Literature DB >> 16831837

Electrical isolation of pulmonary veins in patients with atrial fibrillation: reduction of fluoroscopy exposure and procedure duration by the use of a non-fluoroscopic navigation system (NavX).

Heidi Luise Estner1, Isabel Deisenhofer, Armin Luik, Gjin Ndrepepa, Christian von Bary, Bernhard Zrenner, Claus Schmitt.   

Abstract

AIMS: The aim of the study was to investigate the feasibility of performing segmental pulmonary vein (PV) isolation guided by the NavX (Endocardial Solutions, St Jude Medical, Inc., St Paul, MN, USA) system without the three-dimensional (3D) geometric reconstruction option and whether the use of NavX system will reduce the radiation exposure and procedure duration. METHODS AND
RESULTS: The study included 64 patients with symptomatic paroxysmal or permanent atrial fibrillation, in whom PV isolation was performed using fluoroscopic guidance (n=32) or the NavX system (n=32). Pulmonary vein mapping with a circular mapping catheter allowed the identification and localization of myocardial connections between the PV and the left atrium. PV isolation was performed by radiofrequency ablation of these connections at the atrial aspect of the PV ostium. Primary success rate for isolated PVs did not differ significantly in patients ablated under fluoroscopic guidance vs. those ablated under guidance of NavX system [100/107 PVs (93.5%) vs. 120/124 PV (96.8%; P=n.s.)]. Compared with fluoroscopy guided procedures, NavX-guided procedures showed a significant reduction in the fluoroscopy time (75.8+/-24.5 vs. 38.9+/-19.3 min, P<0.05), total X-ray exposure (93.2+/-51.6 vs. 56.6+/-37.9 Gy cm(2), P=0.03), and total procedural time (237.7+/-65.4 vs. 188.6+/-62.7 min, P=0.01). The mean follow-up was 9.5+/-3.0 months. One patient in each group was lost to follow-up. Seven-day Holter monitoring showed that 23 of 31 patients (74.2%) in the NavX-guided group and 21 of 31 patients (67.7%) in the fluoroscopy-guided group were in sinus rhythm (P=0.57).
CONCLUSION: The 3D visualization of the catheters by NavX system allows a rapid and precise visualization of the mapping and ablation catheters at the PV ostia and markedly reduces fluoroscopy time, total X-ray exposure, and procedural duration during PV isolation compared with ablation performed under fluoroscopy guidance.

Entities:  

Mesh:

Year:  2006        PMID: 16831837     DOI: 10.1093/europace/eul079

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  41 in total

1.  Point-by-point pulmonary vein antrum isolation guided by intracardiac echocardiography and 3D mapping and duty-cycled multipolar AF ablation: effect of multipolar ablation on procedure duration and fluoroscopy time.

Authors:  Yaariv Khaykin; Lauren Zarnett; Daniel Friedlander; Zaev A Wulffhart; Bonnie Whaley; David Giewercer; Bernice Tsang; Atul Verma
Journal:  J Interv Card Electrophysiol       Date:  2012-03-11       Impact factor: 1.900

2.  Impact of 3D mapping on procedural characteristics and outcomes in cryoballoon pulmonary vein isolation for atrial fibrillation.

Authors:  Emily N Guhl; Evan Adelstein; Andrew Voigt; Norman C Wang; Samir Saba; Sandeep K Jain
Journal:  J Interv Card Electrophysiol       Date:  2018-01-05       Impact factor: 1.900

3.  Cost analysis of periprocedural imaging in patients undergoing catheter ablation for atrial fibrillation.

Authors:  Sean D Pokorney; Bradley G Hammill; Laura G Qualls; Benjamin A Steinberg; Lesley H Curtis; Jonathan P Piccini
Journal:  Am J Cardiol       Date:  2014-05-02       Impact factor: 2.778

4.  CARTO-guided vs. NavX-guided pulmonary vein antrum isolation and pulmonary vein antrum isolation performed without 3-D mapping: effect of the 3-D mapping system on procedure duration and fluoroscopy time.

Authors:  Yaariv Khaykin; Richard Oosthuizen; Lauren Zarnett; Zaev A Wulffhart; Bonnie Whaley; Carol Hill; David Giewercer; Atul Verma
Journal:  J Interv Card Electrophysiol       Date:  2011-01-21       Impact factor: 1.900

5.  2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation.

Authors:  Hugh Calkins; Gerhard Hindricks; Riccardo Cappato; Young-Hoon Kim; Eduardo B Saad; Luis Aguinaga; Joseph G Akar; Vinay Badhwar; Josep Brugada; John Camm; Peng-Sheng Chen; Shih-Ann Chen; Mina K Chung; Jens Cosedis Nielsen; Anne B Curtis; D Wyn Davies; John D Day; André d'Avila; N M S Natasja de Groot; Luigi Di Biase; Mattias Duytschaever; James R Edgerton; Kenneth A Ellenbogen; Patrick T Ellinor; Sabine Ernst; Guilherme Fenelon; Edward P Gerstenfeld; David E Haines; Michel Haissaguerre; Robert H Helm; Elaine Hylek; Warren M Jackman; Jose Jalife; Jonathan M Kalman; Josef Kautzner; Hans Kottkamp; Karl Heinz Kuck; Koichiro Kumagai; Richard Lee; Thorsten Lewalter; Bruce D Lindsay; Laurent Macle; Moussa Mansour; Francis E Marchlinski; Gregory F Michaud; Hiroshi Nakagawa; Andrea Natale; Stanley Nattel; Ken Okumura; Douglas Packer; Evgeny Pokushalov; Matthew R Reynolds; Prashanthan Sanders; Mauricio Scanavacca; Richard Schilling; Claudio Tondo; Hsuan-Ming Tsao; Atul Verma; David J Wilber; Teiichi Yamane
Journal:  Heart Rhythm       Date:  2017-05-12       Impact factor: 6.343

6.  Non-fluoroscopic catheter visualization using MediGuide™ technology: experience from the first 600 procedures.

Authors:  P Sommer; S Richter; G Hindricks; S Rolf
Journal:  J Interv Card Electrophysiol       Date:  2014-01-16       Impact factor: 1.900

Review 7.  Advanced mapping techniques in atrial fibrillation.

Authors:  Lars Lickfett; Jörg O Schwab; Thorsten Lewalter
Journal:  J Interv Card Electrophysiol       Date:  2008-04-02       Impact factor: 1.900

8.  Impact of respiration gating on image integration guided atrial fibrillation ablation.

Authors:  Emin Evren Özcan; Gabor Szeplaki; Tamas Tahin; Istvan Osztheimer; Szabolcs Szilagyi; Astrid Apor; Pal Maurovich Horvath; Hajnalka Vago; Béla Merkely; Laszlo Geller
Journal:  Clin Res Cardiol       Date:  2014-05-07       Impact factor: 5.460

9.  Recent trends in imaging for atrial fibrillation ablation.

Authors:  Rajesh Kabra; Jagmeet Singh
Journal:  Indian Pacing Electrophysiol J       Date:  2010-05-05

10.  Catheter ablation of atrial fibrillation using the Navx-/Ensite-system and a CT-/MRI-guided approach.

Authors:  Klaus Kettering; Gerald F Greil; Michael Fenchel; Ulrich Kramer; Hans-Joerg Weig; Mathias Busch; Stephan Miller; Ludger Sieverding; Roman Laszlo; Juergen Schreieck
Journal:  Clin Res Cardiol       Date:  2009-03-13       Impact factor: 5.460