Volker Otten1, Sebastian Mukka1, Kjell Nilsson1, Sead Crnalic1, Johan Kärrholm2.
Abstract
Background and purpose - Uncemented cups in total hip arthroplasty (THA) are often augmented with additional screws to enhance their primary stability. We investigated whether there is a difference in the risk for revision between cups with screw holes and cups without screw holes. Patients and methods - We analyzed the risk for cup revision of uncemented cups registered in the Swedish Hip Arthroplasty Register (SHAR) between 2000 and 2017 with respe ct to the presence of screw holes. Only patients with primary osteoarthritis (OA) were included. 22,725 cups, including 12,354 without screw holes and 10,371 with screw holes, were evaluated. Revision rates at 2 and 10 years after the primary operation were analyzed. Results - At a median follow-up time of 3.4 years (0-18), 459 cup revisions were reported. The main reasons for cup revision during the whole observation time were infection, 52% of all cup revisions, and dislocation, 26% of all cup revisions. The survival rate with cup revision due to aseptic loosening as endpoint was 99.9% (95% CI 99.8-99.9) at 2 years for both cups with and cups without screw holes, and the survival rates at 10 years were 99.5% (CI 99.3-99.7) and 99.1% (CI 98.6-99.5), respectively. Cups without screw holes showed a decreased risk of revision due to any reason at both 2 years (adjusted hazard ratio [HR] 0.6, CI 0.5-0.8) and 10 years (HR 0.7, CI 0.5-0.9). Interpretation - We found a very low revision rate for aseptic loosening with modern, uncemented cup designs. Cups with screw holes had an increased risk of revision due to any reason in patients with primary OA.Entities:
Mesh:
Year: 2019 PMID: 30955399 PMCID: PMC6534209 DOI: 10.1080/17453674.2019.1599777
Source DB: PubMed Journal: Acta Orthop ISSN: 1745-3674 Impact factor: 3.717
Figure 1.Procedures not included in the final analysis.
Baseline characteristics of the study population
| Screw holes | ||||
|---|---|---|---|---|
| No n (%) | Yes n (%) | Totals n (%) | p-value a | |
| Sex | ||||
| Male | 6,916 (56) | 5,405 (52) | 12,321 (54) | < 0.001 |
| Female | 5,438 (44) | 4,966 (48) | 10,404 (46) | |
| Age | ||||
| < 45 | 403 (3.3) | 406 (3.9) | 809 (3.6) | < 0.001 |
| 45–64 | 7,428 (60) | 6,395 (62) | 13,823 (61) | |
| 65–74 | 3,662 (30) | 2,607 (25) | 6,269 (28) | |
| ≥ 75 | 861 (7.0) | 963 (9.3) | 1,824 (8.0) | |
| Side | ||||
| Right | 6,550 (53) | 5,572 (54) | 12,122 (53) | 0.3 |
| Left | 5,804 (47) | 4,799 (46) | 10,603 (47) | |
| Approach | ||||
| Posterolateral | 7,769 (63) | 3,252 (31) | 11,021 (49) | < 0.001 |
| Direct lateral | 4,016 (33) | 7,008 (68) | 11,024 (49) | |
| Other | 464 (3.8) | 97 (0.9) | 561 (2.5) | |
| Type of stem: | ||||
| Uncemented | 10,580 (86) | 8,715 (84) | 19,295 (85) | 0.001 |
| Cemented | 1,774 (14) | 1,656 (16) | 3,430 (15) | |
| Cup coating | ||||
| No HA | 6,662 (54) | 3,483 (34) | 10,145 (45) | < 0.001 |
| HA | 5,692 (46) | 6,888 (66) | 12,580 (55) | |
| Head size | ||||
| 28 mm | 1,385 (11) | 2,434 (24) | 3,819 (17) | < 0.001 |
| 32 mm | 7,585 (61) | 7,128 (69) | 14,713 (65) | |
| 36 mm | 3,384 (27) | 809 (7.8) | 4,193 (19) | |
| Head material | ||||
| Metal | 9,260 (75) | 9,542 (92) | 18,802 (83) | < 0.001 |
| Ceramic | 3,094 (25) | 828 (8.0) | 3,922 (17) | |
| Sum | 12,354 (100) | 10,371 (100) | 22,725 (100) | |
Distribution of possible confounding factors between cups with and without screw holes
Pearson’s chi-square.
Missing data in 119 cases.
Number of procedures included per cup design, percentage per group, and median (range) follow-up time in years
| Without screw holes | With screw holes | Total | ||||
|---|---|---|---|---|---|---|
| Cup design | n (%) | follow-up a | n (%) | follow-up a | n (%) | follow-up a |
| Regenerex | 115 (1) | 4.2 (0.1–7.8) | 617 (6) | 3.3 (0–9.6) | 732 (3) | 3.6 (0–9.6) |
| Pinnacle | 1,993 (16) | 3.1 (0–10.9) | 758 (7) | 2.2 (0–12.0) | 2,751 (12) | 2.9 (0–12.0) |
| Pinnacle Gription | 2,916 (24) | 1.2 (0–6.3) | 448 (4) | 0.9 (0–4.8) | 3,364 (15) | 1.2 (0–6.3) |
| Trident AD | 1,055 (9) | 6.3 (0–13.8) | 760 (7) | 5.3 (0–13.7) | 1,815 (8) | 6.0 (0–13.8) |
| Trident hemi | 3,092 (25) | 2.4 (0–11.1) | 213 (2) | 2.3 (0–9.8) | 3,305 (15) | 2.4 (0–11.1) |
| Tritanium | 394 (3) | 4.9 (0–8.1) | 153 (1) | 2.7 (0.1–8.7) | 547 (2) | 4.2 (0–8.7) |
| Continuum | 2,294 (19) | 2.7 (0–7.2) | 1,215 (12) | 3.1 (0–8.2) | 3,509 (15) | 2.9 (0–8.2) |
| Trilogy | 495 (4) | 7.8 (0–15.6) | 6,207 (60) | 6.9 (0–17.7) | 6,702 (29) | 7.0 (0–17.7) |
| Total | 12,354 (100) | 2.6 (0–15.6) | 10,371 (100) | 5.2 (0–17.7) | 22,725 (100) | 3.4 (0–17.7) |
Median (range) follow-up time in years
There was a significant difference in follow-up time between the groups (p < 0.001, ANOVA).
Figure 2.Number of operations per year.
Number of cups at risk of revision in cups with and without screw holes at 2 and 10 years after primary operation
| 0 years | 2 years | 10 years | |
|---|---|---|---|
| Without screw holes | 12,354 | 7,228 | 347 |
| Regenerex | 115 | 85 | 0 |
| Pinnacle | 1,993 | 1,272 | 8 |
| Pinnacle Gription | 2,916 | 1,008 | 0 |
| Trident AD | 1,055 | 904 | 216 |
| Trident Hemi | 3,092 | 1,758 | 9 |
| Tritanium | 394 | 330 | 0 |
| Continuum | 2,294 | 1,430 | 0 |
| Trilogy | 495 | 441 | 114 |
| With screw holes | 10,371 | 7,984 | 1,499 |
| Regenerex | 617 | 450 | 0 |
| Pinnacle | 758 | 397 | 34 |
| Pinnacle Gription | 448 | 114 | 0 |
| Trident AD | 760 | 542 | 106 |
| Trident Hemi | 213 | 123 | 0 |
| Tritanium | 153 | 99 | 0 |
| Continuum | 1,215 | 777 | 0 |
| Trilogy | 6,207 | 5,482 | 1,359 |
Survival after 2 and 10 years and hazard ratio for revision
| Endpoint | 2-year survival (95% CI) | 10-year survival (95% CI) | Crude HR (0–2 years) (95% CI) | p-value | Adjusted HR (0–2 years) (95% CI) | p-value | Crude HR (0–10 years) (95% CI) | p-value | Adjusted HR (0–10 years) (95% CI) | p-value |
|---|---|---|---|---|---|---|---|---|---|---|
| Cup revision for aseptic loosening | 0.9 (0.5–1.8) | 0.8 | ||||||||
| Without screw holes | 99.9 (99.8–99.9) | 99.1 (98.6–99.5) | 0.8 (0.4–1.7) | 0.6 | 0.6 (0.2–1.8) | 0.4 | 1.2 (0.7–2.1) | 0.5 | 1.0 (ref) | |
| With screw holes | 99.9 (99.8–99.9) | 99.5 (99.3–99.7) | 1.0 (ref) | 1.0 (ref) | 1.0 (ref) | |||||
| Cup revision for any reason | 0.7 (0.5–0.9) | 0.004 | ||||||||
| Without screw holes | 98.6 (98.4–98.8) | 96.5 (95.8–97.2) | 0.8 (0.7–1.0) | 0.09 | 0.6 (0.5–0.8) | 0.002 | 1.0 (0.8–1.2) | 0.7 | 1.0 (ref) | |
| With screw holes | 98.4 (98.2–98.7) | 96.8 (96.3–97.2) | 1.0 (ref) | 1.0 (ref) | 1.0 (ref) | |||||
| Revision (cup or stem) for any reason | 0.7 (0.6–0.8) | < 0.001 | ||||||||
| Without screw holes | 98.0 (97.7–98.2) | 95.0 (94.1–95.9) | 0.9 (0.7–1.0) | 0.09 | 0.6 (0.5–0.8) | 0.001 | 1.0 (0.8–1.1) | 0.6 | 1.0 (ref) | |
| With screw holes | 97.7 (97.4–98.0) | 95.4 (94.9–95.9) | 1.0 (ref) | 1.0 (ref) | 1.0 (ref) |
The adjusted hazard ratio was calculated based on a Cox regression model with gender, age, surgical approach, type of stem fixation, cup coating, head size, head material, and cup design as covariates.
Number of cups at risk for revision in cups without/with screw holes was 12,354/10,371 at 0 years, 7,228/7,984 at 2 years and 3,47/1,499 at 10 years after primary operation.
Hazard ratio for cup revision within 2 years for any reason
| No. of hips | No. of revisions | Crude HR (95% CI) | Adjusted HR (95% CI) | p-value | |
|---|---|---|---|---|---|
| Screw holes: | |||||
| Without screw holes | 12,354 | 140 | 0.8 (0.7–1.0) | 0.6 (0.5–0.8) | 0.002 |
| With screw holes | 10,371 | 153 | 1.0 (ref) | 1.0 (ref) | |
| Sex | |||||
| Male | 12,321 | 190 | 1.6 (1.2–2.0) | 1.6 (1.2–2.0) | < 0.001 |
| Female | 10,404 | 103 | 1.0 (ref) | 1.0 (ref) | |
| Age | |||||
| < 45 | 809 | 13 | 1.2 (0.7–2.1) | 1.1 (0.6–2.0) | 0.7 |
| 45–64 | 13,823 | 186 | 1.0 (ref) | 1.0 (ref) | |
| 65–74 | 6,269 | 76 | 0.9 (0.7–1.2) | 1.0 (0.7–1.3) | 0.7 |
| > 75 | 1,824 | 18 | 0.8 (0.5–1.2) | 0.9 (0.5–1.5) | 0.7 |
| Approach | |||||
| Posterolateral | 11,021 | 130 | 1.0 (ref) | 1.0 (ref) | |
| Direct lateral | 11,024 | 159 | 1.2 (0.9–1.5) | 1.0 (0.7–1.3) | 0.9 |
| Other | 561 | 3 | 0.4 (0.1–1.3) | 0.5 (0.1–1.5) | 0.2 |
| Unknown | 119 | 1 | 0.6 (0.1–4.6) | 0.9 (0.1–6.9) | 0.9 |
| Type of stem | |||||
| Uncemented | 19,295 | 269 | 1.0 (ref) | 1.0 (ref) | |
| Cemented | 3,430 | 24 | 0.5 (0.3–0.8) | 0.6 (0.4–1.0) | 0.03 |
| Cup coating | |||||
| No HA | 10,145 | 158 | 1.0 (ref) | 1.0 (ref) | |
| HA | 12,580 | 135 | 0.6 (0.5–0.8) | 0.7 (0.5–10.2) | 0.2 |
| Head size | |||||
| 28 mm | 3,819 | 42 | 1.0 (ref) | 1.0 (ref) | |
| 32 mm | 14,713 | 202 | 1.4 (1.0–1.9) | 1.1 (0.8–1.6) | 0.6 |
| 36 mm | 4,193 | 49 | 1.2 (0.8–1.8) | 1.0 (0.6–1.7) | 0.9 |
| Head material | |||||
| Metal | 18,802 | 242 | 1.0 (ref) | 1.0 (ref) | |
| Ceramic | 3,922 | 51 | 1.1 (0.8–1.5) | 1.0 (0.7–1.5) | 0.8 |
| Cup design | |||||
| Regenerex | 732 | 13 | 1.6 (0.9–2.8) | 1.1 (0.5–2.5) | 0.7 |
| Pinnacle | 2,751 | 38 | 1.3 (0.9–1.9) | 1.5 (0.9–2.5) | 0.2 |
| Pinnacle Gription | 3,364 | 36 | 1.2 (0.8–1.7) | 1.2 (0.6–2.5) | 0.6 |
| Trident AD | 1,815 | 18 | 0.9 (0.5–1.5) | 1.3 (0.7–2.2) | 0.4 |
| Trident Hemi | 3,305 | 31 | 0.9 (0.6–1.4) | 1.5 (0.9–2.5) | 0.2 |
| Tritanium | 547 | 7 | 1.1 (0.5–2.4) | 1.2 (0.4–3.0) | 0.8 |
| Continuum | 3,509 | 72 | 1.9 (1.4–2.6) | 1.8 (1.0–3.3) | 0.06 |
| Trilogy | 6,702 | 78 | 1.0 (ref) | 1.0 (ref) | |
| Hospital | |||||
| Frequent user | 14,985 | 169 | 1.0 (ref) | 1.0 (ref) | |
| Intermediate user | 6,482 | 105 | 1.4 (1.1–1.7) | 1.5 (1.2–1.8) | 0.001 |
| Low-volume user | 1,258 | 19 | 1.3 (0.9–2.0)) | 1.4 (0.9–2.1) | 0.1 |
The adjusted hazard ratio was calculated based a Cox regression model with the following covariates: screw holes, gender, age, surgical approach, stem fixation, cup coating, head size, head material, and cup design.