.013
Distribution of the microorganisms isolated from the flashlights.Positive rates of the flashlights between the two groups before disinfection.
Comparison of the types of bacteria between flashlights and health care workers' hands before disinfection
Up to 53.6% (22/41) of the health care workers' hands were contaminated with bacteria. Moreover, 77.3% (17/22) of the types of bacteria detected in the hands were the same as those in the corresponding flashlights.
Comparison of the disinfection effectiveness of alcohol and disinfecting wipes
Colony reduction after disinfecting either with alcohol or CaviWipes was significant, and the eligibility rates of alcohol and CaviWipes were not significantly different(17/17 vs. 14/14) (P > 0.05) (Table 2).
Table 2
Positive rates of the flashlights before and after disinfection.
Group
Before
After
χ2
P value
75% alcohol (n = 20)
17/20
0/20
26.189
<0.001
CaviWipes (n = 21)
14/21
0/21
18.107
<0.001
Positive rates of the flashlights before and after disinfection.
Discussion
In clinics, the flashlight is a common tool for healthcare workers. Oral and maxillofacial surgery patients usually have partial wounds in the oral cavities with saliva or other secretions. The flashlight does not have direct contact with patients; however, the contamination probability and cross-infection risk are high if the flashlight is used during wound cleaning. Usually, only one healthcare worker changes the patients' dressings; therefore, the worker must remove the gauze from wounds using his hands and clean wounds with one hand while holding a flashlight in the other. Therefore, microorganisms are likely to be transmitted from the patients' wounds to the healthcare personnel's hands. However, flashlight disinfection is not consciously practiced by the users. Flashlights can harbor high amounts of pathogenic microorganisms, which can be transmitted to other patients and cause NIs. Li-sha Shi et al. [12]. found that S. aureus or MRSA cross-contamination between medical workers' hands and contact surfaces occurs within and between departments of a large metropolitan hospital. Healthcare workers' hands are major vectors of pathogenic cross-transmission, causing an estimated 20%–40% of health care-associated infections [13]. Uneke [8] reported that NCMD contamination rates range from 25% to 100%, with Staphylococcus as the predominant contaminant; however, healthcare workers rarely disinfect NCMDs used by different patients. Longtin Y et al. [6] found substantial contamination levels in stethoscopes after each physical examination, and these contamination levels are comparable with those in parts of the physician's dominant hand. Similar results are shown in this study. The results indicated that all flashlights have been used; however, none of the flashlights was disinfected after each use. Therefore, 75.6%(31/41) of the flashlights had microbial contamination, with Staphylococcus, especially S. aureus, being predominant. Moreover, 77.3%(17/22) of the types of bacteria detected in the hands were same as those in the corresponding flashlights. Therefore, flashlights may be crucial in transmitting pathogenic bacteria in hospitals. The results also showed that both the bacterial overstandard and S. aureus detection rates of the doctors' flashlights were higher than those of the nurses' flashlights. The reasons could be that the doctors' flashlights were for public use and the nurses' flashlights were for personal use, which meant that the former were frequently and widely shared by more staff and in more disease types, thereby leading to increased opportunities for spreading pathogenic microorganisms. Therefore, healthcare workers in cases of contact with patients or hospital settings are supposed to practice proper hand hygiene and disinfect flashlights after each use. Moreover, public flashlights should be placed in designated positions and disinfected after every working shift. P. aeruginosa is a leading cause of NIs and responsible for opportunistic infections in patients with low resistance [14]. In the present study, one flashlight was exclusively used by one isolated patient with P. aeruginosa infection. The purpose of sampling from the exclusively used flashlight was to determine a correlation between the bacteria that the flashlight carried and those of the isolated patient. The results verified this assumption and showed that the total amount of bacteria in this sampled flashlight exceeded the standard, with the main pathogen being P. aeruginosa. Thus, flashlights used by isolated patients carry high contamination risks, and isolated patients need to use special flashlights. Furthermore, a strict terminal disinfection of flashlights should be implemented in hospitals.To eliminate the probability of NIs, appropriate disinfection measures should be undertaken. Alcohol (75%) and CaviWipes are common disinfectants in stomatology wards. Alcohol, which causes bacterial proteins to coagulate, has the highest disinfection effectiveness (70%–75%). CaviWipes is a multi-purpose disinfectant containing isopropanol and diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride as active ingredients [10], [11]. In the present study, the results proved that both disinfectants are effective.
Conclusion
Flashlights are potential nosocomial pathogen reservoirs, but decontaminating them is usually ignored. Disinfecting flashlights could be an effective and practical infection control method. Alcohol and CaviWipes are excellent flashlight disinfectants, and both are recommended as potent disinfectants in practice.
Funding
This research was supported by the Extracurricular Scientific Research Program for Students of Sun Yat-Sen University.
Conflict of interest
No conflict of interest had been declared by the authors.
Authors: David J Weber; William A Rutala; Melissa B Miller; Kirk Huslage; Emily Sickbert-Bennett Journal: Am J Infect Control Date: 2010-06 Impact factor: 2.918
Authors: M J Bruins; D Wijshake; S V de Vries-van Rossum; R G J Klein Overmeen; G J H M Ruijs Journal: J Hosp Infect Date: 2013-06-10 Impact factor: 3.926