Clinical Evaluation of the New Rebound Tonometers Icare PRO and Icare ONE Compared With the Goldmann Tonometer.

Javier Moreno-Montañés1, José M Martínez-de-la-Casa, Alfonso L Sabater, Laura Morales-Fernandez, Cristina Sáenz, Julián Garcia-Feijoo.   

Abstract

PURPOSE: To compare the participant-obtained intraocular pressures (IOPs) using Icare ONE and the clinician-obtained values using Icare PRO, both rebound tonometers, with Goldmann tonometry (GAT) values and analyze the ease of use of Icare ONE.
METHODS: One hundred fifty participants were included (60 normal controls, 90 patients with glaucoma/ocular hypertension). The participants measured the IOP 3 times using Icare ONE; a clinician measured the IOP once using Icare PRO. The instruments were used randomly. Clinical data were evaluated to analyze the difficulty of the technique, the effect on the results, and the ease of use of Icare ONE.
RESULTS: The mean IOPs with GAT, Icare ONE, and Icare PRO were 16.6±4.43, 17.5±5.42, and 16.6±4.77 mm Hg, respectively. The participant-measured IOP values were within +3 mm Hg of the GAT values in 67.1% of eyes with Icare ONE and in 79.6% with Icare PRO. The limits of agreement were higher with Icare ONE compared with Icare PRO. IOP value errors were found with Icare ONE in eyes with low and high GAT-IOP. The areas under the curve to detect IOPs of 21 mm Hg or higher (GAT) exceeded 0.80 with both tonometers. Young participants reported better ease of use with Icare ONE. No other factors were related to the results.
CONCLUSIONS: Icare ONE may be useful for patients monitoring their IOP values; most individuals can use the device after a short training session. Icare PRO had better results compared with Icare ONE in all IOP ranges.

Entities:  

Mesh:

Year:  2015        PMID: 24844537     DOI: 10.1097/IJG.0000000000000058

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  15 in total

1.  Agreement among Goldmann applanation tonometer, iCare, and Icare PRO rebound tonometers; non-contact tonometer; and Tonopen XL in healthy elderly subjects.

Authors:  Yoshitake Kato; Shunsuke Nakakura; Naoko Matsuo; Kayo Yoshitomi; Marina Handa; Hitoshi Tabuchi; Yoshiaki Kiuchi
Journal:  Int Ophthalmol       Date:  2017-04-09       Impact factor: 2.031

2.  A comparison of Icare PRO and Perkins tonometers in anesthetized children.

Authors:  Massimiliano Serafino; Edoardo Villani; Andrea Lembo; Giovanni Rabbiolo; Claudia Specchia; Rupal H Trivedi; Paolo Nucci
Journal:  Int Ophthalmol       Date:  2019-07-16       Impact factor: 2.031

3.  Measurement of Intraocular Pressure by Patients With Glaucoma.

Authors:  Savva Pronin; Lyndsay Brown; Roly Megaw; Andrew J Tatham
Journal:  JAMA Ophthalmol       Date:  2017-10-01       Impact factor: 7.389

4.  Twenty-four-hour efficacy of preservative-free tafluprost for open-angle glaucoma patients, assessed by home intraocular pressure (Icare-ONE) and blood-pressure monitoring.

Authors:  Soon Young Cho; Yong Yeon Kim; Chungkwon Yoo; Tae-Eun Lee
Journal:  Jpn J Ophthalmol       Date:  2015-09-28       Impact factor: 2.447

5.  [Ease of handling of first and second generation rebound tonometers].

Authors:  N Mihailovic; J Termühlen; M Alnawaiseh; N Eter; T S Dietlein; A Rosentreter
Journal:  Ophthalmologe       Date:  2016-04       Impact factor: 1.059

Review 6.  Systematic Review of Current Devices for 24-h Intraocular Pressure Monitoring.

Authors:  Sabita M Ittoop; Jeffrey R SooHoo; Leonard K Seibold; Kaweh Mansouri; Malik Y Kahook
Journal:  Adv Ther       Date:  2016-08-16       Impact factor: 3.845

7.  Effects of short-term mild hypercapnia during head-down tilt on intracranial pressure and ocular structures in healthy human subjects.

Authors:  Steven S Laurie; Gianmarco Vizzeri; Giovanni Taibbi; Connor R Ferguson; Xiao Hu; Stuart M C Lee; Robert Ploutz-Snyder; Scott M Smith; Sara R Zwart; Michael B Stenger
Journal:  Physiol Rep       Date:  2017-06

8.  Evaluation of rebound tonometer iCare IC200 as compared with IcarePRO and Goldmann applanation tonometer in patients with glaucoma.

Authors:  Shunsuke Nakakura; Ryo Asaoka; Etsuko Terao; Yuki Nagata; Yasuko Fukuma; Satomi Oogi; Miku Shiraishi; Yoshiaki Kiuchi
Journal:  Eye Vis (Lond)       Date:  2021-07-01

9.  Diurnal Intraocular Pressure Fluctuations with Self-tonometry in Glaucoma Patients and Suspects: A Clinical Trial.

Authors:  Jessie Huang; Paula Katalinic; Michael Kalloniatis; Michael P Hennessy; Barbara Zangerl
Journal:  Optom Vis Sci       Date:  2018-02       Impact factor: 1.973

10.  Comparison of intraocular pressure measurements using Goldmann tonometer, I-care pro, Tonopen XL, and Schiotz tonometer in patients after Descemet stripping endothelial keratoplasty.

Authors:  Oded Ohana; David Varssano; Gabi Shemesh
Journal:  Indian J Ophthalmol       Date:  2017-07       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.