| Literature DB >> 31269951 |
Zhongqing Wang1,2, Cheng Peng3, Hui Kang4, Xia Fan4, Runqing Mu4, Liping Zhou4, Miao He2, Bo Qu5.
Abstract
BACKGROUND: The autoverification system for coagulation consists of a series of rules that allow normal data to be released without manual verification. With new advances in medical informatics, the laboratory information system (LIS) has growing potential for the autoverification, allowing rapid and accurate verification of clinical laboratory tests. The purpose of the study is to develop and evaluate a LIS-based autoverification system for validation and efficiency.Entities:
Keywords: Autoverification; Coagulation; Laboratory information systems; Medical safety; Turnaround time
Mesh:
Year: 2019 PMID: 31269951 PMCID: PMC6609390 DOI: 10.1186/s12911-019-0848-2
Source DB: PubMed Journal: BMC Med Inform Decis Mak ISSN: 1472-6947 Impact factor: 2.796
Fig. 1Autoverification decision-making tree. Autoverification procedures for single tests and whole samples in the order of quality control check, analytical error flags, critical values, limited range check, delta check, logical rules and patient’s clinical diagnosis
Clinical reference intervals, critical range, 5 and 95% percentiles, and the limited range
| Clinical reference intervals | Critical value | Percentiles (5 and 95%) | Limited Range | |
|---|---|---|---|---|
| PT | 11.00–14.30 s | ≤ 9.00 s or ≥ 70.00 s | 11.80–16.30 s | 11.00–16.30 s |
| APTT | 32.00–43.00 s | ≤ 15.00 s or ≥ 100.00 s | 30.40–46.40 s | 30.40–46.40 s |
| TT | 14.0 0–21.00 s | > 150.00 s | 15.00–19.40 s | 14.00–21.00 s |
| FBG | 2.00–4.00 g/L | < 1.00 g/L | 2.21–6.51 s | 2.00–6.51 g/L |
PT Prothrombin time, APTT Activated partial thromboplastin time, TT Thrombin time, FBG Fibrinogen
Distribution intervals of clinical large-scale data in coagulation
| n | Mean | SD | Median | Range | Minimum | Maximum | Percentiles | ||
|---|---|---|---|---|---|---|---|---|---|
| 5% | 95% | ||||||||
| PT | 157,097 | 13.60 | 2.51 | 13.10 | 109.20 | 10.00 | 119.20 | 11.80 | 16.30 |
| APTT | 157,097 | 36.97 | 5.76 | 36.10 | 153.30 | 20.80 | 174.10 | 30.40 | 46.40 |
| TT | 157,097 | 17.12 | 4.04 | 16.80 | 216.30 | 13.00 | 229.30 | 15.00 | 19.40 |
| FBG | 157,097 | 3.80 | 1.35 | 3.50 | 12.94 | 0.60 | 13.54 | 2.21 | 6.51 |
SD Standard deviation, PT Prothrombin time, APTT Activated partial thromboplastin time, TT Thrombin time, FBG Fibrinogen
Validation of the initial online test
| May 2017 | June 2017 | |
|---|---|---|
| n | 13,949 | 14,295 |
| Pass | 10,828 | 11,257 |
| Fail | 3121 | 3038 |
| True-positive | 858 | 892 |
| True-negative | 10,823 | 11,257 |
| False-positive | 2263 | 2146 |
| False-negative | 5 | 0 |
| Sensitivity | 99.42% | 100% |
| Specificity | 82.71% | 83.99% |
| Passing rate | 77.63% | 78.75% |
True-positive, intercepted problematic reports; True-negative, auto-released correct reports; False-positive, intercepted correct problematic reports; False-negative, auto-released problematic reports
Passing rate of the autoverification system of single tests and whole samples
| n | Autoverification results | PT | APTT | TT | FBG | Whole sample | |
|---|---|---|---|---|---|---|---|
| July 2017 | 13,416 | Pass | 12,485 | 12,389 | 12,536 | 12,557 | 10,453 |
| Fail | 931 | 1027 | 880 | 859 | 2963 | ||
| Passing rate (%) | 93.06 | 92.34 | 93.44 | 93.60 | 77.91 | ||
| August 2017 | 14,477 | Pass | 13,349 | 13,413 | 13,723 | 13,573 | 11,354 |
| Fail | 1128 | 1064 | 754 | 904 | 3123 | ||
| Passing rate (%) | 92.21 | 92.65 | 94.79 | 93.76 | 78.43 | ||
| September 2017 | 13,673 | Pass | 12,784 | 12,456 | 12,585 | 12,932 | 10,740 |
| Fail | 889 | 1217 | 1088 | 741 | 2933 | ||
| Passing rate (%) | 93.50 | 91.10 | 92.04 | 94.58 | 78.55 | ||
| October 2017 | 13,004 | Pass | 12,235 | 12,059 | 11,989 | 12,304 | 10,289 |
| Fail | 769 | 945 | 1015 | 700 | 2715 | ||
| Passing rate (%) | 94.09 | 92.73 | 92.19 | 94.62 | 79.12 | ||
| November 2017 | 14,866 | Pass | 13,832 | 13,905 | 14,203 | 13,857 | 11,934 |
| Fail | 1034 | 961 | 663 | 1009 | 2932 | ||
| Passing rate (%) | 93.04 | 93.54 | 95.54 | 93.21 | 80.28 | ||
| December 2017 | 14,263 | Pass | 12,995 | 13,378 | 13,254 | 13,157 | 11,258 |
| Fail | 1268 | 885 | 1009 | 1106 | 3005 | ||
| Passing rate (%) | 91.11 | 93.80 | 92.93 | 92.25 | 78.93 |
PT Prothrombin time, APTT Activated partial thromboplastin time, TT Thrombin time, FBG Fibrinogen
Comparision of different autoverification systems in the studies
| Study | Year | Country | Assay | System Type | Sample Number | Passing Rate |
|---|---|---|---|---|---|---|
| Randell EW et al [ | 2018 | Canada | Clinical Chemistry and immunoassay | Middleware | 80,867 | 91.1 to 94.7% |
| Xia LY et al [ | 2017 | China | Clinical Chemistry and immunoassay | LIS | 31,349 | 74.0% |
| Krasowski MD et al [ | 2014 | USA | Clinical Chemistry and immunoassay | Middleware | – | 99.5% |
| Palmieri R et al [ | 2018 | Italy | Urinalysis | AutionMAX-SediMAX | 1002 | 52.4% |
| Sediq AM et al [ | 2014 | Egypt | Thyroid Function | LIS | 563 | 63.8% |
| Zhao Y et al [ | 2014 | China | Coagulation | Middleware | 2353 | 82.0% |
| Our Study | – | China | Coagulation | LIS | 83,699 | 78.86% |
LIS Laboratory information system