Measurement variability in single-breath diffusing capacity of the lung.

Naresh M Punjabi1, David Shade, Anshul M Patel, Robert A Wise.   

Abstract

STUDY
OBJECTIVES: The single-breath diffusing capacity of the lung (DLCO) is a commonly performed pulmonary function test. The current American Thoracic Society (ATS) recommendations for reproducibility of DLCO measurements suggest that two measurements for the DLCO agree within 10% or 3 mL/min/mm Hg of the average value. The European Respiratory Society (ERS) recommends that two measurements should agree within 10%. The objectives of the present study were to examine whether the current reproducibility criteria were met in a general pulmonary function laboratory and to determine whether alternative criteria might be appropriate.
DESIGN: Cross-sectional study.
SETTING: University-based pulmonary function laboratory. PATIENTS OR PARTICIPANTS: Patients referred for spirometry, helium lung volumes, and DLCO measurement.
INTERVENTIONS: None. MEASUREMENTS AND
RESULTS: In a sample of 6,193 patients referred for clinical testing, 98.3% had two DLCO values that fulfilled the current ATS criteria for reproducibility. The coefficient of variation (CV) and the percentage difference between two repeat measurements were inversely associated with the baseline DLCO and the FEV(1). As the baseline DLCO (percentage of predicted) or FEV(1) (percentage of predicted) decreased, there was an increase in the CV and the percentage difference. In contrast, the absolute difference between repeat measurements was relatively stable irrespective of the baseline DLCO or FEV(1) values. Other patient factors, such as gender and race, were not associated with measurement variability. Using an absolute difference of 2 to 2.5 mL/min/mm Hg between two DLCO measurements as alternative criteria for reproducibility, 91.5% and 95.8% of the patient sample fulfilled these criteria, respectively.
CONCLUSIONS: Reproducibility of the DLCO measurement is generally much better than current standards allow. Future standards should consider an absolute difference rather than a percentage difference criterion for DLCO reproducibility.

Entities:  

Mesh:

Year:  2003        PMID: 12684297     DOI: 10.1378/chest.123.4.1082

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  7 in total

1.  Repeatability of regional pulmonary functional metrics of Hyperpolarized 129 Xe dissolved-phase MRI.

Authors:  Andrew D Hahn; Jeff Kammerman; Michael Evans; Wei Zha; Robert V Cadman; Keith Meyer; Nathan Sandbo; Sean B Fain
Journal:  J Magn Reson Imaging       Date:  2019-04-10       Impact factor: 4.813

2.  Oxygen-weighted Hyperpolarized (3)He MR Imaging: A Short-term Reproducibility Study in Human Subjects.

Authors:  Masaru Ishii; Hooman Hamedani; Justin T Clapp; Stephen J Kadlecek; Yi Xin; Warren B Gefter; Milton D Rossman; Rahim R Rizi
Journal:  Radiology       Date:  2015-06-23       Impact factor: 11.105

3.  Carbon monoxide diffusion capacity: how low can you go for hematopoietic cell transplantation eligibility?

Authors:  Jason W Chien; Keith M Sullivan
Journal:  Biol Blood Marrow Transplant       Date:  2009-04       Impact factor: 5.742

4.  Variability of spirometry in chronic obstructive pulmonary disease: results from two clinical trials.

Authors:  Laura B Herpel; Richard E Kanner; Shing M Lee; Henry E Fessler; Frank C Sciurba; John E Connett; Robert A Wise
Journal:  Am J Respir Crit Care Med       Date:  2006-02-23       Impact factor: 21.405

5.  Intersession variability in single-breath diffusing capacity in diabetics without overt lung disease.

Authors:  Michael B Drummond; Pamela F Schwartz; William T Duggan; John G Teeter; Richard J Riese; Richard C Ahrens; Robert O Crapo; Richard D England; Neil R Macintyre; Robert L Jensen; Robert A Wise
Journal:  Am J Respir Crit Care Med       Date:  2008-05-08       Impact factor: 21.405

Review 6.  Diffusing capacity.

Authors:  Matthew J Hegewald
Journal:  Clin Rev Allergy Immunol       Date:  2009-03-31       Impact factor: 8.667

7.  Diffusing Capacity for Carbon Monoxide Predicts Response to Balloon Pulmonary Angioplasty in Patients With Inoperable Chronic Thromboembolic Pulmonary Hypertension.

Authors:  Xin Li; Yi Zhang; Qin Luo; Qing Zhao; Qixian Zeng; Tao Yang; Qi Jin; Lu Yan; Anqi Duan; Jiaran Liu; Chenhong An; Xiuping Ma; Changming Xiong; Zhihui Zhao; Zhihong Liu
Journal:  Front Cardiovasc Med       Date:  2021-12-02
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.