The reproducibility of organ position using active breathing control (ABC) during liver radiotherapy.

L A Dawson1, K K Brock, S Kazanjian, D Fitch, C J McGinn, T S Lawrence, R K Ten Haken, J Balter.   

Abstract

PURPOSE: To evaluate the intrafraction and interfraction reproducibility of liver immobilization using active breathing control (ABC). METHODS AND MATERIALS: Patients with unresectable intrahepatic tumors who could comfortably hold their breath for at least 20 s were treated with focal liver radiation using ABC for liver immobilization. Fluoroscopy was used to measure any potential motion during ABC breath holds. Preceding each radiotherapy fraction, with the patient setup in the nominal treatment position using ABC, orthogonal radiographs were taken using room-mounted diagnostic X-ray tubes and a digital imager. The radiographs were compared to reference images using a 2D alignment tool. The treatment table was moved to produce acceptable setup, and repeat orthogonal verification images were obtained. The positions of the diaphragm and the liver (assessed by localization of implanted radiopaque intra-arterial microcoils) relative to the skeleton were subsequently analyzed. The intrafraction reproducibility (from repeat radiographs obtained within the time period of one fraction before treatment) and interfraction reproducibility (from comparisons of the first radiograph for each treatment with a reference radiograph) of the diaphragm and the hepatic microcoil positions relative to the skeleton with repeat breath holds using ABC were then measured. Caudal-cranial (CC), anterior-posterior (AP), and medial-lateral (ML) reproducibility of the hepatic microcoils relative to the skeleton were also determined from three-dimensional alignment of repeat CT scans obtained in the treatment position.
RESULTS: A total of 262 fractions of radiation were delivered using ABC breath holds in 8 patients. No motion of the diaphragm or hepatic microcoils was observed on fluoroscopy during ABC breath holds. From analyses of 158 sets of positioning radiographs, the average intrafraction CC reproducibility (sigma) of the diaphragm and hepatic microcoil position relative to the skeleton using ABC repeat breath holds was 2.5 mm (range 1.8-3.7 mm) and 2.3 mm (range 1.2-3.7 mm) respectively. However, based on 262 sets of positioning radiographs, the average interfraction CC reproducibility (sigma) of the diaphragm and hepatic microcoils was 4.4 mm (range 3.0-6.1 mm) and 4.3 mm (range 3.1-5.7 mm), indicating a change of diaphragm and microcoil position relative to the skeleton over the course of treatment with repeat breath holds at the same phase of the respiratory cycle. The average population absolute intrafraction CC offset in diaphragm and microcoil position relative to skeleton was 2.4 mm and 2.1 mm respectively; the average absolute interfraction CC offset was 5.2 mm. Analyses of repeat CT scans demonstrated that the average intrafraction excursion of the hepatic microcoils relative to the skeleton in the CC, AP, and ML directions was 1.9 mm, 0.6 mm, and 0.6 mm respectively and the average interfraction CC, AP, and ML excursion of the hepatic microcoils was 6.6 mm, 3.2 mm, and 3.3 mm respectively.
CONCLUSION: Radiotherapy using ABC for patients with intrahepatic cancer is feasible, with good intrafraction reproducibility of liver position using ABC. However, the interfraction reproducibility of organ position with ABC suggests the need for daily on-line imaging and repositioning if treatment margins smaller than those required for free breathing are a goal.

Entities:  

Mesh:

Year:  2001        PMID: 11728702     DOI: 10.1016/s0360-3016(01)02653-0

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  53 in total

Review 1.  Charged-particle therapy for hepatocellular carcinoma.

Authors:  Heath D Skinner; Theodore S Hong; Sunil Krishnan
Journal:  Semin Radiat Oncol       Date:  2011-10       Impact factor: 5.934

Review 2.  Imaging and image-guided radiation therapy in liver cancer.

Authors:  Kristy K Brock
Journal:  Semin Radiat Oncol       Date:  2011-10       Impact factor: 5.934

3.  Digital tomosynthesis for respiratory gated liver treatment: clinical feasibility for daily image guidance.

Authors:  Q Jackie Wu; Jeffrey Meyer; Jessica Fuller; Devon Godfrey; Zhiheng Wang; Junan Zhang; Fang-Fang Yin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-06-18       Impact factor: 7.038

4.  In vivo reproducibility of robotic probe placement for a novel ultrasound-guided radiation therapy system.

Authors:  Muyinatu A Lediju Bell; H Tutkun Sen; Iulian Iordachita; Peter Kazanzides; John Wong
Journal:  J Med Imaging (Bellingham)       Date:  2014-07-23

Review 5.  Radiation-associated liver injury.

Authors:  Charlie C Pan; Brian D Kavanagh; Laura A Dawson; X Allen Li; Shiva K Das; Moyed Miften; Randall K Ten Haken
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-03-01       Impact factor: 7.038

6.  RapidArc combined with the active breathing coordinator provides an effective and accurate approach for the radiotherapy of hepatocellular carcinoma.

Authors:  G Z Gong; Y Yin; L G Xing; Y J Guo; T Liu; J Chen; J Lu; C Ma; T Sun; T Bai; G Zhang; R Wang
Journal:  Strahlenther Onkol       Date:  2012-02-08       Impact factor: 3.621

7.  Breathing-synchronized delivery: a potential four-dimensional tomotherapy treatment technique.

Authors:  Tiezhi Zhang; Weiguo Lu; Gustavo H Olivera; Harry Keller; Robert Jeraj; Rafael Manon; Minesh Mehta; Thomas R Mackie; Bhudatt Paliwal
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-06-14       Impact factor: 7.038

Review 8.  MRI in medical practice and its future use in radiation oncology. Resume of XXV GOCO Congress (Montpellier) 2017.

Authors:  Xavier Druet; Estrella Acosta Sanchez; Ken Soleakhena; Anne Laprie; Jordi Sáez; Stéphanie Nougaret; Olivier Riou; Elodie Rigal; Laura Kibranian; Miguel Palacios; Ismael Membrive
Journal:  Rep Pract Oncol Radiother       Date:  2019-06-05

Review 9.  Solutions that enable ablative radiotherapy for large liver tumors: Fractionated dose painting, simultaneous integrated protection, motion management, and computed tomography image guidance.

Authors:  Christopher H Crane; Eugene J Koay
Journal:  Cancer       Date:  2016-03-07       Impact factor: 6.860

10.  Magnitude of shift of tumor position as a function of moderated deep inspiration breath-hold: An analysis of pooled data of lung patients with active breath control in image-guided radiotherapy.

Authors:  K R Muralidhar; P Narayana Murthy; D Shankar Mahadev; K Subramanyam; G Sudarshan; A Krishnam Raju
Journal:  J Med Phys       Date:  2008-10
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.