Dual-energy approach to contrast-enhanced mammography using the balanced filter method: spectral optimization and preliminary phantom measurement.

Masatoshi Saito1.   

Abstract

Dual-energy contrast agent-enhanced mammography is a technique of demonstrating breast cancers obscured by a cluttered background resulting from the contrast between soft tissues in the breast. The technique has usually been implemented by exploiting two exposures to different x-ray tube voltages. In this article, another dual-energy approach using the balanced filter method without switching the tube voltages is described. For the spectral optimization of dual-energy mammography using the balanced filters, we applied a theoretical framework reported by Lemacks et al. [Med. Phys. 29, 1739-1751 (2002)] to calculate the signal-to-noise ratio (SNR) in an iodinated contrast agent subtraction image. This permits the selection of beam parameters such as tube voltage and balanced filter material, and the optimization of the latter's thickness with respect to some critical quantity-in this case, mean glandular dose. For an imaging system with a 0.1 mm thick CsI:T1 scintillator, we predict that the optimal tube voltage would be 45 kVp for a tungsten anode using zirconium, iodine, and neodymium balanced filters. A mean glandular dose of 1.0 mGy is required to obtain an SNR of 5 in order to detect 1.0 mg/cm2 iodine in the resulting clutter-free image of a 5 cm thick breast composed of 50% adipose and 50% glandular tissue. In addition to spectral optimization, we carried out phantom measurements to demonstrate the present dual-energy approach for obtaining a clutter-free image, which preferentially shows iodine, of a breast phantom comprising three major components-acrylic spheres, olive oil, and an iodinated contrast agent. The detection of iodine details on the cluttered background originating from the contrast between acrylic spheres and olive oil is analogous to the task of distinguishing contrast agents in a mixture of glandular and adipose tissues.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18072488     DOI: 10.1118/1.2790841

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  Dosimetric characterization of a dedicated breast computed tomography clinical prototype.

Authors:  Ioannis Sechopoulos; Steve Si Jia Feng; Carl J D'Orsi
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

2.  Quantification of breast density with dual energy mammography: a simulation study.

Authors:  Justin L Ducote; Sabee Molloi
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

3.  X-ray scatter correction method for dedicated breast computed tomography.

Authors:  Ioannis Sechopoulos
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

4.  Quantification of breast density with dual energy mammography: an experimental feasibility study.

Authors:  Justin L Ducote; Sabee Molloi
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

5.  Clinical digital breast tomosynthesis system: dosimetric characterization.

Authors:  Steve Si Jia Feng; Ioannis Sechopoulos
Journal:  Radiology       Date:  2012-02-13       Impact factor: 11.105

6.  Pulse pileup analysis for a double-sided silicon strip detector using variable pulse shapes.

Authors:  Jinghui Wang; Linchuan Chen; Mats Persson; Paurakh L Rajbhandary; Praneeth Kandlakunta; Gabriella Carini; Rebecca Fahrig
Journal:  IEEE Trans Nucl Sci       Date:  2019-05-15       Impact factor: 1.679

7.  A Quantification Method for Breast Tissue Thickness and Iodine Concentration Using Photon-Counting Detector.

Authors:  Seokmin Han
Journal:  J Digit Imaging       Date:  2015-10       Impact factor: 4.056

Review 8.  Breast cancer imaging: a perspective for the next decade.

Authors:  Andrew Karellas; Srinivasan Vedantham
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

9.  Direct measurement of mammographic X-ray spectra with a digital CdTe detection system.

Authors:  Leonardo Abbene; Gaetano Gerardi; Fabio Principato; Stefano Del Sordo; Giuseppe Raso
Journal:  Sensors (Basel)       Date:  2012-06-18       Impact factor: 3.576

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.