Andrew D Hahn1, Jeff Kammerman1, Michael Evans2, Wei Zha1, Robert V Cadman1, Keith Meyer3, Nathan Sandbo3, Sean B Fain1,4,5. 1. Department of Medical Physics, University of Wisconsin, Madison, Wisconsin, USA. 2. Department of Biostatistics and Medical Informatics, University of Wisconsin, Madison, Wisconsin, USA. 3. Department of Medicine, University of Wisconsin, Madison, Wisconsin, USA. 4. Department of Radiology, University of Wisconsin, Madison, Wisconsin, USA. 5. Department of Biomedical Engineering, University of Wisconsin, Madison, Wisconsin, USA.
Abstract
BACKGROUND: MRI of hyperpolarized 129 Xenon (HP 129 Xe) is increasingly utilized for investigating pulmonary function. The solubility of HP 129 Xe in lung tissue, blood plasma (Barrier), and red blood cells (RBC), with unique chemical shifts, enables spectroscopic imaging of potential imaging biomarkers of gas exchange and microstructural pulmonary physiology. PURPOSE: To quantify global average and regional repeatability of Barrier:gas, RBC:gas, and RBC:Barrier ratios derived from dissolved-phase 129 Xe imaging and their dependence on intervisit changes in lung inflation volume. STUDY TYPE: Prospective. POPULATION: Fourteen healthy volunteers. One subject was unable to complete the study resulting in 13 subjects for analysis (eight female, five male, ages 24-69, 53.8 ± 13.9). FIELD STRENGTH: 1.5T. ASSESSMENT: Subjects were imaged using a 3D radial 1-point Dixon method to separate Barrier and RBC component signals, at two different timepoints, with ~1 month between visits. RBC:Gas, Barrier:Gas, and RBC:Barrier measures were compared across time and with pulmonary function tests (PFTs). STATISTICAL TESTS: Repeatablilty was quantified using Bland-Altman plots, coefficient of repeatability, coefficient of variation (CV), and intraclass correlation coefficients (ICCs). Dependence of imaging measures on PFTs and lung volume was evaluated using Spearman and Pearson correlation coefficients, respectively. Statistical significance was determined by F-test for intraclass correlations, and t-test for Spearman correlations and regression. RESULTS: Mean RBC:Gas, Barrier:Gas, and RBC:Barrier had CVs of 19.2%, 20.0%, and 11.5%, respectively, and had significant ICCs, equal to 0.78, 0.79, and 0.92, respectively. Intervisit differences in RBC:Barrier were significantly correlated with intervisit differences in DLCO (r = 0.93, P = 0.007). Significant correlations with intervisit lung volume differences and intervisit differences in mean RBC:Gas (r = -0.73, P = 0.005) and Barrier:Gas (r = -0.69, P = 0.009) were found. DATA CONCLUSION: Three commonly used 129 Xe MRI-based measures of gas-exchange show good repeatability, particularly the Barrier:RBC ratio, which did not depend on lung inflation volume and was strongly associated with intervisit changes in DLCO . LEVEL OF EVIDENCE: 1 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2019;50:1182-1190.
BACKGROUND: MRI of hyperpolarized 129 Xenon (HP 129 Xe) is increasingly utilized for investigating pulmonary function. The solubility of HP 129 Xe in lung tissue, blood plasma (Barrier), and red blood cells (RBC), with unique chemical shifts, enables spectroscopic imaging of potential imaging biomarkers of gas exchange and microstructural pulmonary physiology. PURPOSE: To quantify global average and regional repeatability of Barrier:gas, RBC:gas, and RBC:Barrier ratios derived from dissolved-phase 129 Xe imaging and their dependence on intervisit changes in lung inflation volume. STUDY TYPE: Prospective. POPULATION: Fourteen healthy volunteers. One subject was unable to complete the study resulting in 13 subjects for analysis (eight female, five male, ages 24-69, 53.8 ± 13.9). FIELD STRENGTH: 1.5T. ASSESSMENT: Subjects were imaged using a 3D radial 1-point Dixon method to separate Barrier and RBC component signals, at two different timepoints, with ~1 month between visits. RBC:Gas, Barrier:Gas, and RBC:Barrier measures were compared across time and with pulmonary function tests (PFTs). STATISTICAL TESTS: Repeatablilty was quantified using Bland-Altman plots, coefficient of repeatability, coefficient of variation (CV), and intraclass correlation coefficients (ICCs). Dependence of imaging measures on PFTs and lung volume was evaluated using Spearman and Pearson correlation coefficients, respectively. Statistical significance was determined by F-test for intraclass correlations, and t-test for Spearman correlations and regression. RESULTS: Mean RBC:Gas, Barrier:Gas, and RBC:Barrier had CVs of 19.2%, 20.0%, and 11.5%, respectively, and had significant ICCs, equal to 0.78, 0.79, and 0.92, respectively. Intervisit differences in RBC:Barrier were significantly correlated with intervisit differences in DLCO (r = 0.93, P = 0.007). Significant correlations with intervisit lung volume differences and intervisit differences in mean RBC:Gas (r = -0.73, P = 0.005) and Barrier:Gas (r = -0.69, P = 0.009) were found. DATA CONCLUSION: Three commonly used 129 Xe MRI-based measures of gas-exchange show good repeatability, particularly the Barrier:RBC ratio, which did not depend on lung inflation volume and was strongly associated with intervisit changes in DLCO . LEVEL OF EVIDENCE: 1 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2019;50:1182-1190.
Authors: John P Mugler; Talissa A Altes; Iulian C Ruset; Isabel M Dregely; Jaime F Mata; G Wilson Miller; Stephen Ketel; Jeffrey Ketel; F William Hersman; Kai Ruppert Journal: Proc Natl Acad Sci U S A Date: 2010-11-22 Impact factor: 11.205
Authors: Susan M Galbraith; Martin A Lodge; N Jane Taylor; Gordon J S Rustin; Søren Bentzen; J James Stirling; Anwar R Padhani Journal: NMR Biomed Date: 2002-04 Impact factor: 4.044
Authors: Philip H Quanjer; Sanja Stanojevic; Tim J Cole; Xaver Baur; Graham L Hall; Bruce H Culver; Paul L Enright; John L Hankinson; Mary S M Ip; Jinping Zheng; Janet Stocks Journal: Eur Respir J Date: 2012-06-27 Impact factor: 16.671
Authors: Zackary I Cleveland; Gary P Cofer; Gregory Metz; Denise Beaver; John Nouls; S Sivaram Kaushik; Monica Kraft; Jan Wolber; Kevin T Kelly; H Page McAdams; Bastiaan Driehuys Journal: PLoS One Date: 2010-08-16 Impact factor: 3.240
Authors: Leith J Rankine; Ziyi Wang; Jennifer M Wang; Mu He; H Page McAdams; Joseph Mammarappallil; Craig R Rackley; Bastiaan Driehuys; Robert M Tighe Journal: Ann Am Thorac Soc Date: 2020-01
Authors: Peter J Niedbalski; Elianna A Bier; Ziyi Wang; Matthew M Willmering; Bastiaan Driehuys; Zackary I Cleveland Journal: J Appl Physiol (1985) Date: 2020-06-18
Authors: Laura L Walkup; Kasiani C Myers; Matthew M Willmering; Parinda A Mehta; Adam S Nelson; Robert J Fleck; Jason C Woods; Stella M Davies; Christopher T Towe Journal: Am J Respir Crit Care Med Date: 2021-12-01 Impact factor: 21.405
Authors: Peter J Niedbalski; Junlan Lu; Chase S Hall; Mario Castro; John P Mugler; Yun M Shim; Bastiaan Driehuys Journal: Magn Reson Med Date: 2021-10-13 Impact factor: 4.668
Authors: Tahmina Achekzai; Kai Ruppert; Luis Loza; Faraz Amzajerdian; Harrilla Profka; Ian F Duncan; Stephen J Kadlecek; Rahim R Rizi Journal: Magn Reson Med Date: 2022-08-31 Impact factor: 3.737
Authors: Ziyi Wang; Leith Rankine; Elianna A Bier; David Mummy; Junlan Lu; Alex Church; Robert M Tighe; Aparna Swaminathan; Yuh-Chin T Huang; Loretta G Que; Joseph G Mammarappallil; Sudarshan Rajagopal; Bastiaan Driehuys Journal: J Appl Physiol (1985) Date: 2021-03-18
Authors: Faraz Amzajerdian; Kai Ruppert; Hooman Hamedani; Ryan Baron; Yi Xin; Luis Loza; Tahmina Achekzai; Ian F Duncan; Yiwen Qian; Mehrdad Pourfathi; Stephen Kadlecek; Rahim R Rizi Journal: Magn Reson Med Date: 2020-12-07 Impact factor: 4.668
Authors: Leith J Rankine; Ziyi Wang; Chris R Kelsey; Elianna Bier; Bastiaan Driehuys; Lawrence B Marks; Shiva K Das Journal: Int J Radiat Oncol Biol Phys Date: 2021-07-13 Impact factor: 7.038
Authors: David G Mummy; Elianna A Bier; Ziyi Wang; Jennifer Korzekwinski; Lake Morrison; Christina Barkauskas; H Page McAdams; Robert M Tighe; Bastiaan Driehuys; Joseph G Mammarappallil Journal: Radiology Date: 2021-07-27 Impact factor: 29.146