Literature DB >> 31193931

Determination of initial electron parameters by means of Monte Carlo simulations for the Siemens Artiste Linac 6 MV photon beam.

Taylan Tuğrul1, Osman Eroğul2.   

Abstract

AIM: In this study, we investigated initial electron parameters of Siemens Artiste Linac with 6 MV photon beam using the Monte Carlo method.
BACKGROUND: It is essential to define all the characteristics of initial electrons hitting the target, i.e. mean energy and full width of half maximum (FWHM) of the spatial distribution intensity, which is needed to run Monte Carlo simulations. The Monte Carlo is the most accurate method for simulation of radiotherapy treatments.
MATERIALS AND METHODS: Linac head geometry was modeled using the BEAMnrc code. The phase space files were used as input file to DOSXYZnrc simulation to determine the dose distribution in a water phantom. We obtained percent depth dose curves and the lateral dose profile. All the results were obtained at 100 cm of SSD and for a 10 × 10 cm2 field.
RESULTS: We concluded that there existed a good conformity between Monte Carlo simulation and measurement data when we used electron mean energy of 6.3 MeV and 0.30 cm FWHM value as initial parameters. We observed that FWHM values had very little effect on PDD and we found that the electron mean energy and FWHM values affected the lateral dose profile. However, these effects are between tolerance values.
CONCLUSIONS: The initial parameters especially depend on components of a linac head. The phase space file which was obtained from Monte Carlo Simulation for a linac can be used as calculation of scattering, MLC leakage, to compare dose distribution on patients and in various studies.

Entities:  

Keywords:  Artiste; BEAMnrc; DOSXYZnrc; Monte Carlo; Simulation

Year:  2019        PMID: 31193931      PMCID: PMC6545364          DOI: 10.1016/j.rpor.2019.05.002

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  10 in total

1.  Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version.

Authors:  I Kawrakow
Journal:  Med Phys       Date:  2000-03       Impact factor: 4.071

2.  Sensitivity of megavoltage photon beam Monte Carlo simulations to electron beam and other parameters.

Authors:  Daryoush Sheikh-Bagheri; D W O Rogers
Journal:  Med Phys       Date:  2002-03       Impact factor: 4.071

Review 3.  Monte Carlo modelling of external radiotherapy photon beams.

Authors:  Frank Verhaegen; Jan Seuntjens
Journal:  Phys Med Biol       Date:  2003-11-07       Impact factor: 3.609

4.  Determining the incident electron fluence for Monte Carlo-based photon treatment planning using a standard measured data set.

Authors:  Paul J Keall; Jeffrey V Siebers; Bruce Libby; Radhe Mohan
Journal:  Med Phys       Date:  2003-04       Impact factor: 4.071

5.  Influence of initial electron beam parameters on Monte Carlo calculated absorbed dose distributions for radiotherapy photon beams.

Authors:  Antonis Tzedakis; John E Damilakis; Michael Mazonakis; John Stratakis; Haralambos Varveris; Nicholas Gourtsoyiannis
Journal:  Med Phys       Date:  2004-04       Impact factor: 4.071

6.  Monte Carlo linear accelerator simulation of megavoltage photon beams: independent determination of initial beam parameters.

Authors:  Sigrun Saur Almberg; Jomar Frengen; Arve Kylling; Tore Lindmo
Journal:  Med Phys       Date:  2012-01       Impact factor: 4.071

7.  Determination of the initial beam parameters in Monte Carlo linac simulation.

Authors:  Khaled Aljarrah; Greg C Sharp; Toni Neicu; Steve B Jiang
Journal:  Med Phys       Date:  2006-04       Impact factor: 4.071

8.  Monte Carlo Simulation of Siemens ONCOR Linear Accelerator with BEAMnrc and DOSXYZnrc Code.

Authors:  Keyvan Jabbari; Hossein Saberi Anvar; Mohammad Bagher Tavakoli; Alireza Amouheidari
Journal:  J Med Signals Sens       Date:  2013-07

9.  Monte Carlo modeling of a Novalis Tx Varian 6 MV with HD-120 multileaf collimator.

Authors:  Luis Alberto Vazquez-Quino; Brian Massingill; Chengyu Shi; Alonso Gutierrez; Carlos Esquivel; Tony Eng; Nikos Papanikolaou; Sotirios Stathakis
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102

10.  Dosimetry of small bone joint calculated by the analytical anisotropic algorithm: a Monte Carlo evaluation using the EGSnrc.

Authors:  James C L Chow; Runqing Jiang; Amir M Owrangi
Journal:  J Appl Clin Med Phys       Date:  2013-01-06       Impact factor: 2.102

  10 in total
  2 in total

1.  Monte Carlo evaluation of particle interactions within the patient-dependent part of Elekta 6 MV photon beam applying IAEA phase space data.

Authors:  Deae-Eddine Krim; Dikra Bakari; Mustapha Zerfaoui; Abdeslem Rrhioua; Yassine Oulhouq
Journal:  Rep Pract Oncol Radiother       Date:  2021-12-30

2.  The Effect of Algorithms on Dose Distribution in Inhomogeneous Phantom: Monaco Treatment Planning System versus Monte Carlo Simulation.

Authors:  Taylan Tuğrul
Journal:  J Med Phys       Date:  2021-08-07
  2 in total