A novel deformable lung phantom with programably variable external and internal correlation.

Maida Ranjbar1, Pouya Sabouri1, Carlo Repetto1, Amit Sawant1.   

Abstract

PURPOSE: Lung motion phantoms used to validate radiotherapy motion management strategies have fairly simplistic designs that do not adequately capture complex phenomena observed in human respiration such as external and internal deformation, variable hysteresis and variable correlation between different parts of the thoracic anatomy. These limitations make reliable evaluation of sophisticated motion management techniques quite challenging. In this work, we present the design and implementation of a programmable, externally and internally deformable lung motion phantom that allows for a reproducible change in external-internal and internal-internal correlation of embedded markers.
METHODS: An in-house-designed lung module, made from natural latex foam was inserted inside the outer shell of a commercially available lung phantom (RSD, Long Beach, CA, USA). Radiopaque markers were placed on the external surface and embedded into the lung module. Two independently programmable high-precision linear motion actuators were used to generate primarily anterior-posterior (AP) and primarily superior-inferior (SI) motion in a reproducible fashion in order to enable (a) variable correlation between the displacement of interior volume and the exterior surface, (b) independent changes in the amplitude of the AP and SI motions, and (c) variable hysteresis. The ability of the phantom to produce complex and variable motion accurately and reproducibly was evaluated by programming the two actuators with mathematical and patient-recorded lung tumor motion traces, and recording the trajectories of various markers using kV fluoroscopy. As an example application, the phantom was used to evaluate the performance of lung motion models constructed from kV fluoroscopy and 4DCT images.
RESULTS: The phantom exhibited a high degree of reproducibility and marker motion ranges were reproducible to within 0.5 mm. Variable correlation was observed between the displacements of internal-internal and internal-external markers. The SI and AP components of motion of a specific marker had a correlation parameter that varied from -11 to 17. Monitoring a region of interest on the phantom's surface to estimate internal marker motion led to considerably lower uncertainties than when a single point was monitored.
CONCLUSIONS: We successfully designed and implemented a programmable, externally and internally deformable lung motion phantom that allows for a reproducible change in external-internal and internal-internal correlation of embedded markers.
© 2019 American Association of Physicists in Medicine.

Entities:  

Keywords:  deformable lung phantom; programmable; respiratory motion management; surrogate-based lung motion model; variable correlation

Mesh:

Year:  2019        PMID: 30919974      PMCID: PMC6510612          DOI: 10.1002/mp.13507

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  38 in total

1.  A method for incorporating organ motion due to breathing into 3D dose calculations.

Authors:  A E Lujan; E W Larsen; J M Balter; R K Ten Haken
Journal:  Med Phys       Date:  1999-05       Impact factor: 4.071

2.  Acquiring a four-dimensional computed tomography dataset using an external respiratory signal.

Authors:  S S Vedam; P J Keall; V R Kini; H Mostafavi; H P Shukla; R Mohan
Journal:  Phys Med Biol       Date:  2003-01-07       Impact factor: 3.609

3.  4D-CT imaging of a volume influenced by respiratory motion on multi-slice CT.

Authors:  Tinsu Pan; Ting-Yim Lee; Eike Rietzel; George T Y Chen
Journal:  Med Phys       Date:  2004-02       Impact factor: 4.071

4.  Prediction of respiratory tumour motion for real-time image-guided radiotherapy.

Authors:  Gregory C Sharp; Steve B Jiang; Shinichi Shimizu; Hiroki Shirato
Journal:  Phys Med Biol       Date:  2004-02-07       Impact factor: 3.609

5.  The reproducibility of organ position using active breathing control (ABC) during liver radiotherapy.

Authors:  L A Dawson; K K Brock; S Kazanjian; D Fitch; C J McGinn; T S Lawrence; R K Ten Haken; J Balter
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-12-01       Impact factor: 7.038

6.  Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor.

Authors:  H Shirato; S Shimizu; K Kitamura; T Nishioka; K Kagei; S Hashimoto; H Aoyama; T Kunieda; N Shinohara; H Dosaka-Akita; K Miyasaka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2000-09-01       Impact factor: 7.038

7.  Detection of lung tumor movement in real-time tumor-tracking radiotherapy.

Authors:  S Shimizu; H Shirato; S Ogura; H Akita-Dosaka; K Kitamura; T Nishioka; K Kagei; M Nishimura; K Miyasaka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-10-01       Impact factor: 7.038

8.  Evaluation of respiratory movement during gated radiotherapy using film and electronic portal imaging.

Authors:  E C Ford; G S Mageras; E Yorke; K E Rosenzweig; R Wagman; C C Ling
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-02-01       Impact factor: 7.038

9.  Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy.

Authors:  Yvette Seppenwoolde; Hiroki Shirato; Kei Kitamura; Shinichi Shimizu; Marcel van Herk; Joos V Lebesque; Kazuo Miyasaka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-07-15       Impact factor: 7.038

10.  Synchronized moving aperture radiation therapy (SMART): average tumour trajectory for lung patients.

Authors:  Toni Neicu; Hiroki Shirato; Yvette Seppenwoolde; Steve B Jiang
Journal:  Phys Med Biol       Date:  2003-03-07       Impact factor: 3.609

View more
  2 in total

1.  Development and prospective in-patient proof-of-concept validation of a surface photogrammetry + CT-based volumetric motion model for lung radiotherapy.

Authors:  M Ranjbar; P Sabouri; S Mossahebi; D Leiser; M Foote; J Zhang; G Lasio; S Joshi; A Sawant
Journal:  Med Phys       Date:  2019-10-25       Impact factor: 4.071

2.  Validation of a CT-based motion model with in-situ fluoroscopy for lung surface deformation estimation.

Authors:  M Ranjbar; P Sabouri; S Mossahebi; A Sawant; P Mohindra; G Lasio; L D Timmie Topoleski
Journal:  Phys Med Biol       Date:  2021-02-16       Impact factor: 3.609

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.