A novel weighted cumulative delta-check method for highly sensitive detection of specimen mix-up in the clinical laboratory.

Teppei Yamashita1, Kiyoshi Ichihara, Ayaho Miyamoto.   

Abstract

BACKGROUND: We sought to detect specimen mix-up by developing a new cumulative delta-check method applicable to a mixture of test items with heterogeneous units and distribution patterns.
METHODS: The distributions of all test results were successfully made Gaussian using power transformation. Values were then standardized into z-score (zx) based on reference interval (RI) so that limits of RI take zx=±1.96. To find a weight for summing absolute value of delta between current and previous zx (Dz), we evaluated the distribution of Dz. Its central portion was always regarded as Gaussian despite the presence of symmetrical long tails. Thus, an adjusted SD (aSD) representing the center was estimated with an iterative method. By setting 1/aSD2 as a weight factor, we computed a weighted mean of Dz as an index for specimen mix-up (wCDI).
RESULTS: The performance of wCDI was evaluated, using a model laboratory database consisting of 32 basic test items, by a simulation study generating artificial cases of mix-up. When wCDI was computed from three commonly ordered test sets consisting of 6-9 items each, its diagnostic efficiency in detecting the artificial cases was 0.937-0.967 expressed as area under ROC curves (AUC). When the performance of wCDI was evaluated simply by the number of test items (p) included in the computation, AUC gradually increased from 0.944 (p=5) to 0.976 (p=8). However, when p≥10, AUC stayed at approximately 0.98.
CONCLUSIONS: wCDI was proven to be highly effective in uncovering cases of specimen mix-up. The diagnostic efficiency of wCDI depends only on the number of test items included in the computation.

Entities:  

Mesh:

Year:  2013        PMID: 23388451     DOI: 10.1515/cclm-2012-0752

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  2 in total

1.  Design and evaluation of a LIS-based autoverification system for coagulation assays in a core clinical laboratory.

Authors:  Zhongqing Wang; Cheng Peng; Hui Kang; Xia Fan; Runqing Mu; Liping Zhou; Miao He; Bo Qu
Journal:  BMC Med Inform Decis Mak       Date:  2019-07-03       Impact factor: 2.796

2.  A study on quality control using delta data with machine learning technique.

Authors:  Yufang Liang; Zhe Wang; Dawei Huang; Wei Wang; Xiang Feng; Zewen Han; Biao Song; Qingtao Wang; Rui Zhou
Journal:  Heliyon       Date:  2022-07-14
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.