Nahid Chegeni1, Mohamad Javad Tahmasebi Birgani2, Fariba Farhadi Birgani1, Daryoush Fatehi3, Gholamreza Akbarizadeh4, Marziyeh Tahmasbi5.
Abstract
BACKGROUND: Recently, magnetic resonance imaging (MRI)-based radiotherapy has become a favorite science field for treatment planning purposes. In this study, a simple algorithm was introduced to create synthetic computed tomography (sCT) of the head from MRI.Entities:
Keywords: Deformable registration; Demons algorithm; radiotherapy; synthetic computed tomography
Year: 2019 PMID: 31316906 PMCID: PMC6601231 DOI: 10.4103/jmss.JMSS_26_18
Source DB: PubMed Journal: J Med Signals Sens ISSN: 2228-7477
Figure 1Steps of creating synthetic computed tomography: registering each magnetic resonance image form dataset to the target-magnetic resonance (1) calculating of displacement fields (2), applying the same transformation to the corresponding computed tomography form dataset (3), generating deformed CTs (4), and Fusing the collection of deformed CTs into a single synthetic computed tomography (5)
Figure 2(a) Axial slices for bone and brain window synthetic computed tomography based on T1 or T2 weighted and real computed tomography. (b) Bone outline. (c) The subtraction of the corresponding Real CT and synthetic CT
The average value and standard deviation of the mean absolute error, Dice similarity coefficient of bone, Hounsfield Unit gamma-index and percentage of pixels with Hounsfield Unit gamma-index >1 are shown for 10 patients with glioblastoma multiforme and 10 patient with other brain tumor
| Synthetic CT | %Γ | |||
|---|---|---|---|---|
| For GBM | ||||
| T1-based (bone window) | 85±14.7 | 0.85±0.06 | 0.52±0.33 | 17±8 |
| T2-based (bone window) | 83±15.3 | 0.80±0.02 | 0.44±0.2 | 15±4 |
| T1-based (brain window) | 80±10.8 | 0.83±0.05 | 0.37±0.21 | 10±6 |
| T2-based (brain window) | 78±8.1 | 0.82±0.02 | 0.29±0.16 | 7±2 |
| For other brain tumors | ||||
| T1-based (bone window) | 83±12.4 | 0.89±0.03 | 0.59±0.19 | 13±7 |
| T2-based (bone window) | 93±13 | 0.82±0.01 | 0.64±0.32 | 19±8 |
| T1-based (brain window) | 80±9.7 | 0.85±0.04 | 0.36±0.18 | 9±1.3 |
| T2-based (brain window) | 82±8.3 | 0.84±0.02 | 0.38±0.21 | 11±5 |
*Percentage of pixels with ГHU >1. CT – Computed tomography; GBM – Glioblastoma multiforme; MAE – Mean absolute error; DSCbone – Dice similarity coefficient of bone; HU – Hounsfield unit; ГHU – HU gamma-index