| Literature DB >> 31114685 |
Marius Andronache1, Nikola Drca2, Graziana Viola3.
Abstract
Ablation of AF through electrical isolation of the pulmonary veins is a well-established technique and a cornerstone in the ablation of AF, although there are a variety of techniques and ablation strategies now available. However, high numbers of patients are returning to hospital after ablation procedures such as pulmonary vein isolation (PVI). Scar tissue (as identified by contact voltage mapping) is found to be present in many of these patients, especially those with persistent AF and even those with paroxysmal AF. This scarring is associated with poor outcomes after PVI. Cardiac mapping is necessary to locate triggers and substrate so that an ablation strategy can be optimised. Multipolar mapping catheters offer more information regarding the status of the tissue than standard ablation catheters. A patient-tailored catheter ablation approach, targeting the patient-specific low voltage/fibrotic substrate can lead to improved outcomes.Entities:
Keywords: AF; Atrial fibrosis; catheter ablation; multipolar catheters; pulmonary vein isolation
Year: 2019 PMID: 31114685 PMCID: PMC6528045 DOI: 10.15420/aer.2018.57.1
Source DB: PubMed Journal: Arrhythm Electrophysiol Rev ISSN: 2050-3369
Currently Available Multipolar Catheters for Mapping Atrial Scar
| Name | Details |
|---|---|
| PentaRay (Biosense Webster) | 20 electrodes arranged in 5 soft radiating splines (1 mm electrodes separated by 2 mm interelectrode spacing if 2–6–2 mm used) |
| AFocus II Duo-decapolar (Abbott) | 20 electrodes with 4-4-4 mm spacing |
| LASSO (Biosense Webster) | Circular mapping catheter, 20-polar catheter, 1 mm electrodes, 2–6–2 mm spacing |
| INTELLAMAP ORION (Boston Scientific) | 64 electrodes: 8 splines with 8 electrodes per spline, 0.4 mm2 electrode size, 2.5 mm spacing |
| EnSite (Abbott) | Grid catheter, 16 electrodes (1 mm electrodes with 4 mm electrode spacing) |
| Advisor HD (Abbott) | Circular mapping catheter, 10 electrodes. 2 sizes: 15 mm diameter (3–3–3 mm spacing) and 20 mm (5–5–5 spacing) |
| Achieve (Medtronic) | Circular mapping catheter, 8 electrodes. 2 sizes: 15 mm diameter (4 mm spacing) and 20 mm diameter (6 mm spacing) |