Otitis externa following aural irrigation linked to instruments contaminated with Pseudomonas aeruginosa.

M J Bruins1, D Wijshake, S V de Vries-van Rossum, R G J Klein Overmeen, G J H M Ruijs.   

Abstract

BACKGROUND: The incidence of acute otitis externa, an infection of the external auditory canal, in general practitioners' (GP) practices in The Netherlands is about 14 per 1000 patients per year. In early 2010, one of the authors noted that some of the otitis externa patients in his GP practice had undergone cerumen removal by ear syringing a few weeks earlier. Bacterial cultures of samples taken from the instruments used showed contamination of an ear syringe by Pseudomonas aeruginosa. From then on, P. aeruginosa isolates from patients' ears were stored in the laboratory. AIM: It was assessed whether cross-contamination with P. aeruginosa between patients in the same GP practice could occur through the use of contaminated ear lavage instruments.
METHODS: From 17 GP practices, the otolaryngology Outpatient Department and the Out-of-Hours GP Service, instruments used for examining and cleaning the outer ear were swabbed. Strains of P. aeruginosa cultured from the instruments were genotyped together with isolates of patients registered in the same practice.
FINDINGS: In four practices where contaminated instruments were found, genotyping showed similarity between P. aeruginosa strains isolated from a patient and the ear syringe, and/or between strains of different patients in the same practice.
CONCLUSIONS: Transmission of P. aeruginosa from ear lavage instruments to patients appears to occur with otitis externa as a result. Together with the Infection Control Unit of our hospital we have formulated recommendations for the appropriate cleaning, disinfection and storage of re-usable ear lavage instruments for the GP practices to implement.
Copyright © 2013 The Healthcare Infection Society. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23764317     DOI: 10.1016/j.jhin.2013.04.010

Source DB:  PubMed          Journal:  J Hosp Infect        ISSN: 0195-6701            Impact factor:   3.926


  2 in total

1.  Flashlight contamination and effectiveness of two disinfectants in a stomatology hospital.

Authors:  Dongye Yang; Liting Lin; Penghao Guo; Weina Zhang; Xingfang He; Qiuyu Huang; Longhui Lai; Weiqing Long
Journal:  Int J Nurs Sci       Date:  2017-03-27

2.  Treatment of impacted ear wax: a case for increased community-based microsuction.

Authors:  Jonathan C Radford
Journal:  BJGP Open       Date:  2020-06-23
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.